Device-induced platelet dysfunction in mechanically assisted circulation increases the risks of thrombosis and bleeding

被引:37
作者
Chen, Zengsheng [1 ]
Zhang, Jiafeng [1 ]
Kareem, Kafayat [1 ]
Tran, Douglas [1 ]
Conway, Robert G. [1 ]
Arias, Katherin [1 ,2 ]
Griffith, Bartley P. [1 ]
Wu, Zhongjun J. [1 ,2 ]
机构
[1] Univ Maryland, Sch Med, Dept Surg, 10 South Pine St,MSTF 434A, Baltimore, MD 21201 USA
[2] Univ Maryland, A James Clark Sch Engn, Fischell Dept Bioengn, College Pk, MD 20742 USA
关键词
continuous flow ventricular assist device; hemostasis dysfunction; high mechanical shear stress; platelet; von Willebrand factor; VON-WILLEBRAND SYNDROME; GLYCOPROTEIN IB-ALPHA; SHEAR-STRESS; PUMP THROMBOSIS; FLOW; ACTIVATION; TIME; VI; ANTICOAGULATION; HEMOLYSIS;
D O I
10.1111/aor.13445
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Thrombotic and bleeding complications are the major obstacles for expanding mechanical circulatory support (MCS) beyond the current use. While providing the needed hemodynamic support, those devices can induce damage to blood, particularly to platelets. In this study, we investigated device-induced alteration of three major platelet surface receptors, von Willebrand factor (VWF) and associated hemostatic functions relevant to thrombosis and bleeding. Fresh human whole blood was circulated in an extracorporeal circuit with a clinical rotary blood pump (CentriMag, Abbott, Chicago, IL, USA) under the clinically relevant operating condition for 4hours. Blood samples were examined every hour for glycoprotein (GP) IIb/IIIa activation and receptor loss of GPVI and GPIb alpha on the platelet surface with flow cytometry. Soluble P-selectin in hourly collected blood samples was measured by enzyme linked immunosorbent assay to characterize platelet activation. Adhesion of device-injured platelets to fibrinogen, collagen, and VWF was quantified with fluorescent microscopy. Device-induced damage to VWF was characterized with western blotting. The CentriMag blood pump induced progressive platelet activation with blood circulating time. Particularly, GPIIb/IIIa activation increased from 1.1% (Base) to 11% (4hours) and soluble P-selectin concentration increased from 14.1ng/mL (Base) to 26.5ng/mL (4hours). Those device-activated platelets exhibited increased adhesion capacity to fibrinogen. Concurrently, the CentriMag blood pump caused progressive platelet receptor loss (GPVI and GPIb alpha) with blood circulating time. Specifically, MFI of the GPVI and GPIb alpha receptors decreased by 17.2% and 16.1% for the 4-hours sample compared to the baseline samples, respectively. The device-injured platelets exhibited reduced adhesion capacities to collagen and VWF. The high molecular weight multimers (HMWM) of VWF in the blood disappeared within the first hour of the circulation. Thereafter the multimeric patterns of VWF were stable. The change in the VWF multimeric pattern was different from the progressive structural and functional changes of platelets with the circulation time. This study suggested that the CentriMag blood pump could induce two opposite effects on platelets and associated hemostatic functions under a clinically relevant operating condition. The device-altered hemostatic function may contribute to thrombosis and bleeding simultaneously as occurring in patients supported by a rotary blood pump. Device-induced damage of platelets may be an important cause for bleeding in patients supported with rotary blood pump MCS systems relative to device-induced loss of HMWM-VWF.
引用
收藏
页码:745 / 755
页数:11
相关论文
共 39 条
  • [1] Anticoagulation management in mechanical circulatory support
    Adatya, Sirtaz
    Bennett, Mosi K.
    [J]. JOURNAL OF THORACIC DISEASE, 2015, 7 (12) : 2129 - 2138
  • [2] Haemolysis as a first sign of thromboembolic event and acute pump thrombosis in patients with the continuous-flow left ventricular assist device HeartMate II
    Akin, S.
    Soliman, O. I.
    Constantinescu, A. A.
    Akca, F.
    Birim, O.
    van Domburg, R. T.
    Manintveld, O.
    Caliskan, K.
    [J]. NETHERLANDS HEART JOURNAL, 2016, 24 (02) : 134 - 142
  • [3] Pathologic shear triggers shedding of vascular receptors: a novel mechanism for down-regulation of platelet glycoprotein VI in stenosed coronary vessels
    Al-Tamimi, Mohammad
    Tan, Chee Wee
    Qiao, Jianlin
    Pennings, Gabrielle J.
    Javadzadegan, Ashkan
    Yong, Andy S. C.
    Arthur, Jane F.
    Davis, Amanda K.
    Jing, Jing
    Mu, Fi-Tjen
    Hamilton, Justin R.
    Jackson, Shaun P.
    Ludwig, Andreas
    Berndt, Michael C.
    Ward, Christopher M.
    Kritharides, Leonard
    Andrews, Robert K.
    Gardiner, Elizabeth E.
    [J]. BLOOD, 2012, 119 (18) : 4311 - 4320
  • [4] ARABNIA HR, 2016, EMERGING TRENDS IN A
  • [5] Platelet glycoprotein VI-related clinical defects
    Arthur, Jane F.
    Dunkley, Scott
    Andrews, Robert K.
    [J]. BRITISH JOURNAL OF HAEMATOLOGY, 2007, 139 (03) : 363 - 372
  • [6] Pump thrombosis-A riddle wrapped in a mystery inside an enigma
    Blitz, Arie
    [J]. ANNALS OF CARDIOTHORACIC SURGERY, 2014, 3 (05) : 450 - 471
  • [7] Flow features and device-induced blood trauma in CF-VADs under a pulsatile blood flow condition: A CFD comparative study
    Chen, Zengsheng
    Jena, Sofen K.
    Giridharan, Guruprasad A.
    Koenig, Steven C.
    Slaughter, Mark S.
    Griffith, Bartley P.
    Wu, Zhongjun J.
    [J]. INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2018, 34 (02)
  • [8] Paradoxical Effect of Nonphysiological Shear Stress on Platelets and von Willebrand Factor
    Chen, Zengsheng
    Mondal, Nandan K.
    Ding, Jun
    Koenig, Steven C.
    Slaughter, Mark S.
    Wu, Zhongjun J.
    [J]. ARTIFICIAL ORGANS, 2016, 40 (07) : 659 - 668
  • [9] Activation and shedding of platelet glycoprotein IIb/IIIa under non-physiological shear stress
    Chen, Zengsheng
    Mondal, Nandan K.
    Ding, Jun
    Koenig, Steven C.
    Slaughter, Mark S.
    Griffith, Bartley P.
    Wu, Zhongjun J.
    [J]. MOLECULAR AND CELLULAR BIOCHEMISTRY, 2015, 409 (1-2) : 93 - 101
  • [10] Shear-induced platelet receptor shedding by non-physiological high shear stress with short exposure time: Glycoprotein Ibα and glycoprotein VI
    Chen, Zengsheng
    Mondal, Nandan K.
    Ding, Jun
    Gao, Jingya
    Griffith, Bartley P.
    Wu, Zhongjun J.
    [J]. THROMBOSIS RESEARCH, 2015, 135 (04) : 692 - 698