Development of a Retention Time Interpolation scale (RTi) for liquid chromatography coupled to mass spectrometry in both positive and negative ionization modes

被引:12
作者
Celma, Alberto [1 ]
Bijlsma, Lubertus [1 ]
Lopez, Francisco J. [1 ]
Sancho, Juan V. [1 ]
机构
[1] Univ Jaume 1, Res Inst Pesticides & Water, Avda Sos Baynat S-N, E-12071 Castellon de La Plana, Spain
关键词
Retention time interpolation scale; Liquid chromatography; High resolution mass spectrometry; Isotopically labelled reference standards; Kovats index; Screening; SYNTHETIC CANNABINOIDS; NONTARGET ANALYSIS; WASTE-WATER; IDENTIFICATION; DRUGS; SURFACE; CONTAMINANTS; METABOLITES; POLLUTANTS; PREDICTION;
D O I
10.1016/j.chroma.2018.07.030
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The accuracy and sensitivity of high resolution mass spectrometry (HRMS) enables the identification of candidate compounds with the use of mass spectrometric databases among other tools. However, retention time (RT) data in identification workflows has been sparingly used since it could be strongly affected by matrix or chromatographic performance. Retention Time Interpolation scaling (RTi) strategies can provide a more robust and valuable information than RT, gaining more confidence in the identification of candidate compounds in comparison to an analytical standard. Up to our knowledge, no RTi has been developed for LC-HRMS systems providing information when acquiring in either positive or negative ionization modes. In this work, an RTi strategy was developed by means of the use of 16 isotopically labelled reference standards, which can be spiked into a real sample without resulting in possible false positives or negatives. For testing the RTi performance, a mixture of several reference standards, emulating suspect analytes, were used. RTi values for these compounds were calculated both in solvent and spiked in a real matrix to assess the effect of either chromatographic parameters or matrix in different scenarios. It has been demonstrated that the variation of injection volume, chromatographic gradient and initial percentage of organic solvent injected does not considerably affect RTi calculation. Column aging and solid support of the stationary phase of the column, however, showed strong effects on the elution of several test compounds. Yet, RTi permitted the correction of elution shifts of most compounds. Furthermore, RTi was tested in 47 different matrices from food, biological, animal feeding and environmental origin. The application of RTi in both positive and negative ionization modes showed in general satisfactory results for most matrices studied. The RTi developed can be used in future LC-HRMS screening analysis giving an additional parameter, which facilitates tedious processing tasks and gain more confidence in the identification of (non)-suspect analytes. (C) 2018 Elsevier B.V. All rights reserved.
引用
收藏
页码:101 / 107
页数:7
相关论文
共 29 条
[1]   Advances in liquid chromatography-high-resolution mass spectrometry for quantitative and qualitative environmental analysis [J].
Acena, Jaume ;
Stampachiacchiere, Serena ;
Perez, Sandra ;
Barcelo, Damia .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2015, 407 (21) :6289-6299
[2]  
[Anonymous], 2012, EMEACHMPEWP192217200
[3]  
Ardrey B., 2003, LIQUID CHROMATOGRAPH, DOI [10.1002/0470867299, DOI 10.1002/0470867299]
[4]   Chromatographic retention indices in identification of chemical compounds [J].
Babushok, V. I. .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2015, 69 :98-104
[5]   Suspect screening of large numbers of emerging contaminants in environmental waters using artificial neural networks for chromatographic retention time prediction and high resolution mass spectrometry data analysis [J].
Bade, Richard ;
Bijlsma, Lubertus ;
Miller, Thomas H. ;
Barron, Leon P. ;
Vicente Sancho, Juan ;
Hernandez, Felix .
SCIENCE OF THE TOTAL ENVIRONMENT, 2015, 538 :934-941
[6]   Determination of six microcystins and nodularin in surface and drinking waters by on-line solid phase extraction-ultra high pressure liquid chromatography tandem mass spectrometry [J].
Beltran, Eduardo ;
Ibanez, Maria ;
Vicente Sancho, Juan ;
Hernandez, Felix .
JOURNAL OF CHROMATOGRAPHY A, 2012, 1266 :61-68
[7]   Surface and wastewater quality monitoring: combination of liquid chromatography with (geno)toxicity detection, diode array detection and tandem mass spectrometry for identification of pollutants [J].
Bobeldijk, I ;
Stoks, PGM ;
Vissers, JPC ;
Emke, E ;
van Leerdam, JA ;
Muilwijk, B ;
Berbee, R ;
Noij, THM .
JOURNAL OF CHROMATOGRAPHY A, 2002, 970 (1-2) :167-181
[8]   The Chromatographic Role in High Resolution Mass Spectrometry for Non-Targeted Analysis [J].
Croley, Timothy R. ;
White, Kevin D. ;
Callahan, John H. ;
Musser, Steven M. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2012, 23 (09) :1569-1578
[9]  
European Commission,, 2018, SANTE118132017 EUR C
[10]  
Grosse S., 2013, LANGENAUER WATER FOR