Comparative study on the emission spectrometric determination of manganese using nitrogen-oxygen Okamoto-cavity microwave induced plasma and argon radio-frequency inductively-coupled plasma

被引:4
作者
Arai, Yuuki [1 ]
Sato, Shigeo [1 ]
Wagatsuma, Kazuaki [1 ]
机构
[1] Tohoku Univ, Inst Mat Res, Sendai, Miyagi 9808577, Japan
关键词
Optical emission spectrometry; Microwave-induced plasma; Okamoto-cavity; Nitrogen-oxygen mixed gas; Inductively-coupled plasma; Manganese; ATOMIC-EMISSION;
D O I
10.1016/j.microc.2014.04.017
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Microwave-induced plasma optical emission spectrometry (MIP-OES) using Okamoto-cavity was investigated in the determination of manganese, in comparison with argon radio-frequency inductively-coupled plasma (ICP)-OES. An important point of this study is that, in the Okamoto-cavity MIP, the emission intensity of particular atomic emission lines is much elevated by adding oxygen of ca. 10% to the nitrogen matrix plasma gas. Intensity variations in emission lines of manganese atom as well as the ion and in a band head of nitrogen molecule were observed, when oxygen was added to the nitrogen plasma, indicating that the emission intensity of the atomic lines was drastically elevated while the intensities of the ionic lines and the nitrogen band head decreased in a similar manner. This observation can be explained by the fact that the nitrogen excited species, contributing to the transition of the band head, are consumed through collisions with oxygen molecules to cause their dissociation, and eventually manganese atoms in the nitrogen-oxygen plasma would be less ionized than in the pure nitrogen plasma, because the ionization of manganese is also caused by collisions with the nitrogen excited species. Optimization of the measuring parameters in the Okamoto-cavity MIP-OES was conducted for the determination of manganese; as a result, the Mn I 403.076-nm line was recommended as the analytical line in the nitrogen-oxygen mixed gas MIP, while the most intense ionic line, Mn II 257.065 nm, should be employed in the argon ICP. Their detection limits were of the same order, implying that the Okamoto-cavity MIP had as good analytical performance as the conventional ICP. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:135 / 141
页数:7
相关论文
共 21 条
[1]   Emission Spectrometric Analysis Using an Okamoto-cavity Microwave-induced Plasma with Nitrogen-Oxygen Mixed Gas [J].
Arai, Yuuki ;
Sato, Shigeo ;
Wagatsuma, Kazuaki .
ISIJ INTERNATIONAL, 2013, 53 (11) :1993-1999
[2]   CAVITY FOR MICROWAVE-INDUCED PLASMAS OPERATED IN HELIUM AND ARGON AT ATMOSPHERIC-PRESSURE [J].
BEENAKKER, CIM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1976, 31 (8-9) :483-486
[3]  
Corliss C.H., 1962, NBS Monograph, V53
[4]  
Gaydon A., 1953, Dissociation Energies and Spectra of Diatomic Molecules, V2nd
[5]   NEW MICROWAVE PLASMA AT ATMOSPHERIC-PRESSURE [J].
HUBERT, J ;
MOISAN, M ;
RICARD, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1979, 34 (01) :1-10
[6]   Two-dimensional observation of emission spectra excited from laser induced Plasmas and the application to emission spectrometric analysis [J].
Kitaoka, Chizuru ;
Wagatsuma, Kazuaki .
ANALYTICAL SCIENCES, 2007, 23 (11) :1261-1265
[7]  
Maeda T, 2005, TETSU TO HAGANE, V91, P35
[8]   Direct loading of ethanol solution into high-power nitrogen-oxygen mixed gas microwave-induced plasma and the emission characteristics [J].
Maeda, T ;
Wagatsuma, K .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2005, 60 (01) :81-87
[9]   Emission characteristics of high-powered microwave induced plasma optical emission spectrometry by using nitrogen-oxygen mixture gas [J].
Maeda, T ;
Wagatsuma, K .
MICROCHEMICAL JOURNAL, 2004, 76 (1-2) :53-60
[10]  
Nakahara T., 2002, ISIJ INT, V42, ps114