Clinical Evaluation of a New Swept-Source Optical Coherence Biometer That Uses Individual Refractive Indices to Measure Axial Length in Cataract Patients

被引:34
作者
Tamaoki, Akeno [1 ,3 ]
Kojima, Takashi [2 ]
Hasegawa, Asato [1 ]
Yamamoto, Mana [1 ]
Kaga, Tatsushi [1 ]
Tanaka, Kiyoshi [3 ]
Ichikawa, Kazuo [1 ,4 ]
机构
[1] Japan Community Healthcare Org, Dept Ophthalmol, Chukyo Hosp, Nagoya, Aichi, Japan
[2] Keio Univ, Dept Ophthalmol, Sch Med, Tokyo, Japan
[3] Shinshu Univ, Dept Math & Syst Dev, Interdisciplinary Grad Sch Sci & Technol, Nagano, Japan
[4] Chukyo Eye Clin, Nagoya, Aichi, Japan
关键词
Optical coherence tomography biometer; Individual refractive index; Axial length; Cataract surgery; Equivalent refractive index; Refractive error; LENS POWER CALCULATION; TOMOGRAPHY; EYES; INTERFEROMETRY; REPEATABILITY; FORMULAS; AGREEMENT; IOLMASTER; ACCURACY;
D O I
10.1159/000496690
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Backgrounds: Although the OCT biometer using individual refractive index is available, comparisons of measurement value and intraocular lens (IOL) power calculation error with other SS-OCT biometers are not known. Objectives: To compare the new SS-OCT biometer ARGOS (OCTB1), which uses individual refractive indices to measure axial length, with the IOLMaster 700 (OCTB2) and OA-2000 (OLCR), which use equivalent refractive index. Method: Six hundred and twenty-two eyes of 622 patients who had been diagnosed with cataract were enrolled in the study. Among the 158 eyes that had undergone cataract surgery, the postoperative refractive error was evaluated using the Haigis formula. Results: The axial length measured by the OCTB1 showed a proportional bias in comparison with the other two biometers and a fixed bias in eyes with an axial length 26 mm. No significant difference was found in the median absolute refractive prediction error (p = 0.3278). However, in eyes with an axial length 26 mm, the OCTB1 showed myopic error compared with the other two biometers (p < 0.0001). Conclusions: In eyes with long axial length, when the conventional IOL calculation was optimized with the equivalent refractive index-based instrument, we need to consider that IOL calculation using OCTB1 tends to cause slightly myopic refractive prediction error.
引用
收藏
页码:11 / 23
页数:13
相关论文
共 25 条
[1]   Intraocular lens power calculation for eyes with an axial length greater than 26.0 mm: Comparison of formulas and methods [J].
Abulafia, Adi ;
Barrett, Graham D. ;
Rotenberg, Michael ;
Kleinmann, Guy ;
Levy, Adi ;
Reitblat, Olga ;
Koch, Douglas D. ;
Wang, Li ;
Assia, Ehud I. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (03) :548-556
[2]   The thickness of the aging human lens obtained from corrected scheimpflug images [J].
Dubbelman, M ;
Van der Heijde, GL ;
Weeber, HA .
OPTOMETRY AND VISION SCIENCE, 2001, 78 (06) :411-416
[3]   Accuracy and feasibility of axial length measurements by a new optical low-coherence reflectometry-based device in eyes with posterior subcapsular cataract [J].
Einan-Lifshitz, Adi ;
Rozenberg, Assaf ;
Wang, Li ;
Koch, Douglas D. ;
Shoshany, Nadav ;
Zadok, David ;
Avni, Isaac ;
Abulafia, Adi .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (07) :898-901
[4]  
Findl Oliver, 2005, Curr Opin Ophthalmol, V16, P61, DOI 10.1097/00055735-200502000-00011
[5]   Understanding Bland Altman analysis [J].
Giavarina, Davide .
BIOCHEMIA MEDICA, 2015, 25 (02) :141-151
[6]   Comparison of 3 biometry devices in cataract patients [J].
Goebels, Susanne ;
Pattmoeller, Max ;
Eppig, Timo ;
Cayless, Alan ;
Seitz, Berthold ;
Langenbucher, Achim .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (11) :2387-2393
[7]   Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis [J].
Haigis, W ;
Lege, B ;
Miller, N ;
Schneider, B .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2000, 238 (09) :765-773
[8]   Evaluation of a new IOLMaster algorithm to measure axial length [J].
Hill, Warren ;
Angeles, Raymund ;
Otani, Todd .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (06) :920-924
[9]   Enhanced Penetration for Axial Length Measurement of Eyes with Dense Cataracts Using Swept Source Optical Coherence Tomography: A Consecutive Observational Study [J].
Hirnschall, Nino ;
Varsits, Ralph ;
Doeller, Birgit ;
Findl, Oliver .
OPHTHALMOLOGY AND THERAPY, 2018, 7 (01) :119-124
[10]   Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry [J].
Hoffer, Kenneth J. ;
Hoffmann, Peter C. ;
Savini, Giacomo .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (08) :1165-1172