Permanent Leadless Cardiac Pacemaker Therapy: A Comprehensive Review

被引:173
作者
Tjong, Fleur V. Y. [1 ]
Reddy, Vivek Y. [2 ]
机构
[1] Acad Med Ctr, AMC Heart Ctr, Dept Clin & Expt Cardiol, Amsterdam, Netherlands
[2] Icahn Sch Med Mt Sinai, Helmsley Electrophysiol Ctr, One Gustave L Levy Pl,Box 1030, New York, NY 10029 USA
关键词
arrhythmias; cardiac; cardiac pacing; artificial; leadless pacing; pacemaker; RISK-FACTORS; CARDIOVERTER-DEFIBRILLATORS; DEVICE INFECTION; SINGLE-CHAMBER; TASK-FORCE; FOLLOW-UP; COMPLICATIONS; IMPLANTATION; PREDICTORS; FEASIBILITY;
D O I
10.1161/CIRCULATIONAHA.116.025037
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A new technology, leadless pacemaker therapy, was recently introduced clinically to address lead- and pocket-related complications in conventional transvenous pacemaker therapy. These leadless devices are self-contained right ventricular single-chamber pacemakers implanted by using a femoral percutaneous approach. In this review of available clinical data on leadless pacemakers, early results with leadless devices are compared with historical results with conventional single-chamber pacing. Both presently manufactured leadless pacemakers show similar complications, which are mostly related to the implant procedure: cardiac perforation, device dislocation, and femoral vascular access site complications. In comparison with conventional transvenous single-chamber pacemakers, slightly higher short-term complication rates have been observed: 4.8% for leadless pacemakers versus 4.1% for conventional pacemakers. The complication rate of the leadless pacemakers is influenced by the implanter learning curve for this new procedure. No long-term outcome data are yet available for the leadless pacemakers. Larger leadless pacing trials, with long-term follow-up and direct randomized comparison with conventional pacing systems, will be required to define the proper clinical role of these leadless systems. Although current leadless pacemakers are limited to right ventricular pacing, future advanced, communicating, multicomponent systems are expected to expand the potential benefits of leadless therapy to a larger patient population.
引用
收藏
页码:1458 / 1470
页数:13
相关论文
共 59 条
[1]  
AGGARWAL RK, 1995, BRIT HEART J, V73, P571
[2]   Long-term follow-up of patients from a randomised trial of atrial versus ventricular pacing for sick-sinus syndrome [J].
Andersen, HR ;
Nielsen, JC ;
Thomsen, PEB ;
Thuesen, L ;
Mortensen, PT ;
Vesterlund, T ;
Pedersen, AK .
LANCET, 1997, 350 (9086) :1210-1216
[3]  
[Anonymous], 2016, EX SUMM CIRC SYST DE
[4]   Feasibility, safety, and short-term outcome of leadless ultrasound-based endocardial left ventricular resynchronization in heart failure patients: results of the Wireless Stimulation Endocardially for CRT (WiSE-CRT) study [J].
Auricchio, Angelo ;
Delnoy, Peter-Paul ;
Butter, Christian ;
Brachmann, Johannes ;
Van Erven, Lieselot ;
Spitzer, Stefan ;
Moccetti, Tiziano ;
Seifert, Martin ;
Markou, Thanasie ;
Laszo, Karolyi ;
Regoli, Francois .
EUROPACE, 2014, 16 (05) :681-688
[5]   Outcomes of patients requiring emergent surgical or endovascular intervention for catastrophic complications during transvenous Lead extraction [J].
Brunner, Michael P. ;
Cronin, Edmond M. ;
Wazni, Oussama ;
Baranowski, Bryan ;
Saliba, Walid I. ;
Sabik, Joseph F. ;
Lindsay, Bruce D. ;
Wilkoff, Bruce L. ;
Tarakji, Khaldoun G. .
HEART RHYTHM, 2014, 11 (03) :419-425
[6]   Multiple leadless pacemakers implanted in the right ventricle of swine [J].
Chen, Keping ;
Zheng, Xiaolin ;
Dai, Yan ;
Wang, Hao ;
Tang, Yue ;
Lan, Tingyu ;
Zhang, Jinping ;
Tian, Yi ;
Zhang, Baojie ;
Zhou, Xiaohong ;
Bonner, Matthew ;
Zhang, Shu .
EUROPACE, 2016, 18 (11) :1748-1752
[7]   Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes [J].
Connolly, SJ ;
Kerr, CR ;
Gent, M ;
Roberts, RS ;
Yusuf, S ;
Gillis, AM ;
Sami, MH ;
Talajic, M ;
Tang, ASL ;
Klein, GJ ;
Lau, C ;
Newman, DM .
NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (19) :1385-1391
[8]   Multimodal imaging of a transcatheter pacemaker implantation within reanimated human heart [J].
Eggen, Michael D. ;
Bonner, Matthew D. ;
Williams, Eric R. ;
Iaizzo, Paul A. .
HEART RHYTHM, 2014, 11 (12) :2331-2332
[9]   Complications arising after implantation of DDD pacemakers: The MOST experience [J].
Ellenbogen, KA ;
Hellkamp, AS ;
Wilkoff, BL ;
Camunas, JL ;
Love, JC ;
Hadjis, TA ;
Lee, KL ;
Lamas, GA .
AMERICAN JOURNAL OF CARDIOLOGY, 2003, 92 (06) :740-741
[10]  
Ellenbogen KA, 2000, CLIN CARDIAC PACING