摘要
The purpose of the study was to obtain useful information for diagnosing internal abnormalities of bones by ultrasound. We fabricated a transmitting and receiving system mainly targeting the bone and an image processing system using a computer, and studied the attenuation of ultrasound in animal bone (in vitro). The average value was about 37 dB. The examination was able to determine successfully the proper dynamic range for detection, and a B-mode image of the inside of the bone could be displayed. Here we describe a new transmitting and receiving system developed to improve the detection of echo signals from inside bones. In this method, two kinds of high sensitivity transducer were made as follows: (1) 5 HMz frequency, 6.5 mm diameter; (2) 10 HMz frequency, 13 mm diameter. By using this method, detectability of the echo signals can be increased more than with the former system. The results showed that the structure of the bones could be clearly seen in the B-mode images. (C) 2000 Published by Elsevier Science B.V. All rights reserved.
引用
收藏
页码:698 / 702
页数:5
相关论文
共 50 条
  • [21] Pulse-echo ultrasound method for detection of post-menopausal women with osteoporotic BMD
    Karjalainen, J. P.
    Riekkinen, O.
    Kroger, H.
    OSTEOPOROSIS INTERNATIONAL, 2018, 29 (05) : 1193 - 1199
  • [22] Pulse-echo ultrasound method for detection of post-menopausal women with osteoporotic BMD
    J. P. Karjalainen
    O. Riekkinen
    H. Kröger
    Osteoporosis International, 2018, 29 : 1193 - 1199
  • [23] Fast algorithm for Probabilistic bone edge detection (FAPBED)
    Scepanovic, D
    Kirshtein, J
    Jain, AK
    Taylor, RH
    MEDICAL IMAGING 2005: IMAGE PROCESSING, PT 1-3, 2005, 5747 : 1753 - 1765
  • [24] Effect of a constant magnetic field on echo signals in high-temperature superconductor powders
    Apushkinskii, E. G.
    Astrov, M. S.
    TECHNICAL PHYSICS, 2013, 58 (03) : 455 - 458
  • [25] Simulation on Bone Porosity Estimation Using Pulse Echo Technique Ultrasound
    Adam, Izzat Irfan Muhammad
    Mahmood, Nasrul Humaimi
    Sudirman, Rubita
    Abd Wahab, Muhamad Amin
    2022 2ND INTERNATIONAL CONFERENCE ON INTELLIGENT CYBERNETICS TECHNOLOGY & APPLICATIONS (ICICYTA), 2022, : 250 - 254
  • [26] Robust Bone Detection in Ultrasound Using Combined Strain Imaging and Envelope Signal Power Detection
    Hussain, Mohammad Arafat
    Hodgson, Antony
    Abugharbieh, Rafeef
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2014, PT I, 2014, 8673 : 356 - +
  • [27] Application of an Improved Ultrasound Full-Waveform Inversion in Bone Quantitative Measurement
    Suo, Meng
    Zhang, Dong
    Yang, Yan
    SYMMETRY-BASEL, 2021, 13 (02): : 1 - 29
  • [28] Improved digital SEM of cancellous bone: scanning direction of detection, through focus for in-focus and sample orientation
    Boyde, A
    JOURNAL OF ANATOMY, 2003, 202 (02) : 183 - 194
  • [29] Bone Remodeling Is Regulated by Inner Ear Vestibular Signals
    Vignaux, Guillaume
    Besnard, Stephane
    Ndong, Jean
    Philoxene, Bruno
    Denise, Pierre
    Elefteriou, Florent
    JOURNAL OF BONE AND MINERAL RESEARCH, 2013, 28 (10) : 2136 - 2144
  • [30] Echo decorrelation from displacement gradients
    Cespedes, EI
    de Korte, CL
    van der Steen, AFW
    ULTRASONICS, 1998, 36 (1-5) : 661 - 666