The artificial urinary sphincter in 86 patients with intrinsic sphincter deficiency:: Satisfactory actuarial adequate function rates

被引:16
作者
Bosch, JLHR
Klijn, AJ
Schröder, FH
Hop, WCJ
机构
[1] Acad Hosp Rotterdam Dijkzigt, Dept Urol H1073, NL-3000 CA Rotterdam, Netherlands
[2] Acad Hosp Rotterdam Dijkzigt, Dept Epidemiol & Biostat, NL-3000 CA Rotterdam, Netherlands
[3] Erasmus Univ, Rotterdam, Netherlands
关键词
artificial urinary sphincter; intrinsic sphincter deficiency; incontinence;
D O I
10.1159/000020273
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objective: To evaluate the long-term results of the AMS-800 artificial urinary sphincter (AUS) in patients with incontinence due to intrinsic sphincter deficiency, taking into account the continence status and the durability of the device. Patients and Methods: Between 1984 and 1997, an AUS was implanted in 86 patients. Kaplan-Meier survival analysis was used to determine the 'primary adequate function' rate (P-AF) and the 'additional procedure assisted adequate function' rate (APA-AF). These rates define adequate function as satisfactory continence (use of maximally 1 pad per 24 h) in combination with good AUS function. In the case of P-AF, no revisions have been required. In the case of APA-AF, one or more revisions of parts of the AUS are acceptable but a complete exchange or explantation of the AUS marks its endpoint. Results: At last follow-up, satisfactory continence was found in 76% of the patients. Continence was markedly improved in another 7%. The 5-year P-AF and APA-AF rates were 46 and 67%, respectively. The 5-year P-AF rates for the periods before and after the introduction of the narrow backed cuff were 33 and 61%, respectively (p = 0.03). Conclusions: The AUS can give excellent results as far as urinary continence is concerned, but only at the expense of a considerable reoperation rate. The 5-year actuarial primary adequate function rate has almost doubled since the introduction of the narrow backed cuff design. Copyright (C) 2000 S. Karger AG. Basel.
引用
收藏
页码:156 / 160
页数:5
相关论文
共 14 条
[1]  
*AHPCR, 1996, AHPCR PUBL
[2]  
BARRETT DM, 1989, UROL CLIN N AM, V16, P119
[3]   Mayo Clinic long-term analysis of the functional durability of the AMS 800 artificial urinary sphincter: A review of 323 cases [J].
Elliott, DS ;
Barrett, DM .
JOURNAL OF UROLOGY, 1998, 159 (04) :1206-1208
[4]   EXPERIENCE WITH THE ARTIFICIAL URINARY SPHINCTER MODEL-AS800 IN 148 PATIENTS [J].
FISHMAN, IJ ;
SHABSIGH, R ;
SCOTT, FB .
JOURNAL OF UROLOGY, 1989, 141 (02) :307-310
[5]   The fate of the 'modern' artificial urinary sphincter with a follow-up of more than 10 years [J].
Fulford, SCV ;
Sutton, C ;
Bales, G ;
Hickling, M ;
Stephenson, TP .
BRITISH JOURNAL OF UROLOGY, 1997, 79 (05) :713-716
[6]   THE MODEL AS-800 ARTIFICIAL URINARY SPHINCTER - MAYO-CLINIC EXPERIENCE [J].
GOLDWASSER, B ;
FURLOW, WL ;
BARRETT, DM .
JOURNAL OF UROLOGY, 1987, 137 (04) :668-671
[7]   NONPARAMETRIC-ESTIMATION FROM INCOMPLETE OBSERVATIONS [J].
KAPLAN, EL ;
MEIER, P .
JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1958, 53 (282) :457-481
[8]  
Klijn AJ, 1998, BRIT J UROL, V82, P530
[9]   SUCCESS OF THE NARROW-BACKED CUFF DESIGN OF THE AMS800 ARTIFICIAL URINARY SPHINCTER - ANALYSIS OF 144 PATIENTS [J].
LEO, ME ;
BARRETT, DM .
JOURNAL OF UROLOGY, 1993, 150 (05) :1412-1414
[10]   IMPACT OF THE NEW CUFF DESIGN ON RELIABILITY OF THE AS800 ARTIFICIAL URINARY SPHINCTER [J].
LIGHT, JK ;
REYNOLDS, JC .
JOURNAL OF UROLOGY, 1992, 147 (03) :609-611