CT-based attenuation correction of whole-body radiotherapy treatment positioning devices in PET/MRI hybrid imaging

被引:4
作者
Taeubert, Leticia [1 ,2 ]
Berker, Yannick [3 ,4 ,5 ]
Beuthien-Baumann, Bettina [6 ,7 ]
Hoffmann, Aswin L. [8 ,9 ,10 ,11 ,12 ]
Troost, Esther G. C. [7 ,8 ,9 ,10 ,11 ,12 ,13 ,14 ,15 ,16 ,17 ,18 ]
Kachelriess, Marc [3 ,19 ]
Gillmann, Clarissa [1 ,2 ]
机构
[1] German Canc Res Ctr, Div Med Phys Radiat Oncol, Heidelberg, Germany
[2] Heidelberg Inst Radiat Oncol, Natl Ctr Radiat Res Oncol NCRO, Heidelberg, Germany
[3] German Canc Res Ctr, Div Xray Imaging & Comp Tomog, Heidelberg, Germany
[4] German Canc Res Ctr, Clin Cooperat Unit Pediat Oncol, Translat Pediat Pharmacol, Heidelberg, Germany
[5] Hopp Childrens Canc Ctr Heidelberg KiTZ, Heidelberg, Germany
[6] German Canc Res Ctr, Div Radiol, Heidelberg, Germany
[7] Natl Ctr Tumor Dis NCT, Partner Site, Heidelberg, Germany
[8] Tech Univ Dresden, Helmholtz Zentrum Dresden Rossendorf, Fac Med, OncoRay Natl Ctr Radiat Res Oncol, Dresden, Germany
[9] Tech Univ Dresden, Helmholtz Zentrum Dresden Rossendorf, Univ Hosp Carl Gustav Carus, Dresden, Germany
[10] Fac Med, Dept Radiotherapy & Radiat Oncol, Dresden, Germany
[11] Univ Hosp Carl Gustav Carus, Dresden, Germany
[12] Helmholtz Zentrum Dresden Rossendorf, Inst Radiooncol OncoRay, Dresden, Germany
[13] German Canc Consortium DKTK, Partner Site Dresden, Heidelberg, Germany
[14] German Canc Res Ctr, Heidelberg, Germany
[15] Partner Site, Dresden, Germany
[16] Tech Univ Dresden, Fac Med, Dresden, Germany
[17] Tech Univ Dresden, Univ Hosp Carl Gustav Carus, Dresden, Germany
[18] Helmholtz Zentrum Dresden Rossendorf, Helmholtz Assoc, Dresden, Germany
[19] Heidelberg Univ, Fac Med, Heidelberg, Germany
关键词
PET; MRI; RT treatment planning; PET attenuation correction; RF coil holders; RT table overlay; whole-body PET; RADIATION-THERAPY; PET/CT; MRI; IMAGES; COILS; BONE;
D O I
10.1088/1361-6560/abb7c3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective.To implement computed tomography (CT)-based attenuation maps of radiotherapy (RT) positioning hardware and radiofrequency (RF) coils to enable hybrid positron emission tomography/magnetic resonance imaging (PET/MRI)-based RT treatment planning. Materials and Methods. The RT positioning hardware consisted of a flat RT table overlay, coil holders for abdominal scans, coil holders for head and neck scans and an MRI compatible hip and leg immobilization device. CT images of each hardware element were acquired on a CT scanner. Based on the CT images, attenuation maps of the devices were created. Validation measurements were performed on a PET/MR scanner using a Ge-68 phantom (48 MBq, 10 min scan time). Scans with each device in treatment position were performed. Then, reference scans containing only the phantom were taken. The scans were reconstructed online (at the PET/MRI scanner) and offline (via e7tools on a PC) using identical reconstruction parameters. Average reconstructed activity concentrations of the device and reference scans were compared. Results. The device attenuation maps were successfully implemented. The RT positioning devices caused an average decrease of reconstructed PET activity concentration in the range between -8.3 +/- 2.1% (mean +/- SD) (head and neck coil holder with coils) to -1.0 +/- 0.5% (abdominal coil holder). With attenuation correction taking into account RT hardware, these values were reduced to -2.0 +/- 1.2% and -0.6 +/- 0.5%, respectively. The results of the offline and online reconstructions were nearly identical, with a difference of up to 0.2%. Conclusion. The decrease in reconstructed activity concentration caused by the RT positioning devices is clinically relevant and can successfully be corrected using CT-based attenuation maps. Both the offline and online reconstruction methods are viable options.
引用
收藏
页数:10
相关论文
共 24 条
[1]   Multi-modality functional image guided dose escalation in the presence of uncertainties [J].
Alber, Markus ;
Thorwarth, Daniela .
RADIOTHERAPY AND ONCOLOGY, 2014, 111 (03) :354-359
[2]   [68Ga-]PSMA-11 PET/CT and multiparametric MRI for gross tumor volume delineation in a slice by slice analysis with whole mount histopathology as a reference standard - Implications for focal radiotherapy planning in primary prostate cancer [J].
Bettermann, Alisa S. ;
Zamboglou, Constantinos ;
Kiefer, Selina ;
Jilg, Cordula A. ;
Spohn, Simon ;
Kranz-Rudolph, Jasmin ;
Fassbender, Thomas F. ;
Bronsert, Peter ;
Nicolay, Nils H. ;
Gratzke, Christian ;
Bock, Michael ;
Ruf, Juri ;
Benndorf, Matthias ;
Grosu, Anca L. .
RADIOTHERAPY AND ONCOLOGY, 2019, 141 :214-219
[3]   Method for transforming CT images for attenuation correction in PET/CT imaging [J].
Carney, JPJ ;
Townsend, DW ;
Rappoport, V ;
Bendriem, B .
MEDICAL PHYSICS, 2006, 33 (04) :976-983
[4]  
Catana C, 2006, J NUCL MED, V47, P1968
[5]  
Delso G, 2011, J NUCL MED, V52, P1914, DOI 10.2967/jnumed.111.092726
[6]   Markerless attenuation correction for carotid MRI surface receiver coils in combined PET/MR imaging [J].
Eldib, Mootaz ;
Bini, Jason ;
Robson, Philip M. ;
Calcagno, Claudia ;
Faul, David D. ;
Tsoumpas, Charalampos ;
Fayad, Zahi A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (12) :4705-4717
[7]   Comparison of hybrid 68Ga-PSMA PET/MRI and 68Ga-PSMA PET/CT in the evaluation of lymph node and bone metastases of prostate cancer [J].
Freitag, Martin T. ;
Radtke, Jan P. ;
Hadaschik, Boris A. ;
Kopp-Schneider, A. ;
Eder, Matthias ;
Kopka, Klaus ;
Haberkorn, Uwe ;
Roethke, Matthias ;
Schlemmer, Heinz-Peter ;
Afshar-Oromieh, Ali .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 (01) :70-83
[8]   MLAA-based attenuation correction of flexible hardware components in hybrid PET/MR imaging [J].
Heusser, Thorsten ;
Rank, Christopher M. ;
Berker, Yannick ;
Freitag, Martin T. ;
Kachelriess, Marc .
EJNMMI PHYSICS, 2017, 4
[9]   Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques [J].
Hofmann, Matthias ;
Pichler, Bernd ;
Schoelkopf, Bernhard ;
Beyer, Thomas .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 :93-104
[10]   Integrated PET/MR imaging: Automatic attenuation correction of flexible RF coils [J].
Kartmann, Rene ;
Paulus, Daniel H. ;
Braun, Harald ;
Aklan, Bassim ;
Ziegler, Susanne ;
Navalpakkam, Bharath K. ;
Lentschig, Markus ;
Quick, Harald H. .
MEDICAL PHYSICS, 2013, 40 (08)