Multiple breath washout with a sidestream ultrasonic flow sensor and mass spectrometry: A comparative study

被引:43
作者
Fuchs, Susanne I.
Buess, Christian
Lum, Sooky
Kozowska, Wanda
Stocks, Janet
Gappa, Monika [1 ]
机构
[1] Hannover Med Sch, Dept Paediat Pulmonol & Neonatol, D-30623 Hannover, Germany
[2] NDD Med Technol, Zurich, Switzerland
[3] UCL, Inst Child Hlth, Portex Unit, London, England
关键词
multiple breath washout; validation; mass spectrometer; molar mass; lung clearance index; comparison; ultrasonic flow sensor; child; ventilation inhomogeneity;
D O I
10.1002/ppul.20524
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Over recent years, there has been renewed interest in the multiple breath wash-out (MBW) technique for assessing ventilation inhomogeneity (VI) as a measure of early lung disease in children. While currently considered the gold standard, use of mass spectrometry (MS) to measure MBW is not commercially available, thereby limiting widespread application of this technique. A mainstream ultrasonic flow sensor was marketed for MBW a few years ago, but its use was limited to infants. We have recently undertaken intensive modifications of both hardware and software for the ultrasonic system to extend its use for older children. The aim of the current in vivo study was to compare simultaneous measurements of end-tidal tracer gas concentrations and lung clearance index (LCI) from this modified ultrasonic device with those from a mass spectrometer. Paired measurements of three MBW, using 4% sulfur hexafluoride (SF6) as the tracer gas and the two systems in series, were obtained in nine healthy adult volunteers. End-tidal tracer gas concentrations (n = 675 paired values) demonstrated close agreement (95% Cl of difference -0.23; -0.17%, r(2) = 1). FRC was slightly higher from the MS (95%CI 0.08;0.17 L), but there was no difference in LCI (95%Cl -0.10; 0.3). We conclude, that this ultrasonic prototype system measures end-tidal tracer gas concentration accurately and may therefore be a valid tool for MBW beyond early childhood. This prototype system could be the basis for a commercial device allowing more widespread application of MBW in the near future.
引用
收藏
页码:1218 / 1225
页数:8
相关论文
共 29 条
[1]   Gas mixing efficiency from birth to adulthood measured by multiple-breath washout [J].
Aurora, P ;
Kozlowska, W ;
Stocks, J .
RESPIRATORY PHYSIOLOGY & NEUROBIOLOGY, 2005, 148 (1-2) :125-139
[2]   Multiple breath inert gas washout as a measure of ventilation distribution in children with cystic fibrosis [J].
Aurora, P ;
Gustafsson, P ;
Bush, A ;
Lindblad, A ;
Oliver, C ;
Wallis, CE ;
Stocks, J .
THORAX, 2004, 59 (12) :1068-1073
[3]   Multiple-breath washout as a marker of lung disease in preschool children with cystic fibrosis [J].
Aurora, P ;
Bush, A ;
Gustafsson, P ;
Oliver, C ;
Wallis, C ;
Price, J ;
Stroobant, J ;
Carr, S ;
Stocks, J .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2005, 171 (03) :249-256
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   A PULSED DIAGONAL-BEAM ULTRASONIC AIR-FLOW METER [J].
BUESS, C ;
PIETSCH, P ;
GUGGENBUHL, W ;
KOLLER, EA .
JOURNAL OF APPLIED PHYSIOLOGY, 1986, 61 (03) :1195-1199
[6]   DESIGN AND CONSTRUCTION OF A PULSED ULTRASONIC AIR FLOWMETER [J].
BUESS, C ;
PIETSCH, P ;
GUGGENBUHL, W ;
KOLLER, EA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (08) :768-774
[7]  
FUCHS S, 2003, EUR RESPIR J, V22, pS129
[8]  
FUCHS S, 2004, EUR RESPIR J, V24, pS376
[9]  
FUCHS S, 2006, EUR RESPIR J, V28, pS704
[10]  
FUCHS S, 2003, EUR RESPIR J, V22, pS383