The effect of airflow on site-specific electrical conductance measurements used in the diagnosis of pit and fissure caries in vitro

被引:12
作者
Ricketts, DNJ
Kidd, EAM
Wilson, RF
机构
[1] Guy’s Hospital Dental School United Medical, Dental Schools of Guy’s, St. Thomas’s Hospitals, London
关键词
diagnosis; electrical resistance; occlusal caries;
D O I
10.1159/000262385
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
There has been renewed interest in the electronic diagnosis of occlusal caries using measurement of conductance or impedance. One of two previously manufactured electronic caries detectors (the Vanguard electronic caries detector, Massachusetts Manufacturing Corporation, Cambridge, Mass., USA) had a probe tip with an integral air supply. Airflow is essential for removing superficial moisture and preventing surface conduction to the gingival margin. The aim of this study was to determine the effect of airflow on electronic diagnosis of occlusal caries using a prototype electronic caries meter (ECM II. LODE, Groningen, The Netherlands) fitted with a flow meter. Stable conductance readings were taken at 76 discrete sites on 32 extracted teeth with no visible signs of cavitation, at 3 airflows: 5, 7.5 and 10 litres/min. The stable conductance scale was a continuous scale from -0.45 to 13.25 and set by the manufacturer. Histological validation was carried out on macroradiographs of sections cut to include each sample site. The histological picture was compared with the stable conductance readings taken at various airflows. Sensitivity and specificity were calculated using different conductance readings to differentiate sound and carious sites, and receiver operating characteristic (ROC) curves constructed. Of the sites, 32% had enamel and dentine caries and 33% had enamel caries. The ROC curves showed airflow to be highly relevant. An airflow of 5 litres/min was shown to be inadequate and led to large numbers of false-positive diagnoses. A minimum airflow of 7.5 litres/min was required to achieve optimum sensitivity (92%) and specificity (87%) for dentine caries diagnosis.
引用
收藏
页码:111 / 118
页数:8
相关论文
共 16 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   ESTIMATING THE RELIABILITY OF CONTINUOUS MEASURES WITH CRONBACH ALPHA OR THE INTRACLASS CORRELATION-COEFFICIENT - TOWARD THE INTEGRATION OF 2 TRADITIONS [J].
BRAVO, G ;
POTVIN, L .
JOURNAL OF CLINICAL EPIDEMIOLOGY, 1991, 44 (4-5) :381-390
[3]  
CAMPBELL MJ, 1990, MED STAT COMMON SENS, P36
[4]   SURFACE-AREA AND PORE-SIZE ANALYSIS FOR HUMAN-ENAMEL AND DENTIN BY WATER-VAPOR SORPTION [J].
DIBDIN, GH ;
POOLE, DFG .
ARCHIVES OF ORAL BIOLOGY, 1982, 27 (03) :235-241
[5]  
FLAITZ C M, 1986, Pediatric Dentistry, V8, P24
[6]   THE MEANING AND USE OF THE AREA UNDER A RECEIVER OPERATING CHARACTERISTIC (ROC) CURVE [J].
HANLEY, JA ;
MCNEIL, BJ .
RADIOLOGY, 1982, 143 (01) :29-36
[7]   ELECTRICAL CONDUCTANCE AND ELECTRODE AREA ON SOUND SMOOTH ENAMEL IN EXTRACTED TEETH [J].
HUYSMANS, MCDNJM ;
VERDONSCHOT, EH ;
RONDEL, P .
CARIES RESEARCH, 1995, 29 (02) :88-93
[8]   OCCLUSAL CARIES DIAGNOSIS - A CHANGING CHALLENGE FOR CLINICIAN AND EPIDEMIOLOGISTS [J].
KIDD, EAM ;
RICKETTS, DNJ ;
PITTS, NB .
JOURNAL OF DENTISTRY, 1993, 21 (06) :323-331
[9]   IN-VIVO DIAGNOSIS OF FISSURE CARIES USING A NEW ELECTRICAL-RESISTANCE MONITOR [J].
LUSSI, A ;
FIRESTONE, A ;
SCHOENBERG, V ;
HOTZ, P ;
STICH, H .
CARIES RESEARCH, 1995, 29 (02) :81-87
[10]   PORE STRUCTURE OF HUMAN DENTAL ENAMEL [J].
MORENO, EC ;
ZAHRADNIK, RT .