Factors affecting the stimulus artifact tail in surface-recorded somatosensory-evoked potentials

被引:6
作者
Hua, Y. [1 ]
Lovely, D. F. [1 ]
Doraiswami, R. [1 ]
机构
[1] Univ New Brunswick, Dept Elect & Comp Engn, Fredericton, NB E3B 5A3, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
evoked potentials; stimulus artifact; tissue modeling; instrumentation; time series analysis;
D O I
10.1007/s11517-006-0034-4
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Surface-recorded somatosensory-evoked potentials (SEPs) are neural signals elicited by an external stimulus. In the case of electrically induced SEPs, the artifact generated by the stimulation process can severely distort the signal. In some cases, the artifact tail often lasts well into the initiation of the SEP making the determination of absolute latency very difficult. In this work, a new approach was taken to identify factors that affect the tail of the artifact. The methodology adopted was the development of a lumped electrical circuit model of the artifact generation process. While the modeling of the instrumentation hardware is relatively simple, this is not the case with tissue and electrode/skin interface effects. Consequently, this paper describes a novel tissue modeling approach that uses an autoregressive moving average (ARMA) parametric technique and an artificial neural network ( ANN) to estimate tissue parameters from experimental data. This coupled with an estimation of the stimulation electrode-skin impedance completes the lumped circuit model. Simulink (R) ( The Mathworks Inc.) was used to evaluate the model under several different conditions. These results show that both the stimulation electrode-skin interface impedance and nature of the body tissue directly under the recording electrodes have a profound effect on the appearance of the stimulus artifact tail. This was verified by experimentally recorded data obtained from the median nerve using surface electrodes. Conclusions drawn from this work include that stimulation electrodes with low series capacitance should be used whenever possible to minimize the duration of the artifact tail.
引用
收藏
页码:226 / 241
页数:16
相关论文
共 25 条
[1]   LOW NOISE ALTERNATING CURRENT AMPLIFIER AND COMPENSATOR TO REDUCE STIMULUS ARTEFACT [J].
ANDERSEN, VO ;
BUCHTHAL, F .
MEDICAL & BIOLOGICAL ENGINEERING, 1970, 8 (05) :501-&
[2]  
BABB TL, 1978, ELECTROEN CLIN NEURO, V44, P528, DOI 10.1016/0013-4694(78)90038-X
[3]  
BROWN RH, 1981, 34 IEEE ANN C ENG ME, P308
[4]  
De Luca G, 1997, CANADIAN MEDICAL AND BIOLOGICAL ENGINEERING SOCIETY, PROCEEDINGS, P28
[5]   HYBRID STIMULATOR FOR CHRONIC EXPERIMENTS [J].
DELPOZO, F ;
DELGADO, JMR .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1978, 25 (01) :92-94
[6]   PERFORMANCE MONITORING AND FAULT PREDICTION USING A LINEAR PREDICTIVE CODING ALGORITHM [J].
DORAISWAMI, R .
AUTOMATICA, 1993, 29 (04) :1115-1120
[7]  
FREEMAN JA, 1970, CLIN NEUROPHYSIOL, V31, P170
[8]   DIFFERENTIATION OF PSEUDO-INVERSES AND NONLINEAR LEAST-SQUARES PROBLEMS WHOSE VARIABLES SEPARATE [J].
GOLUB, GH ;
PEREYRA, V .
SIAM JOURNAL ON NUMERICAL ANALYSIS, 1973, 10 (02) :413-432
[9]  
GRIEVE R, 1995, ANN INT C IEEE ENG M, V17, P801
[10]   Influence of recording instrumentation on the stimulus artifact tail in the surface acquisition of somatosensory evoked potentials [J].
Hamming, N. ;
Lovely, D. F. .
MEDICAL ENGINEERING & PHYSICS, 2007, 29 (01) :148-153