Photothermal spectroscopy using multilayer cantilever for chemical detection

被引:24
作者
Li, GM
Burggraf, LW
Baker, WP
机构
[1] USAF, Inst Technol, Dept Engn Phys, Wright Patterson AFB, OH 45344 USA
[2] USAF, Inst Technol, Dept Math & Stat, Wright Patterson AFB, OH 45344 USA
关键词
D O I
10.1063/1.125988
中图分类号
O59 [应用物理学];
学科分类号
摘要
Photothermal spectroscopy is demonstrated using a high-aspect-ratio multilayer cantilever to measure adsorbed dimethyl methylphosphonate, which has an optical absorbance of similar to 10(-5) in the near infrared range. Detection sensitivity was 160 pW for a reed of 6 mm in length, 2 mm in width, and 10 mu m. Sensitivity of the cantilever is compared to a thermal diffusion model that accounts for conduction loss through the cantilever clamp and to air along the length of the cantilever surface. [S0003-6951(00)02309-3].
引用
收藏
页码:1122 / 1124
页数:3
相关论文
共 6 条
[1]   PHOTOTHERMAL SPECTROSCOPY WITH FEMTOJOULE SENSITIVITY USING A MICROMECHANICAL DEVICE [J].
BARNES, JR ;
STEPHENSON, RJ ;
WELLAND, ME ;
GERBER, C ;
GIMZEWSKI, JK .
NATURE, 1994, 372 (6501) :79-81
[2]   A FEMTOJOULE CALORIMETER USING MICROMECHANICAL SENSORS [J].
BARNES, JR ;
STEPHENSON, RJ ;
WOODBURN, CN ;
OSHEA, SJ ;
WELLAND, ME ;
RAYMENT, T ;
GIMZEWSKI, JK ;
GERBER, C .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1994, 65 (12) :3793-3798
[3]   Force-balancing microforce sensor with an optical-fiber interferometer [J].
Kato, N ;
Suzuki, I ;
Kikuta, H ;
Iwata, K .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (06) :2475-2478
[4]   Optimization and performance of high-resolution micro-optomechanical thermal sensors [J].
Lai, J ;
Perazzo, T ;
Shi, Z ;
Majumdar, A .
SENSORS AND ACTUATORS A-PHYSICAL, 1997, 58 (02) :113-119
[5]  
ROSENCWAIG A, 1980, PHOTOACOUSTICS PHOTO, P93
[6]   FORCE MICROSCOPE USING A FIBER-OPTIC DISPLACEMENT SENSOR [J].
RUGAR, D ;
MAMIN, HJ ;
ERLANDSSON, R ;
STERN, JE ;
TERRIS, BD .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (11) :2337-2340