Can current minute ventilation rate adaptive pacemakers provide appropriate chronotropic response in pediatric patients?

被引:3
作者
Cabrera, ME
Portzline, G
Aach, S
Condie, C
Dorostkar, P
Mianulli, M
机构
[1] Case Western Reserve Univ, Cleveland, OH 44106 USA
[2] Rainbow Babies & Childrens Hosp, Cleveland, OH 44106 USA
[3] Medtronic Inc, Minneapolis, MN USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2002年 / 25卷 / 06期
关键词
minute ventilation; rate responsive; children; pacemaker; dual sensor; algorithm;
D O I
10.1046/j.1460-9592.2002.00907.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Since children have different activity patterns and exercise responses, uncertainty exists as to whether minute ventilation (MV) sensors designed for adults provide adequate chronotropic response in pediatrics. In particular, high respiratory rates (RR > 48 breaths/min), which are characteristic of the ventilatory response to exercise in children, cannot be sensed by MV rate responsive pacemakers. The purpose of this study was to evaluate the MV sensor rate response of the Medtronic Kappa 400 using exercise data from healthy children in a computer simulation of its rate response algorithm. Thirty-eight healthy children, ages 6-14, underwent a treadmill maximal exercise test. Subjects were divided based on body surface area (BSA) and W rate response parameters were selected. Respiratory rates and tidal volumes were entered into the Kappa 400 rate response algorithm to calculate sensor-driven rates. Intrinsic heart rate (HR), oxygen uptake, and sensor-driven rates were normalized to HR reserve (HRR), metabolic reserve (MR), and sensor-driven reserve to compare across groups, Linear regression analysis among sensor-driven rate reserve, HRR, and MR was performed as described by Wilkoff. The mean slopes (+/- SD) of the relationships between the sensor-driven rate reserve and HRR were 1.06 +/- 0,34, 1.07 +/- 0.28, and 1.01 +/- 0.19 for children with BSA < 1.10 m(2), 1.10 < BSA < 1.40 m(2), and BSA > 1.40 m(2), respectively. High correlations were found between sensor-drive rates and HR responses and between sensor-drive rates and MV throughout exercise. No significant differences were noted between sensor-drive rates and HR using the Wilkoff model. From this study the authors conclude that: (1) W is a good physiological parameter to control heart rate and (2) simulated sensor-driven rates closely match intrinsic HRs during exercise in healthy children, which supports the appropriateness of clinical validation in pediatric pacemaker patients.
引用
收藏
页码:907 / 914
页数:8
相关论文
共 26 条
[1]   CONTROL OF PACEMAKER RATE BY IMPEDANCE-BASED RESPIRATORY MINUTE VENTILATION [J].
ALT, E ;
HEINZ, M ;
HIRGSTETTER, C ;
EMSLANDER, HP ;
DAUM, S ;
BLOMER, H .
CHEST, 1987, 92 (02) :247-252
[2]  
ALT E, 1993, RATE ADAPTIVE CARDIA, P58
[3]   THE LEVEL AND TEMPO OF CHILDRENS PHYSICAL ACTIVITIES - AN OBSERVATIONAL STUDY [J].
BAILEY, RC ;
OLSON, J ;
PEPPER, SL ;
PORSZASZ, J ;
BARSTOW, TJ ;
COOPER, DM .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 1995, 27 (07) :1033-1041
[4]  
Bar-Or O., 1983, PEDIAT SPORTS MED PR
[5]  
CAMM AJ, 1999, FDN CARDIAC PACING I, P257
[6]   Dual sensor pacemakers in children: What is the choice of sensor blending? [J].
Celiker, A ;
Tokel, K ;
Lenk, MK ;
Ozme, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (05) :1301-1304
[7]   Comparison of normal sinus rhythm and pacing rate in children with minute ventilation single chamber rate adaptive permanent pacemakers [J].
Celiker, A ;
Ceviz, N ;
Alehan, D ;
Lenk, MK ;
Özme, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (11) :2100-2104
[8]  
Celiker A, 1996, EUR HEART J, V17, P1251
[9]  
COOPER DM, 1996, CHILD ADOLESCENT ATH, P54
[10]  
EARANOWSKI TP, 1987, J DEV BEHAV PEDIATR, V8, P203