High incidence of defective ultrasound transducers in use in routine clinical practice

被引:48
作者
Martensson, Mattias [1 ]
Olsson, Mats [2 ]
Segall, Bjorn [5 ]
Fraser, Alan G. [4 ]
Winter, Reidar [3 ]
Brodin, Lars-Ake [1 ]
机构
[1] KTH, Sch Tech & Hlth, S-14152 Huddinge, Sweden
[2] Karolinska Univ Hosp, Dept Biomed Engn, Huddinge, Sweden
[3] Karolinska Univ Hosp, Dept Clin Physiol, Huddinge, Sweden
[4] Cardiff Univ, Sch Med, Wales Heart Res Inst, Cardiff, S Glam, Wales
[5] BBS Med AB, Vato, Sweden
来源
EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY | 2009年 / 10卷 / 03期
关键词
Ultrasound; Transducer; Probe; Test; Defect; Error; QUALITY-CONTROL; MEDICAL ULTRASOUND; PERFORMANCE; SYSTEM;
D O I
10.1093/ejechocard/jen295
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The objective was to evaluate the function of ultrasound transducers in use in routine clinical practice and thereby estimating the incidence of defective transducers. The study comprised a one-time test of 676 transducers from 7 manufacturers which were in daily use in clinical departments at 32 hospitals. They were tested with the Sonora FirstCall Test System; 39.8% exhibited a transducer error. Delamination was detected in 26.5% and break in the cable was detected in 8.4% of the tested transducers. Errors originating from the piezoelectrical elements were unusual. Delamination and short circuit occurred without significant differences between transducers from all tested manufacturers, but the errors break in the cable, weak and dead element showed a statistically significant higher frequency in transducers from certain manufacturers. The high error frequency and the risk for incorrect medical decisions when using a defective transducer indicate an urgent need for increased testing of the transducers in clinical departments.
引用
收藏
页码:389 / 394
页数:6
相关论文
共 20 条
  • [1] *AIUM, 1990, TECHN STAND COMM MET
  • [2] *AIUM, 1995, TECHN STAND COMM M 2
  • [3] *AIUM, 1995, TECHN STAND COMM QUA
  • [4] *AIUM, 1974, AIUM 100 MM TEST OBJ
  • [5] BRENDEL K, 1976, ULTRASOUND MED BIOL, V2, P343
  • [6] CARSON P L, 1976, Journal of Clinical Ultrasound, V4, P259, DOI 10.1002/jcu.1870040406
  • [7] DUDLEY NJ, 2000, EUR J ULTRASOUND, V12, P233
  • [8] A computerised quality control testing system for B-mode ultrasound
    Gibson, NM
    Dudley, NJ
    Griffith, K
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (12) : 1697 - 1711
  • [9] Real-time B-mode ultrasound quality control test procedures - Report of AAPM ultrasound task group no. 1
    Goodsitt, MM
    Carson, PL
    Witt, S
    Hykes, DL
    Kofler, JM
    [J]. MEDICAL PHYSICS, 1998, 25 (08) : 1385 - 1406
  • [10] METHOD OF CALIBRATING CONTACT B-SCANNERS
    HALL, AJ
    FLEMING, JEE
    [J]. ULTRASONICS, 1978, 16 (06) : 277 - 281