Preliminary system characterization of flat panel detector-based cone beam CT for breast imaging

被引:26
作者
Ning, RL [1 ]
Yu, Y [1 ]
Conover, D [1 ]
Lu, XH [1 ]
He, HG [1 ]
Chen, ZK [1 ]
Schiffhauer, L [1 ]
Cullinan, J [1 ]
机构
[1] Univ Rochester, Med Ctr, Dept Radiol, Rochester, NY 14642 USA
来源
MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 2004年 / 5368卷
关键词
cone beam reconstruction; cone beam volume CT; digital mammography; mammography; flat panel detector; breast imaging;
D O I
10.1117/12.536231
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Conventional film-screen mammography is the most effective tool for the early detection of breast cancer currently available. However, conventional mammography has relatively low sensitivity to detect small breast cancers (under several millimeters) owing to an overlap in the appearances of benign and malignant lesions, and surrounding structure. The limitations accompanying conventional mammography is to be addressed by incorporating a cone beam CT imaging technique with a recently developed flat panel detector. Computer simulation and preliminary studies have been performed to prove the feasibility of developing a flat panel detector-based cone beam CT breast imaging (FPD-CBCTBI) technique. A preliminary system characterization study of flat panel detector-based cone beam CT for breast imaging was performed to confirm the findings in the computer simulation and previous phantom studies using the current prototype cone beam CT scanner. The results indicate that the CBCTBI technique effectively removes structure overlap and significantly improves the detectability of small breast tumors. More importantly, the results also demonstrate CBCTBI offers good image quality with the radiation dose level less than or equal to that of conventional mammography. The results from this study suggest that FPD-CBCTBI is a potentially powerful breast-imaging tool.
引用
收藏
页码:292 / 303
页数:12
相关论文
共 33 条
[1]  
Boone JM, 2000, RADIOLOGY, V217, P346
[2]  
CHANG CHJ, 1978, RADIOLOGY, V129, P209
[3]   Cone-beam volume CT breast imaging: wavelet analysis-based multi-resolution reconstruction and de-noising technique [J].
Chen, B ;
Ning, R .
MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 :236-244
[4]   Cone-beam volume CT breast imaging: Feasibility study [J].
Chen, B ;
Ning, R .
MEDICAL PHYSICS, 2002, 29 (05) :755-770
[5]   Cone-beam volume CT mammographic imaging: feasibility study [J].
Chen, BA ;
Ning, RL .
MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 :655-664
[6]   Accurate perspective projection calculation using a pixel-pyramid model for iterative cone beam reconstruction [J].
Chen, ZK ;
Ning, R ;
Conover, D .
MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 :728-739
[7]   Why should breast tumour detection go three dimensional? [J].
Chen, ZK ;
Ning, R .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (14) :2217-2228
[8]  
CONOVER D, 2003, P SOC PHOTO-OPT INS, V5030, P768
[9]  
FEIG SA, 1999, RISK BENEFIT CONTROV, P99
[10]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619