Ion-selective carbon paste electrode based on new tripodal ligand for determination of cadmium (II)

被引:22
作者
Abbastabar-Ahangar, Hossein [1 ,2 ]
Shirzadmehr, Ali [1 ]
Marjani, Katayoun [1 ]
Khoshsafar, Hamid [1 ]
Chaloosi, Marzieh [1 ]
Mohammadi, Leila [1 ]
机构
[1] Tarbiat Moallem Univ, Fac Chem, Tehran, Iran
[2] Young Researcher Club Iran, Tehran, Iran
关键词
Cd(II); Carbon paste modified electrode; Potentiometry; Tripodal ligand; ATOMIC-ABSORPTION-SPECTROMETRY; MOLECULAR RECOGNITION; BIOLOGICAL SAMPLES; SENSORS; PB; CD;
D O I
10.1007/s10847-008-9519-0
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A new chemically modified carbon paste electrode for Cd(II) ions based on 3,5-dinitro-N-(tri-2-pyridyl methyl) benzamide (DNTPMBA) as an ionophore was prepared. The electrode exhibits a Nernstian response for Cd(II) ions over a wide concentration range (2.16 x 10(-7)-1.00 x 10(-1) mol L-1) with a slope of 30 +/- A 1 mV per decade. It has a response time of about 50 s and can be used for a period of 3 months with good reproducibility. Detection limit obtained in the optimal conditions was 1.82 x 10(-7) mol L-1. The electrode was successfully used for potentiometric determination of Cd(II) in well water. The pH influence and interference of some cations were also studied.
引用
收藏
页码:287 / 293
页数:7
相关论文
共 32 条
[1]   New triiodomercurate-modified carbon paste electrode for the potentiometric determination of mercury [J].
Abbas, MN ;
Mostafa, GAE .
ANALYTICA CHIMICA ACTA, 2003, 478 (02) :329-335
[2]   Influence of remote intramolecular hydrogen bonds on the thermodynamics of molecular recognition of cis-1,3,5-cyclohexanetricarboxylic acid. [J].
Ballester, P ;
Costa, A ;
Deyà, PM ;
Vega, M ;
Morey, J ;
Deslongchamps, G .
TETRAHEDRON LETTERS, 1999, 40 (01) :171-174
[3]   RECOMMENDATIONS FOR NOMENCLATURE OF ION-SELECTIVE ELECTRODES - (IUPAC RECOMMENDATIONS 1994) [J].
BUCK, RP ;
LINDNER, E .
PURE AND APPLIED CHEMISTRY, 1994, 66 (12) :2527-2536
[4]  
Fan A., 2002, J SUPRAMOECULAR CHEM, V2, P247, DOI DOI 10.1016/S1472-7862(03)00079-0
[5]  
FRIBERG L, 1986, HDB TOXICOLOGY METAL, P130
[6]  
FRIES J, 1977, ORGANIC REAGENTS TRA, P227
[7]   Copper(II) ion-selective electrodes based on dithiosalicylic and thiosalicylic acids [J].
Gismera, MJ ;
Procopio, JR ;
Sevilla, MT ;
Hernández, L .
ELECTROANALYSIS, 2003, 15 (02) :126-132
[8]   Electroanalytical studies on cadmium(II) selective potentiometric sensors based on t-butyl thiacalix[4]arene and thiacalix[4]arene in poly(vinyl chloride) [J].
Gupta, Vinod K. ;
Jain, Ajay K. ;
Ludwig, Rainer ;
Maheshwari, Gaurav .
ELECTROCHIMICA ACTA, 2008, 53 (05) :2362-2368
[9]   Cadmium(II)-selective sensors based on dibenzo-24-crown-8 in PVC matrix [J].
Gupta, VK ;
Kumar, P .
ANALYTICA CHIMICA ACTA, 1999, 389 (1-3) :205-212
[10]   Determination of cadmium in urine specimens by graphite furnace atomic absorption spectrometry using a fast atomization program [J].
Hernández-Caraballo, EA ;
Burguera, M ;
Burguera, JL .
TALANTA, 2004, 63 (02) :419-424