High dose-per-pulse electron beam dosimetry: Usability and dose-rate independence of EBT3 Gafchromic films

被引:151
作者
Jaccard, Maud [1 ]
Petersson, Kristoffer [1 ]
Buchillier, Thierry [1 ]
Germond, Jean-Francois [1 ]
Duran, Maria Teresa [1 ]
Vozenin, Marie-Catherine [2 ,3 ]
Bourhis, Jean [2 ,3 ]
Bochud, Francois O. [1 ]
Bailat, Claude [1 ]
机构
[1] Univ Lausanne Hosp, Inst Radiat Phys, Lausanne, Switzerland
[2] Univ Lausanne Hosp, Dept Radiat Oncol, Lausanne, Switzerland
[3] Lausanne Univ Hosp Lausanne, Radio Oncol Lab, DO CHUV, Lausanne, Switzerland
基金
瑞士国家科学基金会;
关键词
dose-rate dependence; EBT3 Gafchromic films; energy dependence; high dose-perpulse; high dose-rate; ENERGY-DEPENDENCE; RECOMBINATION CORRECTION; PHOTON;
D O I
10.1002/mp.12066
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The aim of this study was to assess the suitability of Gafchromic EBT3 films for reference dose measurements in the beam of a prototype high dose-per-pulse linear accelerator (linac), capable of delivering electron beams with a mean dose-rate (D-m) ranging from 0.07 to 3000 Gy/s and a doserate in pulse (D-p) of up to 8 x 10(6) Gy/s. To do this, we evaluated the overall uncertainties in EBT3 film dosimetry as well as the energy and dose-rate dependence of their response. Material and methods: Our dosimetric system was composed of EBT3 Gafchromic films in combination with a flatbed scanner and was calibrated against an ionization chamber traceable to primary standard. All sources of uncertainties in EBT3 dosimetry were carefully analyzed using irradiations at a clinical radiotherapy linac. Energy dependence was investigated with the same machine by acquiring and comparing calibration curves for three different beam energies (4, 8 and 12 MeV), for doses between 0.25 and 30 Gy. D-m dependence was studied at the clinical linac by changing the pulse repetition frequency (f) of the beam in order to vary D-m between 0.55 and 4.40 Gy/min, while D-p dependence was probed at the prototype machine for D-p ranging from 7 x 10(3) to 8 x 10(6) Gy/s. Dp dependence was first determined by studying the correlation between the dose measured by films and the charge of electrons measured at the exit of the machine by an induction torus. Furthermore, we compared doses from the films to independently calibrated thermo-luminescent dosimeters (TLD) that have been reported as being dose-rate independent up to such high dose-rates. Results: We report that uncertainty below 4% (k = 2) can be achieved in the dose range between 3 and 17 Gy. Results also demonstrated that EBT3 films did not display any detectable energy dependence for electron beam energies between 4 and 12 MeV. No D-m dependence was found either. In addition, we obtained excellent consistency between films and TLDs over the entire D-p range attainable at the prototype linac confirming the absence of any dose-rate dependence within the investigated range (7 x 10(3) to 8 x 10(6) Gy/s). This aspect was further corroborated by the linear relationship between the dose-perpulse (D-p) measured by films and the charge per pulse (C-p) measured at the prototype linac exit. Conclusion: Our study shows that the use of EBT3 Gafchromic films can be extended to reference dosimetry in pulsed electron beams with a very high dose rate. The measurement results are associated with an overall uncertainty below 4% (k = 2) and are dose-rate and energy independent. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:725 / 735
页数:11
相关论文
共 33 条
[1]  
Andreo P, 2000, 398 TRS IAEA
[2]   Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies [J].
Arjomandy, Bijan ;
Tailor, Ramesh ;
Anand, Aman ;
Sahoo, Narayan ;
Gillin, Michael ;
Prado, Karl ;
Vicic, Milos .
MEDICAL PHYSICS, 2010, 37 (05) :1942-1947
[3]  
Ashland, 2011, GAFCHR EBT3 SCAN HAN
[4]   Treatment planning for radiotherapy with very high-energy electron beams and comparison of VHEE and VMAT plans [J].
Bazalova-Carter, Magdalena ;
Qu, Bradley ;
Palma, Bianey ;
Hardemark, Bjorn ;
Hynning, Elin ;
Jensen, Christopher ;
Maxim, Peter G. ;
Loo, Billy W., Jr. .
MEDICAL PHYSICS, 2015, 42 (05) :2615-2625
[5]   Comparison of film measurements and Monte Carlo simulations of dose delivered with very high-energy electron beams in a polystyrene phantom [J].
Bazalova-Carter, Magdalena ;
Liu, Michael ;
Palma, Bianey ;
Dunning, Michael ;
McCormick, Doug ;
Hemsing, Erik ;
Nelson, Janice ;
Jobe, Keith ;
Colby, Eric ;
Koong, Albert C. ;
Tantawi, Sami ;
Dolgashev, Valery ;
Maxim, Peter G. ;
Loo, Billy W., Jr. .
MEDICAL PHYSICS, 2015, 42 (04) :1606-1613
[6]   The effect of free-electron collection on the recombination correction to ionization measurements of pulsed radiation [J].
Boag, JW ;
Hochhauser, E ;
Balk, OA .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (05) :885-897
[7]   Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification [J].
Borca, Valeria Casanova ;
Pasquino, Massimo ;
Russo, Giuliana ;
Grosso, Pierangelo ;
Cante, Domenico ;
Sciacero, Piera ;
Girelli, Giuseppe ;
La Porta, Maria Rosa ;
Tofani, Santi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02) :158-171
[8]   On the characterization and uncertainty analysis of radiochromic film dosimetry [J].
Bouchard, Hugo ;
Lacroix, Frederic ;
Beaudoin, Gilles ;
Carrier, Jean-Francois ;
Kawrakow, Iwan .
MEDICAL PHYSICS, 2009, 36 (06) :1931-1946
[9]   Ion recombination corrections for the NACP parallel-plate chamber in a pulsed electron beam [J].
Burns, DT ;
McEwen, MR .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (08) :2033-2045
[10]   Precise radiochromic film dosimetry using a flat-bed document scanner [J].
Devic, S ;
Seuntjens, J ;
Sham, E ;
Podgorsak, EB ;
Schmidtlein, CR ;
Kirov, AS ;
Soares, CG .
MEDICAL PHYSICS, 2005, 32 (07) :2245-2253