Evaluation of MLACF based calculated attenuation brain PET imaging for FDG patient studies

被引:12
作者
Bal, Harshali [1 ]
Panin, Vladimir Y. [1 ]
Platsch, Guenther [1 ]
Defrise, Michel [2 ]
Hayden, Charles [1 ]
Hutton, Chloe [1 ]
Serrano, Benjamin [3 ]
Paulmier, Benoit [3 ]
Casey, Michael E. [1 ]
机构
[1] Siemens Mol Imaging, Res & Clin Collaborat, Knoxville, TN 37932 USA
[2] Vrije Univ Brussel, Dept Nucl Med, Brussels, Belgium
[3] Ctr Hosp Princesse Grace, Med Phys & Radiat Protect Unit, Monaco, Monaco
关键词
attenuation correction; MLACF; FDG brain PET; OF-FLIGHT PET; POSITRON-EMISSION-TOMOGRAPHY; F-18-FDG PET; RECONSTRUCTION; TUMORS; METABOLISM; PATTERN; MAP;
D O I
10.1088/1361-6560/aa5e99
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Calculating attenuation correction for brain PET imaging rather than using CT presents opportunities for low radiation dose applications such as pediatric imaging and serial scans to monitor disease progression. Our goal is to evaluate the iterative time-of-flight based maximum-likelihood activity and attenuation correction factors estimation (MLACF) method for clinical FDG brain PET imaging. FDG PET/CT brain studies were performed in 57 patients using the Biograph mCT (Siemens) four-ring scanner. The time-of-flight PET sinograms were acquired using the standard clinical protocol consisting of a CT scan followed by 10 min of single-bed PET acquisition. Images were reconstructed using CT-based attenuation correction (CTAC) and used as a gold standard for comparison. Two methods were compared with respect to CTAC: a calculated brain attenuation correction (CBAC) and MLACF based PET reconstruction. Plane-by-plane scaling was performed for MLACF images in order to fix the variable axial scaling observed. The noise structure of the MLACF images was different compared to those obtained using CTAC and the reconstruction required a higher number of iterations to obtain comparable image quality. To analyze the pooled data, each dataset was registered to a standard template and standard regions of interest were extracted. An SUVr analysis of the brain regions of interest showed that CBAC and MLACF were each well correlated with CTAC SUVrs. A plane-by-plane error analysis indicated that there were local differences for both CBAC and MLACF images with respect to CTAC. Mean relative error in the standard regions of interest was less than 5% for both methods and the mean absolute relative errors for both methods were similar (3.4% +/- 3.1% for CBAC and 3.5% +/- 3.1% for MLACF). However, the MLACF method recovered activity adjoining the frontal sinus regions more accurately than CBAC method. The use of plane-by-plane scaling of MLACF images was found to be a crucial step in order to obtain improved activity estimates. Presence of local errors in both MLACF and CBAC based reconstructions would require the use of a normal database for clinical assessment. However, further work is required in order to assess the clinical advantage of MLACF over CBAC based method.
引用
收藏
页码:2542 / 2558
页数:17
相关论文
共 29 条
[1]  
Ahn S, 2012, IEEE NUCL SCI CONF R, P3695
[2]   DETERMINATION OF OBJECT CONTOUR FROM PROJECTIONS FOR ATTENUATION CORRECTION IN CRANIAL POSITRON EMISSION TOMOGRAPHY [J].
BERGSTROM, M ;
LITTON, J ;
ERIKSSON, L ;
BOHM, C ;
BLOMQVIST, G .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1982, 6 (02) :365-372
[3]   Accurate PET/MR Quantification Using Time of Flight MLAA Image Reconstruction [J].
Boellaard, R. ;
Hofman, M. B. M. ;
Hoekstra, O. S. ;
Lammertsma, A. A. .
MOLECULAR IMAGING AND BIOLOGY, 2014, 16 (04) :469-477
[4]   Attenuation Correction Synthesis for Hybrid PET-MR Scanners: Application to Brain Studies [J].
Burgos, Ninon ;
Cardoso, M. Jorge ;
Thielemans, Kris ;
Modat, Marc ;
Pedemonte, Stefano ;
Dickson, John ;
Barnes, Anna ;
Ahmed, Rebekah ;
Mahoney, Colin J. ;
Schott, Jonathan M. ;
Duncan, John S. ;
Atkinson, David ;
Arridge, Simon R. ;
Hutton, Brian F. ;
Ourselin, Sebastien .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2014, 33 (12) :2332-2341
[5]   Toward Implementing an MRI-Based PET Attenuation-Correction Method for Neurologic Studies on the MR-PET Brain Prototype [J].
Catana, Ciprian ;
van der Kouwe, Andre ;
Benner, Thomas ;
Michel, Christian J. ;
Hamm, Michael ;
Fenchel, Matthias ;
Fischl, Bruce ;
Rosen, Bruce ;
Schmand, Matthias ;
Sorensen, A. Gregory .
JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (09) :1431-1438
[6]  
COLEMAN RE, 1991, J NUCL MED, V32, P616
[7]   Transmission-less attenuation correction in time-of-flight PET: analysis of a discrete iterative algorithm [J].
Defrise, Michel ;
Rezaei, Ahmadreza ;
Nuyts, Johan .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (04) :1073-1095
[8]   Time-of-flight PET data determine the attenuation sinogram up to a constant [J].
Defrise, Michel ;
Rezaei, Ahmadreza ;
Nuyts, Johan .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (04) :885-899
[9]   Molecular imaging of pediatric brain tumors: comparison of tumor metabolism using 18F-FDG-PET and MRSI [J].
Hipp, Sean J. ;
Steffen-Smith, Emilie A. ;
Patronas, Nicholas ;
Herscovitch, Peter ;
Solomon, Jeffrey M. ;
Bent, Robyn S. ;
Steinberg, Seth M. ;
Warren, Katherine E. .
JOURNAL OF NEURO-ONCOLOGY, 2012, 109 (03) :521-527
[10]   MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas Registration [J].
Hofmann, Matthias ;
Steinke, Florian ;
Scheel, Verena ;
Charpiat, Guillaume ;
Farquhar, Jason ;
Aschoff, Philip ;
Brady, Michael ;
Schoelkopf, Bernhard ;
Pichler, Bernd J. .
JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (11) :1875-1883