Copper Fixed-Point Measurements for Radiation Thermometry at National Research Council

被引:4
作者
Todd, A. D. W. [1 ]
Woods, D. J. [1 ]
机构
[1] Natl Res Council Canada, Ottawa, ON K1A 0R6, Canada
关键词
Copper fixed-points; Heat pipe; Hybrid fixed points; Radiation thermometry;
D O I
10.1007/s10765-014-1666-5
中图分类号
O414.1 [热力学];
学科分类号
摘要
Due to its high transition temperature relative to other fixed points defined in the International Temperature Scale of 1990 (ITS-90) and its relatively low cost compared to silver and gold, copper is often chosen as the fixed point used to define the ITS-90 above 1235 K at national measurement institutes. Measurement of the copper freezing point can be done in a variety of furnaces. Although there are a large number of copper fixed-point designs, we expect the freezing temperatures to be the same. The difference between realizing different sized fixed points and the use of different furnaces in which to realize them is explored here. A traditional, large aperture fixed-point containing over 600 g of copper is compared to a hybrid-type fixed point containing only 15 g of copper and a commercial fixed point. Three types of furnaces including a heat-pipe furnace, a compact furnace, and a high-temperature blackbody were used to realize the copper freezing point. Between the fixed-point types, only the length of the plateau differed. However, a significant difference was found between the freezing temperatures determined in the different furnaces, and this difference was independent of cell type.
引用
收藏
页码:1366 / 1377
页数:12
相关论文
共 11 条
[1]   Small copper fixed-point cells of the hybrid type to be used in place of normal larger cells [J].
Battuello, M. ;
Girard, F. ;
Florio, M. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2012, 23 (10)
[2]   Ultra-High Temperature Isothermal Furnace Liners (IFLS) for Copper Freeze Point Cells [J].
Dussinger, P. M. ;
Tavener, J. P. .
TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 8, 2013, 1552 :851-856
[3]  
Hartmann J., 2003, VDI BER, V1784, P135
[4]   A New Compact Fixed-Point Blackbody Furnace [J].
Hiraka, K. ;
Yamada, Y. ;
Ishii, J. ;
Oikawa, H. ;
Shimizu, T. ;
Kadoya, S. ;
Kobayashi, T. .
TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 8, 2013, 1552 :300-304
[5]   The structural failure of copper-freezing-point blackbodies for radiation thermometry: analysis and design improvements [J].
Ma, C. K. .
METROLOGIA, 2006, 43 (06) :515-521
[6]  
OGAREV SA, 2004, P TEMPMEKO 2004 9 IN, P569
[7]   THE INTERNATIONAL TEMPERATURE SCALE OF 1990 (ITS-90) [J].
PRESTONTHOMAS, H .
METROLOGIA, 1990, 27 (01) :3-10
[8]  
Sadli M, 2008, ACTA METEOROL SIN, V29, P59
[9]   Metal (carbide)-carbon eutectics for thermometry and radiometry: A review of the first seven years [J].
Woolliams, Emma R. ;
MacHin, Graham ;
Lowe, David H. ;
Winkler, Rainer .
Metrologia, 2006, 43 (06) :R11-R25
[10]  
Yamada Y., 2008, ACTA METEOROL SIN, V29, P46