Characterization of the impact to PET quantification and image quality of an anterior array surface coil for PET/MR imaging

被引:22
作者
Wollenweber, Scott D. [1 ]
Delso, Gaspar [1 ,2 ]
Deller, Timothy [1 ]
Goldhaber, David [1 ]
Huellner, Martin [2 ]
Veit-Haibach, Patrick [2 ]
机构
[1] GE Healthcare, Waukesha, WI 53186 USA
[2] Univ Zurich Hosp, CH-8091 Zurich, Switzerland
关键词
Positron-emission tomography; Magnetic resonance imaging; Attenuation correction; ATTENUATION CORRECTION; TIME;
D O I
10.1007/s10334-013-0388-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Object The aim of this study was to determine the impact to PET quantification, image quality and possible diagnostic impact of an anterior surface array used in a combined PET/MR imaging system. Materials and methods An extended oval phantom and 15 whole-body FDG PET/CT subjects were re-imaged for one bed position following placement of an anterior array coil at a clinically realistic position. The CT scan, used for PET attenuation correction, did not include the coil. Comparison, including liver SUVmean, was performed between the coil present and absent images using two methods of PET reconstruction. Due to the time delay between PET scans, a model was used to account for average physiologic time change of SUV. Results On phantom data, neglecting the coil caused a mean bias of -8.2 % for non-TOF/PSF reconstruction, and -7.3 % with TOF/PSF. On clinical data, the liver SUV neglecting the coil presence fell by -6.1 % (+/- 6.5 %) for non-TOF/PSF reconstruction; respectively-5.2 %(+/- 5.3 %) with TOF/PSF. All FDG-avid features seen with TOF/PSF were also seen with non-TOF/PSF reconstruction. Conclusion Neglecting coil attenuation for this anterior array coil results in a small but significant reduction in liver SUVmean but was not found to change the clinical interpretation of the PET images.
引用
收藏
页码:149 / 159
页数:11
相关论文
共 50 条
  • [31] Hybrid PET- and MR-driven attenuation correction for enhanced 18F-NaF and 18F-FDG quantification in cardiovascular PET/MR imaging
    Karakatsanis, Nicolas A.
    Abgral, Ronan
    Trivieri, Maria Giovanna
    Dweck, Marc R.
    Robson, Philip M.
    Calcagno, Claudia
    Boeykens, Gilles
    Senders, Max L.
    Mulder, Willem J. M.
    Tsoumpas, Charalampos
    Fayad, Zahi A.
    JOURNAL OF NUCLEAR CARDIOLOGY, 2020, 27 (04) : 1126 - 1141
  • [32] Toward simultaneous PET/MR breast imaging: Systematic evaluation and integration of a radiofrequency breast coil
    Aklan, Bassim
    Paulus, Daniel H.
    Wenkel, Evelyn
    Braun, Harald
    Navalpakkam, Bharath K.
    Ziegler, Susanne
    Geppert, Christian
    Sigmund, Eric E.
    Melsaether, Amy
    Quick, Harald H.
    MEDICAL PHYSICS, 2013, 40 (02)
  • [33] Quality control for quantitative multicenter whole-body PET/MR studies: A NEMA image quality phantom study with three current PET/MR systems
    Boellaard, Ronald
    Rausch, Ivo
    Beyer, Thomas
    Delso, Gaspar
    Yaqub, Maqsood
    Quick, Harald H.
    Sattler, Bernhard
    MEDICAL PHYSICS, 2015, 42 (10) : 5961 - 5969
  • [34] Amyloid burden quantification depends on PET and MR image processing methodology
    Kolinger, Guilherme D.
    Garcia, David Vallez
    Willemsen, Antoon T. M.
    Reesink, Fransje E.
    de Jong, Bauke M.
    Dierckx, Rudi A. J. O.
    De Deyn, Peter P.
    Boellaard, Ronald
    PLOS ONE, 2021, 16 (03):
  • [35] Impact of improved attenuation correction featuring a bone atlas and truncation correction on PET quantification in whole-body PET/MR
    Mark Oehmigen
    Maike E. Lindemann
    Marcel Gratz
    Julian Kirchner
    Verena Ruhlmann
    Lale Umutlu
    Jan Ole Blumhagen
    Matthias Fenchel
    Harald H. Quick
    European Journal of Nuclear Medicine and Molecular Imaging, 2018, 45 : 642 - 653
  • [36] Impact of improved attenuation correction featuring a bone atlas and truncation correction on PET quantification in whole-body PET/MR
    Oehmigen, Mark
    Lindemann, Maike E.
    Gratz, Marcel
    Kirchner, Julian
    Ruhlmann, Verena
    Umutlu, Lale
    Blumhagen, Jan Ole
    Fenchel, Matthias
    Quick, Harald H.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 (04) : 642 - 653
  • [37] A Novel J-Shape Antenna Array for Simultaneous MR-PET or MR-SPECT Imaging
    Choi, Chang-Hoon
    Hong, Suk-Min
    Felder, Joerg
    Tellmann, Lutz
    Scheins, Juergen
    Kops, Elena Rota
    Lerche, Christoph
    Shah, N. Jon
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2022, 41 (05) : 1104 - 1113
  • [38] Application of Highly Flexible Adaptive Image Receive Coil for Lung MR Imaging Using Zero TE Sequence: Comparison with Conventional Anterior Array Coil
    Bae, Kyungsoo
    Jeon, Kyung Nyeo
    Hwang, Moon Jung
    Jung, Yunsub
    Lee, Joonsung
    DIAGNOSTICS, 2022, 12 (01)
  • [39] Impact of PET acquisition durations on image quality and lesion detectability in whole-body 68Ga-PSMA PET-MRI
    Noto, Benjamin
    Buether, Florian
    Springe, Katharina Auf Der
    Avramovic, Nemanja
    Heindel, Walter
    Schaefers, Michael
    Allkemper, Thomas
    Stegger, Lars
    EJNMMI RESEARCH, 2017, 7
  • [40] Integrated PET/MR breast cancer imaging: Attenuation correction and implementation of a 16-channel RF coil
    Oehmigen, Mark
    Lindemann, Maike E.
    Lanz, Titus
    Kinner, Sonja
    Quick, Harald H.
    MEDICAL PHYSICS, 2016, 43 (08) : 4808 - 4820