Development of a multi-component fiber-reinforced composite implant for load-sharing conditions

被引:62
作者
Zhao, D. S.
Moritz, N.
Laurila, P. [3 ]
Mattila, R. [2 ]
Lassila, L. V. J. [2 ]
Strandberg, N.
Mantyla, T. [3 ]
Vallittu, P. K. [2 ]
Aro, H. T. [1 ]
机构
[1] Univ Turku, Orthopaed Res Unit, Dept Orthopaed Surg & Traumatol, Turku 20520, Finland
[2] Univ Turku, Inst Dent, Dept Prosthet Dent & Biomat Sci, Turku 20520, Finland
[3] Tampere Univ Technol, Inst Mat Sci, FIN-33101 Tampere, Finland
关键词
Fiber-reinforced composites; Orthopaedic surgery; Implant; Finite element; Mechanical testing; Intramedullary nail; TOTAL HIP-REPLACEMENT; FINITE-ELEMENT; MECHANICAL-PROPERTIES; FLEXURAL PROPERTIES; EARLY FAILURE; CANINE FEMUR; BONE; MODELS; FORCES; TISSUE;
D O I
10.1016/j.medengphy.2008.11.006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Fiber-reinforced composites (FRC) have the potential for use as load-bearing orthopaedic implants if the high strength and elastic modulus of FRC implant can be matched with local requirements. This study tested the in vivo performance of novel FRC implants made of unidirectional glass fibers (E-glass fibers in Bis-GMA and TEGDMA polymeric matrix). The implant surface was covered with bioactive glass granules. Control implants were made of surface-roughened titanium. Stress-shielding effects of the implants were predicted by finite element modelling (FEM). Surgical stabilization of bone metastasis in the subtrochanteric region of the femur was simulated in 12 rabbits. An oblong subtrochanteric defect of a standardized size (reducing the torsional strength of the bones approximately by 66%) was created and an intramedullary implant made of titanium or the FRC composite was inserted. The contralateral femur served as the intact control. At 12 weeks of healing, the femurs were harvested and analyzed by radiography, torsional testing, micro-CT imaging and hard tissue histology. The functional recovery was unremarkable in both groups, although the final analysis revealed two healed undisplaced peri-implant fractures in the group of FRC implants. FEM studies demonstrated differences in stress-shielding effects of the titanium and FRC implants, but the expected biological consequences did not become evident during the follow-up time of the animal study. Biomechanical testing of the retrieved femurs showed no significant differences between the groups. The torsional strength of the fixed bones had returned the level of contralateral intact femurs. Both implants showed ongrowth of intramedullary new bone. No adverse tissue reactions were observed. Based on these favorable results, a large-scale EU-project (NewBone. www.hb.se/ih/polymer/newbone) has been launched for development of orthopaedic FRC implants. (C) 2008 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:461 / 469
页数:9
相关论文
共 27 条
[1]   Early failure of a press-fit carbon fiber hip prosthesis with a smooth surface [J].
Adam, F ;
Hammer, DS ;
Pfautsch, S ;
Westermann, K .
JOURNAL OF ARTHROPLASTY, 2002, 17 (02) :217-223
[2]   Clinical and histologic results related to a low-modulus composite total hip replacement stem [J].
Akhavan, Sam ;
Matthiesen, Mary M. ;
Schulte, Leah ;
Penoyar, Tom ;
Kraay, Matthew J. ;
Rimnac, Clare M. ;
Goldberg, Victor M. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (06) :1308-1314
[3]   Early failure of a carbon-fiber composite femoral component [J].
Allcock, S ;
Ali, MA .
JOURNAL OF ARTHROPLASTY, 1997, 12 (03) :356-358
[4]   Effect of bisphosphonates on periprosthetic bone mineral density after total joint arthroplasty -: A meta-analysis [J].
Bhandari, M ;
Bajammal, S ;
Guyatt, GH ;
Griffith, L ;
Busse, JW ;
Schünemann, H ;
Einhorn, TA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (02) :293-301
[5]  
BROWN SA, 1996, ASTM STP, V1272, P103
[6]   EFFECT OF BIOPSY-HOLE SHAPE AND SIZE ON BONE STRENGTH [J].
CLARK, CR ;
MORGAN, C ;
SONSTEGARD, DA ;
MATTHEWS, LS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1977, 59 (02) :213-217
[7]   Composite technology in load-bearing orthopaedic implants [J].
Evans, SL ;
Gregson, PJ .
BIOMATERIALS, 1998, 19 (15) :1329-1342
[8]   Rabbit knee joint biomechanics: Motion analysis and modeling of forces during hopping [J].
Gushue, DL ;
Houck, J ;
Lerner, AL .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2005, 23 (04) :735-742
[9]  
Hench L.L., 1971, Journal of Biomedical Materials Research Symposium, V36, P117, DOI [10.1002/jbm.820050611, DOI 10.1002/JBM.820050611]
[10]   Mechanical properties and failure behaviour of cylindrical CFRP-implant-rods under torsion load [J].
Hiermer, T ;
Schmitt-Thomas, KG ;
Yang, ZG .
COMPOSITES PART A-APPLIED SCIENCE AND MANUFACTURING, 1998, 29 (11) :1453-1461