Quantification of uric acid, xanthine and hypoxanthine in human serum by HPLC for pharmacodynamic studies

被引:178
作者
Cooper, Nancy
Khosravan, Reza
Erdmann, Carol
Fiene, John
Lee, Jean W.
机构
[1] TAP Pharmaceut Prod Inc, Drug Metab & Pharmacokinet, Lake Forest, IL 60045 USA
[2] MDS Pharma Serv, Lincoln, NE 68501 USA
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2006年 / 837卷 / 1-2期
关键词
uric acid; xanthine oxidase inhibitor; HPLC; pharmacodynamic biornarkers; human serum;
D O I
10.1016/j.jchromb.2006.02.060
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple HPLC method was developed and validated for the determination of uric acid (UA), xanthine (X) and hypoxanthine (HX) concentrations in human serum to support pharmacodynamic (PD) studies of a novel xanthine oxidase inhibitor during its clinical development. Serum proteins were removed by ultrafiltration. The hydrophilic analytes and the I.S. were eluted by 100% aqueous phosphate buffer mobile phase. The hydrophobic matrix components (late peaks) were eluted with a step gradient of a higher organic mobile phase. Validation on linearity, sensitivity, precision, accuracy, stability, and robustness of the method for PD biomarkers (UA, X, and HX) was carried out in a similar manner to that for pharmacokinetic (PK) data where applicable. Issues of selectivity for endogenous biomarker analytes and individual concentration variations were addressed during method validation. Standards were prepared in analyte-free phosphate buffer. Quality control samples were prepared in control serum from individuals not dosed with the xanthine oxidase inhibitor. The method was simple and robust with good accuracy and precision for the measurement of serum UA, X, and HX concentrations. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 10
页数:10
相关论文
共 50 条
  • [31] Cerebrospinal fluid hypoxanthine, xanthine and uric acid levels may reflect glutamate-mediated excitotoxicity in different neurological diseases
    Stover, JF
    Lowitzsch, K
    Kempski, OS
    NEUROSCIENCE LETTERS, 1997, 238 (1-2) : 25 - 28
  • [32] Fabrication of niobium metal organic frameworks anchored carbon nanofiber hybrid film for simultaneous detection of xanthine, hypoxanthine and uric acid
    Dey, Baban
    Ahmad, Md. Wasi
    Sarkhel, Gautam
    Lee, Gang Ho
    Choudhury, Arup
    MICROCHEMICAL JOURNAL, 2023, 186
  • [33] An ultrasensitive electrochemical sensor for simultaneous determination of xanthine, hypoxanthine and uric acid based on Co doped CeO2 nanoparticles
    Lavanya, N.
    Sekar, C.
    Murugan, R.
    Ravi, G.
    MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2016, 65 : 278 - 286
  • [34] Electrochemical synthesis of a novel purine-based polymer and its use for the simultaneous determination of dopamine, uric acid, xanthine and hypoxanthine
    Lan, Dexiang
    Zhang, Lei
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2015, 757 : 107 - 115
  • [35] Spectrofluorimetric method for the determination of uric acid in human serum
    W. W. Bian
    L. Wang
    N. Zhang
    C. Q. Jiang
    Journal of Analytical Chemistry, 2014, 69 : 480 - 484
  • [36] Spectrofluorimetric method for the determination of uric acid in human serum
    Bian, W. W.
    Wang, L.
    Zhang, N.
    Jiang, C. Q.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2014, 69 (05) : 480 - 484
  • [37] HYPOXANTHINE, XANTHINE, AND URIC-ACID IN NEWBORN PIGS DURING HYPOXEMIA FOLLOWED BY RESUSCITATION WITH ROOM AIR OR 100-PERCENT OXYGEN
    POULSEN, JP
    OYASAETER, S
    SAUGSTAD, OD
    CRITICAL CARE MEDICINE, 1993, 21 (07) : 1058 - 1065
  • [38] Simultaneous electrochemical determination of uric acid, xanthine and hypoxanthine based on poly(L-arginine)/graphene composite film modified electrode
    Zhang, Fengyuan
    Wang, Zaihua
    Zhang, Yuzhen
    Zheng, Zhixiang
    Wang, Chunming
    Du, Yongling
    Ye, Weichun
    TALANTA, 2012, 93 : 320 - 325
  • [39] Determination of uric acid in human urine and serum by capillary electrophoresis with chemiluminescence detection
    Zhao, Shulin
    Wang, Jianshi
    Ye, Fanggui
    Liu, Yi-Ming
    ANALYTICAL BIOCHEMISTRY, 2008, 378 (02) : 127 - 131
  • [40] Multianalyte sensor for the simultaneous determination of hypoxanthine, xanthine and uric acid based on a preanodized nontronite-coated screen-printed electrode
    Zen, JM
    Lai, YY
    Yang, HH
    Kumar, AS
    SENSORS AND ACTUATORS B-CHEMICAL, 2002, 84 (2-3) : 237 - 244