Effect of generator nonlinearities on the accuracy of respiratory impedance measurements by forced oscillation

被引:8
作者
de Melo, PL
Werneck, MM
Giannella-Neto, A [1 ]
机构
[1] Univ Fed Rio de Janeiro, COPPE, Biomed Engn Program, BR-21941 Rio De Janeiro, Brazil
[2] Univ Fed Rio de Janeiro, COPPE, Met & Mat Engn Program, BR-21941 Rio De Janeiro, Brazil
关键词
forced oscillation; respiratory impedance; nonlinearities; unbiased estimators;
D O I
10.1007/BF02344697
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Measurements of respiratory impedance by means of the forced oscillation technique (FOT) are usually made using a loudspeaker as the excitation device, Its nonlinear nature can introduce artifacts that coincide with the frequencies applied to excite the respiratory system, limiting the accuracy of the impedance estimation. In this paper, this hypothesis is evaluated in the case of both a traditional estimator and the unbiased estimator proposed by DAROCZY and HANTOS (1982). A simulated study under apnoea conditions in the pressure range 0.5-3.0 cmH(2)O peak-to-peak reveals that loudspeaker nonlinearities introduce a characteristic pattern of dispersion in both the resistance and reactance curves that can be significantly decreased (p similar or equal to 0.03, signtest) by reducing the nonlinearities. A simulation of spontaneous breathing shows the same pattern, and is observed in the case of traditional as well as unbiased estimators. The dispersion is quantified by the mean absolute distance between the theoretical and simulated data and decreases with the reduction of nonlinearities when impedance is estimated with a traditional estimator (from 6.63 to 4.72% in real estimates and from 6.78 to 3.47% in imaginary estimates) as well as with an unbiased estimator (real estimates from 4.84 to 1.57% and 5.61 to 2.06% in imaginary estimates). Studies with normal subjects show the same dispersion pattern, which decreases if the generator nonlinearities are reduced, These results supply substantial evidence that reducing generator nonlinearities can contribute to the production of more reliable mechanical impedance FOT measurements.
引用
收藏
页码:102 / 108
页数:7
相关论文
共 29 条
[1]   Respiratory resistance by end-inspiratory occlusion and forced oscillations in intubated patients [J].
Beydon, L ;
Malassine, P ;
Lorino, AM ;
Mariette, C ;
Bonnet, F ;
Harf, A ;
Lorino, H .
JOURNAL OF APPLIED PHYSIOLOGY, 1996, 80 (04) :1105-1111
[2]  
BROCHARD L, 1987, AM REV RESPIR DIS, V135, P579
[3]  
CAUBERGHS M, 1991, EUR RESPIR REV, V1, P206
[4]  
COLLOMS M, 1991, HIGH PERFORMANCE LOU
[5]   AN IMPROVED FORCED OSCILLATORY ESTIMATION OF RESPIRATORY IMPEDANCE [J].
DAROCZY, B ;
HANTOS, Z .
INTERNATIONAL JOURNAL OF BIO-MEDICAL COMPUTING, 1982, 13 (03) :221-235
[6]  
Daroczy B., 1991, EUR RESPIR REV, V1, P183
[7]   CHARACTERIZATION AND VALIDATION OF FORCED INPUT METHOD FOR RESPIRATORY IMPEDANCE MEASUREMENT [J].
DELAVAULT, E ;
SAUMON, G ;
GEORGES, R .
RESPIRATION PHYSIOLOGY, 1980, 40 (01) :119-136
[8]   OSCILLATION MECHANICS OF LUNGS AND CHEST IN MAN [J].
DUBOIS, AB ;
BRODY, AW ;
LEWIS, DH ;
BURGESS, BF .
JOURNAL OF APPLIED PHYSIOLOGY, 1956, 8 (06) :587-594
[9]  
DUVIVIER C, 1990, INNOV TECH BIOL MED, V11, P381
[10]   TIME-DOMAIN DIGITAL-FILTER TO IMPROVE SIGNAL-TO-NOISE RATIO IN RESPIRATORY IMPEDANCE MEASUREMENTS [J].
FARRE, R ;
ROTGER, M ;
NAVAJAS, D .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1991, 29 (01) :18-24