Small-animal MRI: signal-to-noise ratio comparison at 7 and 1.5 T with multiple-animal acquisition strategies

被引:40
作者
Beuf, Olivier [1 ]
Jaillon, Franck [1 ]
Saint-Jalmes, Herve [1 ]
机构
[1] Univ Lyon 1, ESCPE, Lab RMN, CNRS,UMR 5012, F-69616 Villeurbanne, France
关键词
multiple-animal imaging; small animal; 7T; 1.5T; SNR comparison;
D O I
10.1007/s10334-006-0048-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The purpose of this study was to compare the signal-to-noise ratio (SNR) of phantom and rat brain images performed at 1.5 T on a clinical MR system and at 7 T on a small-animal experimental system. Comparison was carried out by taking into account SNR values based on a single sample acquisition at 1.5 and 7 T as well as on simultaneous imaging of multiple samples at 1.5 T. Methods: SNR was experimentally assessed on a phantom and rat brains at 1.5 and 7 T using 25 mm surface coils and compared to theoretical SNR gain estimations. The feasibility of multiple-animal imaging, using the hardware capabilities available on the 1.5 T system, was demonstrated. Finally, rat brain images obtained on a single animal at 7 T and on multiple animals acquired simultaneously at 1.5 T were compared. Results: Experimentally determined SNR at 7 T was far below theoretical estimations. Taking into account chemical shift, susceptibility artifacts and modifications of T1 and T2 relaxation times at higher field, a 7-T system holds limited advantage over a 1.5-T system. Instead, a multiple-animal acquisition methodology was demonstrated on a clinical 1.5-T scanner. This acquisition method significantly increases imaging efficiency and competes with single animal acquisitions at higher field. Conclusion: Multiple-animal imaging using a standard clinical scanner has a great potential as a high-throughput acquisition method for small animals.
引用
收藏
页码:202 / 208
页数:7
相关论文
共 21 条
  • [1] In vivo multiple-mouse MRI at 7 Tesla
    Bock, NA
    Nieman, BJ
    Bishop, JB
    Henkelman, RM
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (05) : 1311 - 1316
  • [2] High-resolution longitudinal screening with magnetic resonance imaging in a murine brain cancer model
    Bock, NA
    Zadeh, G
    Davidson, LM
    Qian, BP
    Sled, JG
    Guha, A
    Henkelman, RM
    [J]. NEOPLASIA, 2003, 5 (06): : 546 - 554
  • [3] Multiple-mouse MRI
    Bock, NA
    Konyer, NB
    Henkelman, RM
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (01) : 158 - 167
  • [4] Callaghan P.T., 1991, Principles of nuclear magnetic resonance microscopy, V1st
  • [5] Perspectives with cryogenic RF probes in biomedical MRI
    Darrasse, L
    Ginefri, JC
    [J]. BIOCHIMIE, 2003, 85 (09) : 915 - 937
  • [6] NMR probe for the simultaneous acquisition of multiple samples
    Fisher, G
    Petucci, C
    MacNamara, E
    Raftery, D
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1999, 138 (01) : 160 - 163
  • [7] Haase A, 2000, CONCEPT MAGNETIC RES, V12, P361, DOI 10.1002/1099-0534(2000)12:6<361::AID-CMR1>3.0.CO
  • [8] 2-L
  • [9] SENSITIVITY OF THE ZEUGMATOGRAPHIC EXPERIMENT INVOLVING HUMAN SAMPLES
    HOULT, DI
    LAUTERBUR, PC
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1979, 34 (02) : 425 - 433
  • [10] Q-FACTOR MEASUREMENT WITH A SCALAR NETWORK ANALYZER
    KAJFEZ, D
    [J]. IEE PROCEEDINGS-MICROWAVES ANTENNAS AND PROPAGATION, 1995, 142 (05) : 369 - 372