Comparison of analytical performances of inductively coupled plasma mass spectrometry and inductively coupled plasma atomic emission spectrometry for trace analysis of bismuth and bismuth oxide

被引:17
作者
Medvedev, Nickolay S. [1 ]
Shaverina, Anastasiya, V [1 ]
Tsygankova, Alphiya R. [1 ,2 ]
Saprykin, Anatoly, I [1 ,2 ]
机构
[1] NIIC SB RAS, 3 Lavrentiev Ave, Novosibirsk 630090, Russia
[2] Novosibirsk Natl Res State Univ, 2 Pirogova St, Novosibirsk 630090, Russia
关键词
ICP-MS; ICP-AES; Matrix interferences; Bismuth; Bismuth oxide; ABSORPTION-SPECTROMETRY; CONCOMITANT ELEMENTS; OPERATING-CONDITIONS; INTERFERENCES; BI4GE3O12;
D O I
10.1016/j.sab.2018.01.017
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The paper presents a comparison of analytical performances of inductively coupled plasma mass spectrometry (ICP-MS) and inductively coupled plasma atomic emission spectrometry (ICP-AES) for trace analysis of high purity bismuth and bismuth oxide. Matrix effects in the ICP-MS and ICP-AES methods were studied as a function of Bi concentration, ICP power and nebulizer flow rate. For ICP-MS the strong dependence of the matrix effects versus the atomic mass of analytes was observed. For ICP-AES the minimal matrix effects were achieved for spectral lines of analytes with low excitation potentials. The optimum degree of sample dilution providing minimum values of the limits of detection (LODs) was chosen. Both methods let us to reach LODs from n.10(-7) to n.10(-4) wt% for more than 50 trace elements. For most elements the LODs of ICP-MS were lower in comparison to ICP-AES. Validation of accuracy of the developed techniques was performed by "added-found" experiments and by comparison of the results of ICP-MS and ICP-AES analysis of high-purity bismuth oxide. (C) 2018 Elsevier B.V. All rights reserved.
引用
收藏
页码:23 / 28
页数:6
相关论文
共 31 条
[1]   Matrix effects in inductively coupled plasma mass spectrometry: A review [J].
Agatemor, Christian ;
Beauchemin, Diane .
ANALYTICA CHIMICA ACTA, 2011, 706 (01) :66-83
[2]  
Chanysheva T.A., 2002, ANAL KONTROL, V6, P298
[3]  
Devyatych G. G., 1988, HIGH-PURITY SUBST, V4, P5
[4]   INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY - A REVIEW [J].
EVANS, EH ;
GIGLIO, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (01) :1-18
[5]  
Földvári I, 2000, OPT COMMUN, V177, P105, DOI 10.1016/S0030-4018(00)00560-5
[6]   Effect of concomitant elements on the distribution of ions in inductively coupled plasma-mass spectroscopy. Part 1. Elemental ions [J].
Fraser, MM ;
Beauchemin, D .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2000, 55 (11) :1705-1731
[8]   Holographic data storage in photorefractive bismuth tellurite [J].
Horn, W. ;
Foldvari, I. ;
Denz, C. .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2008, 41 (22)
[9]   Thermoelectric properties of p-type (Bi2Te3)x(Sb2Te3)1-x crystals prepared via zone melting [J].
Jiang, J ;
Chen, LD ;
Bai, SQ ;
Yao, Q ;
Wang, Q .
JOURNAL OF CRYSTAL GROWTH, 2005, 277 (1-4) :258-263
[10]  
Korzhik M.V., 1992, ZH PRIKL SPEKTROSK, V57, P299