Sources of Error in Toric Intraocular Lens Power Calculation
被引:45
作者:
Hirnschall, Nino
[1
]
Findl, Oliver
[1
,3
]
Bayer, Natascha
[1
]
Leisser, Christoph
[1
]
Norrby, Sverker
Zimper, Eva
[1
]
Hoffmann, Peter
[2
]
机构:
[1] Hanusch Hosp, Vienna Inst Res Ocular Surg, A Karl Landsteiner Inst, Vienna, Austria
[2] Eye & Laser Clin, Castrop Rauxel, Germany
[3] Moorfields Eye Hosp NHS Fdn Trust, London, England
关键词:
OPTICAL COHERENCE TOMOGRAPHY;
RESIDUAL REFRACTIVE ASTIGMATISM;
POSTERIOR CORNEAL SURFACE;
CATARACT-SURGERY;
VECTOR ANALYSIS;
BIOMETER REPEATABILITY;
ROTATIONAL STABILITY;
CLINICAL-OUTCOMES;
TEAR FILM;
IMPLANTATION;
D O I:
10.3928/1081597X-20200729-03
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
PURPOSE: To evaluate the influencing factors on remaining astigmatism after implanting a toric intraocular tens (IOL) during cataract surgery. METHODS: This retrospective study included parameters that were considered to have an influence on toric IOL power calculation. Therefore, data from the literature and the authors' own data were used. This included axial eye length, anterior chamber depth, central corneal thickness, corneal radii (anterior and posterior), diurnal changes of the cornea, interdevice differences, rotational misalignment of the IOL, tilt and decentration of the IOL, pupil size, angle kappa, and surgically induced astigmatism. Ray-tracing and Gaussian error propagation analysis was performed to quantify the sources of error. RESULTS: In total, 4,949 eyes (4,365 eyes of 42 studies and 584 eyes of retrospectively analyzed study data) were included in the study and the difference vector between aimed and calculated remaining astigmatism was 0.81 diopters (D). The main source of error was the preoperative measurement of the cornea (27%), followed by IOL misalignment (14.4%) and IOL tilt (11.3%). Other factors, such as angle kappa (10.9%), pupil size (8.1%), surgically induced astigmatism (7.8%), anterior chamber depth (7.5%), axial eye length (7.5%), and decentration (5.6%), also contributed to the refractive astigmatic error. CONCLUSIONS: The main source of error in toric IOL power calculation is the preoperative corneal measurement followed by IOL misalignment and tilt.
引用
收藏
页码:646 / 652
页数:7
相关论文