Sources of Error in Toric Intraocular Lens Power Calculation

被引:41
作者
Hirnschall, Nino [1 ]
Findl, Oliver [1 ,3 ]
Bayer, Natascha [1 ]
Leisser, Christoph [1 ]
Norrby, Sverker
Zimper, Eva [1 ]
Hoffmann, Peter [2 ]
机构
[1] Hanusch Hosp, Vienna Inst Res Ocular Surg, A Karl Landsteiner Inst, Vienna, Austria
[2] Eye & Laser Clin, Castrop Rauxel, Germany
[3] Moorfields Eye Hosp NHS Fdn Trust, London, England
关键词
OPTICAL COHERENCE TOMOGRAPHY; RESIDUAL REFRACTIVE ASTIGMATISM; POSTERIOR CORNEAL SURFACE; CATARACT-SURGERY; VECTOR ANALYSIS; BIOMETER REPEATABILITY; ROTATIONAL STABILITY; CLINICAL-OUTCOMES; TEAR FILM; IMPLANTATION;
D O I
10.3928/1081597X-20200729-03
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate the influencing factors on remaining astigmatism after implanting a toric intraocular tens (IOL) during cataract surgery. METHODS: This retrospective study included parameters that were considered to have an influence on toric IOL power calculation. Therefore, data from the literature and the authors' own data were used. This included axial eye length, anterior chamber depth, central corneal thickness, corneal radii (anterior and posterior), diurnal changes of the cornea, interdevice differences, rotational misalignment of the IOL, tilt and decentration of the IOL, pupil size, angle kappa, and surgically induced astigmatism. Ray-tracing and Gaussian error propagation analysis was performed to quantify the sources of error. RESULTS: In total, 4,949 eyes (4,365 eyes of 42 studies and 584 eyes of retrospectively analyzed study data) were included in the study and the difference vector between aimed and calculated remaining astigmatism was 0.81 diopters (D). The main source of error was the preoperative measurement of the cornea (27%), followed by IOL misalignment (14.4%) and IOL tilt (11.3%). Other factors, such as angle kappa (10.9%), pupil size (8.1%), surgically induced astigmatism (7.8%), anterior chamber depth (7.5%), axial eye length (7.5%), and decentration (5.6%), also contributed to the refractive astigmatic error. CONCLUSIONS: The main source of error in toric IOL power calculation is the preoperative corneal measurement followed by IOL misalignment and tilt.
引用
收藏
页码:646 / 652
页数:7
相关论文
共 68 条
  • [1] Prediction of refractive outcomes with toric intraocular lens implantation
    Abulafia, Adi
    Barrett, Graham D.
    Kleinmann, Guy
    Ofir, Shay
    Levy, Adi
    Marcovich, Arie L.
    Michaeli, Adi
    Koch, Douglas D.
    Wang, Li
    Assia, Ehud I.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (05) : 936 - 944
  • [2] Visual function and patient experience after bilateral implantation of toric intraocular lenses
    Ahmed, Iqbal Ike K.
    Rocha, Guillermo
    Slomovic, Allan R.
    Climenhaga, Harold
    Gohill, Jit
    Gregoire, Alain
    Ma, Joseph
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (04) : 609 - 616
  • [3] Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500
    Akman, Ahmet
    Asena, Leyla
    Gungor, Sirel Gur
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2016, 100 (09) : 1201 - 1205
  • [4] Vector analysis of astigmatic changes after cataract surgery with toric intraocular lens implantation
    Alio, Jorge L.
    Pinero, David P.
    Tomas, Javier
    Aleson, Alicia
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (06) : 1038 - 1049
  • [5] Dual versus single Scheimpflug camera for anterior segment analysis: Precision and agreement
    Aramberri, Jaime
    Araiz, Luis
    Garcia, Ane
    Illarramendi, Igor
    Olmos, Jaione
    Oyanarte, Izaskun
    Romay, Amaya
    Vigara, Itxaso
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (11) : 1934 - 1949
  • [6] Optical models of the human eye
    Atchison, David A.
    Thibos, Larry N.
    [J]. CLINICAL AND EXPERIMENTAL OPTOMETRY, 2016, 99 (02) : 99 - 106
  • [7] Repeatability, Reproducibility, and Agreement of Two Scheimpflug-Placido Anterior Corneal Analyzers for Posterior Corneal Surface Measurement
    Bao, Fangjun
    Savini, Giacomo
    Shu, Bao
    Zhu, Senmiao
    Gao, Rongrong
    Dang, Guanxin
    Yu, Ayong
    Wang, Qinmei
    Huang, Jinhai
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2017, 33 (08) : 524 - U114
  • [8] Astigmatism management in cataract surgery with the AcrySof toric intraocular lens
    Bauer, Noel J. C.
    de Vries, Niels E.
    Webers, Carroll A. B.
    Hendrikse, Fred
    Nuijts, Rudy M. M. A.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (09) : 1483 - 1488
  • [9] Bayer N, 2017, ASS RES VIS OPHTH M
  • [10] Agreement between a partial coherence interferometer and 2 manual keratometers
    Bullimore, Mark A.
    Buehren, Tobias
    Bissmann, Wilfried
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (10) : 1550 - 1560