Comparison of MLC error sensitivity of various commercial devices for VMAT pre-treatment quality assurance

被引:24
作者
Saito, Masahide [1 ]
Sano, Naoki [1 ]
Shibata, Yuki [1 ]
Kuriyama, Kengo [1 ]
Komiyama, Takafumi [1 ]
Marino, Kan [1 ]
Aoki, Shinichi [1 ]
Ashizawa, Kazunari [1 ]
Yoshizawa, Kazuya [1 ]
Onishi, Hiroshi [1 ]
机构
[1] Univ Yamanashi, Dept Radiol, Yamanashi, Japan
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2018年 / 19卷 / 03期
关键词
MLC error; patient QA; radiotherapy; VMAT; VIVO EPID DOSIMETRY; DOSE RECONSTRUCTION; IMRT QA; PROSTATE IMRT; PATIENT; VERIFICATION; SYSTEMS; ACCURACY; DELIVERY; THERAPY;
D O I
10.1002/acm2.12288
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to compare the MLC error sensitivity of various measurement devices for VMAT pre-treatment quality assurance (QA). This study used four QA devices (Scandidos Delta4, PTW 2D-array, iRT systems IQM, and PTW Farmer chamber). Nine retrospective VMAT plans were used and nine MLC error plans were generated for all nine original VMAT plans. The IQM and Farmer chamber were evaluated using the cumulative signal difference between the baseline and error-induced measurements. In addition, to investigate the sensitivity of the Delta4 device and the 2D-array, global gamma analysis (1%/1, 2%/2, and 3%/3 mm), dose difference (1%, 2%, and 3%) were used between the baseline and error-induced measurements. Some deviations of the MLC error sensitivity for the evaluation metrics and MLC error ranges were observed. For the two ionization devices, the sensitivity of the IQM was significantly better than that of the Farmer chamber (P < 0.01) while both devices had good linearly correlation between the cumulative signal difference and the magnitude of MLC errors. The pass rates decreased as the magnitude of the MLC error increased for both Delta4 and 2D-array. However, the small MLC error for small aperture sizes, such as for lung SBRT, could not be detected using the loosest gamma criteria (3%/3 mm). Our results indicate that DD could be more useful than gamma analysis for daily MLC QA, and that a large-area ionization chamber has a greater advantage for detecting systematic MLC error because of the large sensitive volume, while the other devices could not detect this error for some cases with a small range of MLC error.
引用
收藏
页码:87 / 93
页数:7
相关论文
共 26 条
  • [1] EPID dosimetry for pretreatment quality assurance with two commercial systems
    Bailey, Daniel W.
    Kumaraswamy, Lalith
    Bakhtiari, Mohammad
    Malhotra, Harish K.
    Podgorsak, Matthew B.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (04): : 82 - 99
  • [2] Evaluation of a new VMAT QA device, or the "X" and "O" array geometries
    Feygelman, Vladimir
    Zhang, Geoffrey
    Stevens, Craig
    Nelms, Benjamin E.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02): : 146 - 168
  • [3] Patient QA systems for rotational radiation therapy: A comparative experimental study with intentional errors
    Fredh, Anna
    Scherman, Jonas Bengtsson
    Fog, Lotte S.
    Af Rosenschold, Per Munck
    [J]. MEDICAL PHYSICS, 2013, 40 (03)
  • [4] A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems
    Hussein, Mohammad
    Rowshanfarzad, Pejman
    Ebert, Martin A.
    Nisbet, Andrew
    Clark, Catharine H.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2013, 109 (03) : 370 - 376
  • [5] An integral quality monitoring system for real-time verification of intensity modulated radiation therapy
    Islam, Mohammad K.
    Norrlinger, Bernhard D.
    Smale, Jason R.
    Heaton, Robert K.
    Galbraith, Duncan
    Fan, Cary
    Jaffray, David A.
    [J]. MEDICAL PHYSICS, 2009, 36 (12) : 5420 - 5428
  • [6] Evaluation of patient DVH-based QA metrics for prostate VMAT: correlation between accuracy of estimated 3D patient dose and magnitude of MLC misalignment
    Kadoya, Noriyuki
    Saito, Masahide
    Ogasawara, Makoto
    Fujita, Yukio
    Ito, Kengo
    Sato, Kiyokazu
    Kishi, Kazuma
    Dobashi, Suguru
    Takeda, Ken
    Jingu, Keiichi
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 179 - 189
  • [7] Task Group 142 report: Quality assurance of medical accelerators
    Klein, Eric E.
    Hanley, Joseph
    Bayouth, John
    Yin, Fang-Fang
    Simon, William
    Dresser, Sean
    Serago, Christopher
    Aguirre, Francisco
    Ma, Lijun
    Arjomandy, Bijan
    Liu, Chihray
    Sandin, Carlos
    Holmes, Todd
    [J]. MEDICAL PHYSICS, 2009, 36 (09) : 4197 - 4212
  • [8] Retrospective IMRT dose reconstruction based on cone-beam CT and MLC log-file
    Lee, Louis
    Le, Quynh-Thu
    Xing, Lei
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (02): : 634 - 644
  • [9] Validation of dynamic MLC-controller log files using a two-dimensional diode array
    Li, JG
    Dempsey, JF
    Ding, L
    Liu, CR
    Palta, JR
    [J]. MEDICAL PHYSICS, 2003, 30 (05) : 799 - 805
  • [10] Litzenberg Dale W, 2002, J Appl Clin Med Phys, V3, P63, DOI 10.1120/1.1449362