Evaluation of the EDR-2 film for relative dosimetry of high-energy photon and electron beams

被引:4
作者
Ahmad, Munir [1 ]
Chen, Zhe [1 ]
Song, Haijun [1 ]
Deng, Jun [1 ]
Nath, Ravinder [1 ]
机构
[1] Yale Univ, Sch Med, Dept Therapeut Radiol, New Haven, CT 06520 USA
关键词
D O I
10.1093/rpd/ncj006
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A sensitometric study of Kodak XV and EDR-2 radiographic films (Eastman Kodak Company, Rochester, NY) was performed using photons ranging from 75 kV to 18 MV and electrons ranging from 6 to 20 MeV. To investigate the applicability of the EDR-2 film for clinical radiation dosimetry, percentage depth-doses, profiles and distributions in open and dynamically wedged fields were measured using film and compared with data from a linear diode. Moreover, conventional quality assurance dose parameters were measured, including open-field dose profiles to determine flatness and symmetry of photon and electron beams. Finally, film was employed to validate dose distributions produced by complex computerised treatment planning techniques. Our conclusion is that the EDR-2 film is an effective tool for relative dosimetry of photon and electron beams.
引用
收藏
页码:159 / 162
页数:4
相关论文
共 3 条
[1]  
Leavitt D D, 1997, Med Dosim, V22, P171, DOI 10.1016/S0958-3947(97)00014-9
[2]   Improved treatment of pelvis and inguinal nodes using modified segmental boost technique: Dosimetric evaluation [J].
Moran, M ;
Lund, MW ;
Ahmad, M ;
Trumpore, H ;
Haffty, B ;
Nath, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (05) :1523-1530
[3]   FILM DOSIMETRY OF MEGAVOLTAGE PHOTON BEAMS - A PRACTICAL METHOD OF ISODENSITY-TO-ISODOSE CURVE CONVERSION [J].
WILLIAMSON, JF ;
KHAN, FM ;
SHARMA, SC .
MEDICAL PHYSICS, 1981, 8 (01) :94-98