Characterization of a modular microfluidic photoionization detector

被引:11
作者
Rezende, Gustavo Coelho [1 ]
Le Calve, Stephane [2 ,3 ]
Brandner, Jurgen J. [4 ]
Newport, David [1 ]
机构
[1] Univ Limerick, Sch Engn, Bernal Inst, Limerick V94 T9PX, Ireland
[2] Unistra, Inst Chim & Procede Energie Environm & Sante, Grp Atmospher Phys Chem, ICPEES,UMR 7515,CNRS, 25 Rue Becquerel, F-67087 Strasbourg 02, France
[3] In Air Solut, 25 Rue Becquerel, F-67000 Strasbourg, France
[4] Karlsruhe Inst Technol, Inst Microstruct Technol IMT, Karlsruhe, Germany
基金
欧盟地平线“2020”;
关键词
Photoionization detector; Miniaturization; Microfabrication; Gas chromatography; VOC detection; Air quality; GAS-CHROMATOGRAPHY; EXPOSURE; LEVEL; GC;
D O I
10.1016/j.snb.2020.128667
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Photoionization detectors with a small ionization chamber can contribute to overall gas analyser miniaturization. This work reports the characterization of a microfluidic photoionization detector prototype (mu PID) which is constructed to be modular for easy replacement of the components and maintenance. The device is fabricated by micromilling and electrical discharge machining, dispensing clean room fabrication techniques. The mu PID ionization chamber is a microchannel and four channel designs are presented and tested in experiments in order to evaluate the influence of the geometrical parameters on the detector performance. The chamber volumes of the channel designs range from 1.1 to 6.7 mu L. Experimental characterization of the prototype is performed when it is used without and with a portable gas chromatograph (GC) for volatile organic compounds (VOCs) analysis. When a sample of 100 ppm toluene is injected directly into the mu PID, it can generate a current signal up to similar to 4 nA. When used without a GC, the device showed a linear response for an injection of toluene gas concentrations ranging from 1 to 100 ppm. A combination of high illumination area and electrode area resulted in the highest signal in the mu PID with a detection limit of similar to 40 ppb of toluene. When integrated to the portable GC, the detection limit reached for toluene is similar to 140 ppb. The detection limit for toluene was further reduced to low ppb levels (similar to 5 ppb) when a preconcentrator was integrated into the sampling loop of the portable GC.
引用
收藏
页数:10
相关论文
共 50 条
  • [11] Design of the Photoionization Detector for the Detection of Volatile Organic Compounds
    Sun Ruixia
    Liang Ting
    Fan Qiusheng
    Tan Qiulin
    Liu Jun
    ISTM/2009: 8TH INTERNATIONAL SYMPOSIUM ON TEST AND MEASUREMENT, VOLS 1-6, 2009, : 2317 - 2320
  • [12] Design of photoionization detector based on the miniature gas chromatography
    Chai Shou-chen
    Yang Peng
    Yang Cheng-jia
    Cai Chun-li
    Yu Na
    COMPUTING, CONTROL AND INDUSTRIAL ENGINEERING IV, 2013, 823 : 291 - +
  • [13] Colorimetric Readout Based Photoionization Detector for Gas Chromatographs
    Mao, Jingqin
    Liu, Longze
    Atwa, Yahya
    Hou, Junming
    Shakeel, Hamza
    IEEJ TRANSACTIONS ON ELECTRICAL AND ELECTRONIC ENGINEERING, 2024, 19 (05) : 845 - 851
  • [14] Characterization of chlorinated compounds using a dual chlorine-selective pulsed discharge emission detector-helium-pulsed discharge photoionization detector system
    Sun, KF
    Wentworth, WE
    Stearns, SD
    JOURNAL OF CHROMATOGRAPHY A, 2000, 872 (1-2) : 179 - 189
  • [15] Windowless pulsed-discharge photoionization detector - Application to qualitative analysis of volatile organic compounds
    Gremaud, G
    Wentworth, WE
    Zlatkis, A
    Swatloski, R
    Chen, ECM
    Stearns, SD
    JOURNAL OF CHROMATOGRAPHY A, 1996, 724 (1-2) : 235 - 250
  • [16] Supported capillary membrane sampling-gas chromatography on a valve with a pulsed discharge photoionization detector
    Emmert, GL
    Brown, MA
    Liao, ZH
    Cao, G
    Duty, C
    ANALYTICA CHIMICA ACTA, 2006, 560 (1-2) : 197 - 206
  • [17] A Micro Helium-Discharge PhotoIonization Detector for Gas Sensing
    Narayanan, Shree
    Agah, Masoud
    Rice, Gary
    2013 IEEE SENSORS, 2013, : 125 - 128
  • [18] Determination of short-term exposure with a direct reading photoionization detector
    Poirot, P
    Subra, I
    Gérardin, F
    Baudin, V
    Grossmann, S
    Héry, M
    ANNALS OF OCCUPATIONAL HYGIENE, 2004, 48 (01) : 75 - 84
  • [19] A PMUT Based Photoacoustic System as a Microfluidic Concentration Detector
    Roy, Kaustav
    Shastri, Vijayendra
    Kumar, Akshay
    Rout, Jagi
    Isha
    Kalyan, Kritank
    Prakash, Jaya
    Pratap, Rudra
    PHOTONS PLUS ULTRASOUND: IMAGING AND SENSING 2022, 2022, 11960
  • [20] Application of homemade portable gas chromatography coupled to photoionization detector for the detection of volatile organic compounds in an industrial park
    Pang, Xiaobing
    Li, Wenke
    Wang, Shuaiqi
    Wu, Zhentao
    Sun, Songhua
    Lyu, Yan
    Chen, Dongzhi
    Li, Haiyan
    JOURNAL OF CHROMATOGRAPHY A, 2023, 1704