Water exchange and inflow affect the accuracy of T1-GRE blood volume measurements:: Implications for the evaluation of tumor angiogenesis

被引:55
作者
Kim, YR [1 ]
Rebro, KJ [1 ]
Schmainda, KM [1 ]
机构
[1] Med Coll Wisconsin, Biophys Res Inst, Milwaukee, WI 53226 USA
关键词
angiogenesis; blood volume; water exchange; T(1)contrast agent; inflow; tumor;
D O I
10.1002/mrm.10175
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The goal of this study was to determine the degree to which vascular water exchange and blood flowing into an imaging slice affect the accuracy of blood volume measurements of brain and tumor tissue when using intravascular T-1 contrast agents. The study was performed using 2D and 3D gradient-echo imaging sequences, since these are two of the most commonly used MRI methods used to evaluate tissue blood volume fraction. Computer simulations were performed and measurements made in a rat 9L gliosarcoma brain tumor model. The computer simulations demonstrate that, with either water exchange or inflow effects alone, the dependence on the physiologic and imaging parameters can be well characterized and therefore potentially offset. In the exchange only case, the parametric dependence of 3D simulations suggest that the best accuracy is achieved with high flip angles, short TR, and low blood contrast agent concentrations. However, for a 2D GRE sequence which is influenced by both water exchange and inflow, the simulations predict that the error trend as a function of the imaging and physiologic parameters is unpredictable and therefore difficult to compensate. With both 2D and 3D GRE the measured blood volume data in rat brain and tumor tissue demonstrate tissue-specific trends, which reflect differences in the considered physiologic parameters. The experimental data strongly support the computer simulations and also indicate that minimization of the physiological effects by proper selection of imaging parameters, contrast concentration, and volume calculation methods is crucial for accurate assessment of absolute blood volume fraction.
引用
收藏
页码:1110 / 1120
页数:11
相关论文
共 31 条
[1]  
ARTEMOV D, 2000, P 8 SCI M ISMRM DENV, V2, P1051
[2]  
BHUJWALLA ZM, 2000, P 8 SCI M ISMRM DENV, V2, P1063
[3]  
BHUJWALLA ZM, 1999, P 7 SCI M ISMRM PHIL, V1, P146
[4]   A NEW MACROMOLECULE AS A CONTRAST AGENT FOR MR-ANGIOGRAPHY - PREPARATION, PROPERTIES, AND ANIMAL STUDIES [J].
BOGDANOV, AA ;
WEISSLEDER, R ;
FRANK, HW ;
BOGDANOVA, AV ;
NOSSIF, N ;
SCHAFFER, BK ;
TSAI, E ;
PAPISOV, MI ;
BRADY, TJ .
RADIOLOGY, 1993, 187 (03) :701-706
[5]  
CELIK A, 1999, P 7 SCI M ISMRM PHIL, V3, P1872
[6]  
DEAN BL, 1992, AM J NEURORADIOL, V13, P39
[7]   A MRI spatial mapping technique for microvascular permeability and tissue blood volume based on macromolecular contrast agent distribution [J].
Demsar, F ;
Roberts, TPL ;
Schwickert, HC ;
Shames, DM ;
vanDijke, CF ;
Mann, JS ;
Saeed, M ;
Brasch, RC .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (02) :236-242
[8]   Improving MR quantification of regional blood volume with intravascular T-1 contrast agents: Accuracy, precision, and water exchange [J].
Donahue, KM ;
Weisskoff, RM ;
Chesler, DA ;
Kwong, KK ;
Bogdanov, AA ;
Mandeville, JB ;
Rosen, BR .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (06) :858-867
[9]  
ESKEY CJ, 1992, CANCER RES, V52, P6010
[10]  
FOLKMAN J, 1986, CANCER RES, V46, P467