Contrast sensitivity with bifocal intraocular lenses is halved, as measured with the Freiburg Vision Test (FrACT), yet patients are happy

被引:22
作者
Anton, Alexandra [1 ]
Boehringer, Daniel [1 ]
Bach, Michael [1 ]
Reinhard, Thomas [1 ]
Birnbaum, Florian [1 ,2 ]
机构
[1] Univ Freiburg, Ctr Eye, D-79106 Freiburg, Germany
[2] Klinikum Bremen Mitte GmbH, Hosp Eye, D-28177 Bremen, Germany
关键词
Multifocal intraocular lens; Visual acuity; Contrast sensitivity; VISUAL-ACUITY TEST; BINOCULAR ENHANCEMENT; EYES;
D O I
10.1007/s00417-014-2565-y
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
As the desire for spectacle independence following cataract surgery grows, so does interest in the implantation of multifocal intraocular lenses. However, glare phenomena, reduced intermediate vision and loss of image quality are known problems associated with this new generation of lenses. We compared the functional results achieved by the implantation of the diffractive-refractive Acri.LISA 366D lenses to those achieved from implanting monofocal Acri.Smart 46LC lenses. In a retrospective data analysis we followed ten patients who received bifocal intraocular lenses (Acri.LISA 366D) and ten patients who received monofocal intraocular lenses (Acri.Smart 46LC). Lenses were always implanted in both eyes. In each group we assessed visual acuity and contrast sensitivity with the Freiburg Vison Test (FrACT) at multiple distances ranging from 0.5 to 5 m. Additionally, we assessed near vision with the Birkhaeuser charts. We also evaluated photopic phenomena and patient satisfaction using a standardised questionnaire. One patient in the Acri.LISA group and six patients in the control group missed the recommended follow-up visits. We found good uncorrected distance and near visual acuity. Only three of nine patients wore glasses occasionally. Although their contrast sensitivity decreased by a factor of two compared to the control group, patients did not complain about visual disturbances. Patient satisfaction was predominantly high following Acri.LISA 366D implantation. With the Acri.LISA 366D, patients demonstrate excellent near and distance vision, albeit with reduced contrast sensitivity. Acri Lisa is a viable option in patients that do not want to depend on spectacles.
引用
收藏
页码:539 / 544
页数:6
相关论文
共 19 条
  • [1] Prospective study of the Acri.LISA bifocal intraocular lens
    Alfonso, Jose F.
    Fernandez-Vega, Luis
    Senaris, Ana
    Montes-Mico, Robert
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (11) : 1930 - 1935
  • [2] [Anonymous], 2006, R LANG ENV STAT COMP
  • [3] Improving the design of the letter contrast sensitivity test
    Arditi, A
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (06) : 2225 - 2229
  • [4] The Freiburg Visual Acuity test - Automatic measurement of visual acuity
    Bach, M
    [J]. OPTOMETRY AND VISION SCIENCE, 1996, 73 (01) : 49 - 53
  • [5] Bach M., 2011, OPHTHALMOLOGE, V108, P1195, DOI [10.1007/s00347-011-2488-5, DOI 10.1007/S00347-011-2488-5]
  • [7] BINOCULAR ENHANCEMENT OF VISUAL-ACUITY
    CAGENELLO, R
    ARDITI, A
    HALPERN, DL
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1993, 10 (08): : 1841 - 1848
  • [8] Bilateral implantation of the Acri.LISA bifocal intraocular lens in myopic eyes
    Fernandez-Vega, Luis
    Madrid-Costa, David
    Alfonso, Jose F.
    Poo-Lopez, Arancha
    Montes-Mico, Robert
    [J]. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2010, 20 (01) : 83 - 89
  • [9] Visual and Refractive Outcomes in Hyperopic Pseudophakic Patients Implanted with the Acri.LISA 366D Multifocal Intraocular Lens
    Fernandez-Vega, Luis
    Alfonso, Jose F.
    Baamonde, Begona
    Madrid-Costa, David
    Montes-Mico, Robert
    Lozano, Javier
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 148 (02) : 214 - 220
  • [10] Meta-analysis of accommodating intraocular lenses
    Findl, Oliver
    Leydolt, Christina
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (03) : 522 - 527