A Study of Pullout Forces of the Components of Modular Multi-manufacturer Hybrid Endografts Used for Aortic Aneurysm Repair

被引:16
作者
Cina, D. P. [1 ]
Grant, G. [1 ]
Peterson, M. [2 ]
Campbell, V. [1 ]
Garrido-Olivares, L. [1 ]
Cina, C. S. [1 ,3 ]
机构
[1] Univ Toronto, Div Vasc & Endovasc Surg, St Michaels Hosp, Toronto, ON M5C 1R6, Canada
[2] Univ Toronto, Div Cardiac Surg, St Michaels Hosp, Toronto, ON M5C 1R6, Canada
[3] Univ Toronto, Dept Hlth Policy Management & Evaluat, Toronto, ON M5C 1R6, Canada
关键词
Endovascular aneurysm repair; Endografts; Endostents; ENDOVASCULAR REPAIR; UNITED-STATES; STENT-GRAFTS; OUTCOMES; BLOOD; ENDOLEAK; DECADE;
D O I
10.1016/j.ejvs.2009.02.017
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objectives: Aorto-iliac angulations may be challenging for modular stent-graft systems (SGSs) from a single manufacturer. This study aims to define the pullout forces (POFs) of SGSs derived from the same (non-hybrid) or different manufacturers (hybrid). Methods: The POFs were tested in a vertical position in air and 5% albumin. We studied the POFs between legs from Anaconda (Vascutek (R)), Excluder (Gore (R)), Talent (Medtronic (R)) and Zenith (Cook) with the contralateral limb of bifurcated aortic bodies from Zenith (12 mm), Anaconda and Excluder. Results: For non-hybrid SGSs, the POFs decreased in the following order: Anaconda (11.2 +/- 0.6 N), Talent (6.25 +/- 0.6 N), Zenith (3.5 +/- 0.01 N) and Excluder (2.5 +/- 0.5 N). The Zenith body with the Anaconda limb (15 mm) registered the greatest POF (13.083 +/- 0.821 N); the Zenith and Excluder bodies combined with the Excluder limb (16 mm) registered the weakest POFs (2.397 +/- 0.22 N and 2.500 +/- 0.479 N, respectively). The Zenith body combined with the Excluder limb (16 mm) had a POF similar to the Zenith non-hybrid; combined with Talent 14 mm and Anaconda limb exhibited POFs greater than the Zenith non-hybrid system. For the Limb-to-limb POFs, the greatest was registered for the Anaconda limb, 13 mm within a 12-mm extension for 40-mm overlaps (23.06 +/- 0.480 N); the weakest POFs were recorded for the Excluder limbs at 30-mm overlaps (1.09 +/- 0.167 N and 1.11 +/- 0.250 N). Conclusions: The hybrid SGSs performed as well as or better than the non-hybrid systems, and should be considered for clinical testing in patients whose unique anatomy warrants the flexibility that the use of hybrids provides. (C) 2009 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:671 / 680
页数:10
相关论文
共 29 条
[1]   A statewide experience with endovascular abdominal aortic aneurysm repair: Rapid diffusion with excellent early results - Discussion [J].
Matsumura, JS ;
Anderson, PL ;
Muluk, SC ;
Green, RM ;
Dardik, A ;
Ricotta, JJ .
JOURNAL OF VASCULAR SURGERY, 2004, 39 (01) :18-19
[2]   Long-term outcomes after endovascular abdominal aortic aneurysm repair - The first decade [J].
Brewster, David C. ;
Jones, John E. ;
Chung, Thomas K. ;
Lamuraglia, Glenn M. ;
Kwolek, Christopher J. ;
Watkins, Michael T. ;
Hodgman, Thomas M. ;
Cambria, Richard P. .
ANNALS OF SURGERY, 2006, 244 (03) :426-438
[3]   Malfunctions and Adverse Events Associated With Off-Label Use of Biliary Stents in the Peripheral Vasculature [J].
Bridges, Jonathan ;
Maisel, William H. .
AMERICAN JOURNAL OF THERAPEUTICS, 2008, 15 (01) :12-18
[4]  
Buth Jacob, 2003, Semin Vasc Surg, V16, P95, DOI 10.1016/S0895-7967(03)00007-3
[5]   A decade of change in abdominal aortic aneurysm repair in the United States: Have we improved outcomes equally between men and women? [J].
Dillavou, ED ;
Muluk, SC ;
Makaroun, MS .
JOURNAL OF VASCULAR SURGERY, 2006, 43 (02) :230-236
[6]   Improving aneurysm-related outcomes: Nationwide benefits of endovascular repair [J].
Dillavou, ED ;
Muluk, SC ;
Makaroun, MS .
JOURNAL OF VASCULAR SURGERY, 2006, 43 (03) :446-451
[7]  
GUYTON AC, 2000, TXB MED PHYSL, P355
[8]   In vitro analysis of modular aortic stent-graft failure [J].
Hinchliffe, RJ ;
Natarajan, S ;
Hopkinson, BR .
JOURNAL OF ENDOVASCULAR THERAPY, 2006, 13 (01) :77-84
[9]  
Howell BA, 2007, J ENDOVASC THER, V14, P138, DOI 10.1583/1545-1550(2007)14[138:CFDWBA]2.0.CO
[10]  
2