In vivo accuracy of image guidance performed using optical tracking and optimized registration

被引:44
作者
Mascott, Christopher R. [1 ]
机构
[1] Beacon Hosp, Epilepsy Inst, Beacon Neurosci, Dublin 18, Ireland
关键词
imaging-guided surgery; electrode implantation; depth electrode; frameless stereotaxy;
D O I
10.3171/jns.2006.105.4.561
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. Image guidance systems involving the use of frameless referencing of surgical space to compile volumetric imaging data sets recently have come into widespread use. Few studies have addressed the true intraoperative surgical accuracy (that is the application accuracy) of these systems except in a subjective manner. Calculated accuracies given by the systems do not necessarily reflect true intraoperative accuracy. Methods. To objectively assess the stereotactic accuracy of a frameless image guidance system using optical spatial referencing, the author analyzed postoperative magnetic resonance (MR) images after placement of depth electrodes for the investigation of epilepsy. Preoperative planning for the treatment of seven patients included implanting skull fiducial screws and obtaining computed tomography/MR fusion images by using ImMerge image fusion software on the StealthStation (Medtronic, Inc.). A total of 42 electrodes were placed. Postoperative volumetric MR images were fused with preoperative study images. The difference between the planned electrode trajectories and targets and the visualized electrodes was measured in stereotactic space. Conclusions. The mean distance between the distal electrode contact and the distal end of the planned trajectory for the 42 targets was 3 +/- 1.5 mm. The most common error was in depth. The author's technique did not involve rigid skull fixation of electrodes because they were subsequently tunneled subcutaneously and later removed at the bedside of the patient. Errors in depth were known to be due to traction at the time of tunneling and not due to stereotactic factors. Correcting for depth along the electrode trajectory, the mean accuracy was found to be 2.4 +/- 1 mm.
引用
收藏
页码:561 / 567
页数:7
相关论文
共 30 条
[1]   MAGNETIC-RESONANCE IMAGE-DIRECTED STEREOTAXIC NEUROSURGERY - USE OF IMAGE FUSION WITH COMPUTERIZED-TOMOGRAPHY TO ENHANCE SPATIAL ACCURACY [J].
ALEXANDER, E ;
KOOY, HM ;
VANHERK, M ;
SCHWARTZ, M ;
BARNES, PD ;
TARBELL, N ;
MULKERN, RV ;
HOLUPKA, EJ ;
LOEFFLER, JS .
JOURNAL OF NEUROSURGERY, 1995, 83 (02) :271-276
[2]  
Benabid AL, 1999, REV NEUROL-FRANCE, V155, P869
[3]   Accuracy of MRI-guided stereotactic thalamic functional neurosurgery [J].
Bourgeois, G ;
Magnin, M ;
Morel, A ;
Sartoretti, S ;
Huisman, T ;
Tuncdogan, E ;
Meier, D ;
Jeanmonod, D .
NEURORADIOLOGY, 1999, 41 (09) :636-645
[4]   MRI distortion and stereotactic neurosurgery using the Cosman-Roberts-Wells and Leksell frames [J].
Burchiel, KJ ;
Nguyen, TT ;
Coombs, BD ;
Szumoski, J .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1996, 66 (1-3) :123-136
[5]   EFFECTS OF COREGISTRATION OF MR TO CT IMAGES ON MR STEREOTAXIC ACCURACY [J].
COHEN, DS ;
LUSTGARTEN, JH ;
MILLER, E ;
KHANDJI, AG ;
GOODMAN, RR .
JOURNAL OF NEUROSURGERY, 1995, 82 (05) :772-779
[6]   MRI confirmation of accuracy of freehand placement of mesial temporal lobe depth electrodes in the investigation of intractable epilepsy [J].
Davies, KG ;
Phillips, BLB ;
Hermann, BP .
BRITISH JOURNAL OF NEUROSURGERY, 1996, 10 (02) :175-178
[7]   Accuracy of true frameless stereotaxy: in vivo measurement and laboratory phantom studies - Technical note [J].
Dorward, NL ;
Alberti, O ;
Palmer, JD ;
Kitchen, ND ;
Thomas, DGT .
JOURNAL OF NEUROSURGERY, 1999, 90 (01) :160-168
[8]  
GALLOWAY R L JR, 1991, Biomedical Instrumentation and Technology, V25, P457
[9]   Neurophysiological monitoring for epilepsy surgery: The talairach SEEG method [J].
Guenot, M ;
Isnard, J ;
Ryvlin, P ;
Fischer, C ;
Ostrowsky, K ;
Mauguiere, FO ;
Sindou, M .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2001, 77 (1-4) :29-32
[10]   Frameless stereotaxy using bone fiducial markers for deep brain stimulation [J].
Holloway, KL ;
Gaede, SE ;
Starr, PA ;
Rosenow, JM ;
Ramakrishnan, V ;
Henderson, JM .
JOURNAL OF NEUROSURGERY, 2005, 103 (03) :404-413