An episode of increased hemolysis due to a defective pneumatic air tube delivery system

被引:38
作者
Ellis, Graham [1 ]
机构
[1] St Johns Hosp, Dept Clin Biochem, Livingston W Lothian EH54 6PP, Scotland
关键词
Blood chemical analysis; Blood specimen collection; Humans; Reproducibility of results; Transportation; Hemolysis; Laboratories; Hospital; Specimen handling; Potassium/blood; BLOOD-GAS SAMPLES; TURNAROUND TIME; OXYGEN-TENSION; TRANSPORT; SPECIMENS; INTERFERENCE; BUBBLES; BANK;
D O I
10.1016/j.clinbiochem.2009.05.002
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: To study an episode of excessive sample hemolysis associated with a recently installed pneumatic tube system from the Accident and Emergency Department. Design and methods: Hemolysis rates in samples delivered with a tube system installed 3-4 months previously were followed up retrospectively and prospectively. The laboratory database was searched over a 3-year period for serum potassium results and the number of hemolyzed samples rejected that had been delivered by pneumatic tube or by hand. Results: Before the tube system was installed, the weekly mean hemolysis rate was 3.3%. After installation, the rate was 10.9% for 12 weeks but then it increased to 54%. After a fault was corrected, values fell to 9.0% and to 7.1% after samples were bubble-wrapped. Other wards served as controls. Conclusion: Laboratories should be aware that defects may arise in tube systems that may result in rapid sample deceleration and excessive hemolysis. (C) 2009 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:1265 / 1269
页数:5
相关论文
共 22 条
[11]  
Kratz A, 2007, ARCH PATHOL LAB MED, V131, P293
[12]  
Lu JY, 2003, J FORMOS MED ASSOC, V102, P246
[13]   SENDING BLOOD-GAS SPECIMENS THROUGH PRESSURIZED TRANSPORT TUBE SYSTEMS EXAGGERATES THE ERROR IN OXYGEN-TENSION MEASUREMENTS CREATED BY THE PRESENCE OF AIR BUBBLES [J].
MCKANE, MH ;
SOUTHORN, PA ;
SANTRACH, PJ ;
BURRITT, MF ;
PLEVAK, DJ .
ANESTHESIA AND ANALGESIA, 1995, 81 (01) :179-182
[14]  
POZNANSKI W, 1978, AM J CLIN PATHOL, V70, P291
[15]  
SHEPHERD HT, 2008, ANN CLIN BIOCH S1, V45, P73
[16]   Pneumatic tube system induced haemolysis: assessing sample type susceptibility to haemolysis [J].
Sodi, R ;
Darn, SM ;
Stott, A .
ANNALS OF CLINICAL BIOCHEMISTRY, 2004, 41 :237-240
[17]   HEMOLYSIS OF BLOOD SPECIMENS TRANSPORTED FROM ED TO LABORATORY BY PNEUMATIC TUBE [J].
STAIR, TO ;
HOWELL, JM ;
FITZGERALD, DJ ;
BAILEY, SC ;
BASTASCH, MD .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1995, 13 (04) :484-484
[18]   USE OF A PNEUMATIC TUBE SYSTEM FOR DELIVERY OF BLOOD-BANK PRODUCTS AND SPECIMENS [J].
TANLEY, PC ;
WALLAS, CH ;
ABRAM, MC ;
RICHARDSON, LD .
TRANSFUSION, 1987, 27 (02) :196-198
[19]  
WEAVER DK, 1978, AM J CLIN PATHOL, V70, P400
[20]   Interference in spectrophotometric analysis of cerebrospinal fluid by haemolysis induced by transport through a pneumatic tube system [J].
Wenham, PR ;
Hanson, T ;
Ashby, JP .
ANNALS OF CLINICAL BIOCHEMISTRY, 2001, 38 :371-375