The accuracy and reliability of 3D CT/MRI co-registration in planning epilepsy surgery

被引:44
|
作者
Tao, James X. [1 ]
Hawes-Ebersole, Susan [1 ]
Baldwin, Maria [1 ]
Shah, Sona [1 ]
Erickson, Robert K. [1 ]
Ebersole, John S. [1 ]
机构
[1] Univ Chicago, Dept Neurol & Neurosurg, Adult Epilepsy Ctr, Chicago, IL 60637 USA
关键词
Co-registration; Epilepsy surgery; Subdural electrodes; Temporal lobe epilepsy; Intra-operative digital photography; SUBDURAL ELECTRODES; CT IMAGES; PRESURGICAL EVALUATION; TEMPORAL-LOBE; LOCALIZATION; MRI; COREGISTRATION; RECONSTRUCTION; FUSION;
D O I
10.1016/j.clinph.2009.02.002
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: To investigate the accuracy and reliability of 3D CT/MRI co-registration technique for the localization of implanted subdural electrodes in the routine epilepsy presurgical evaluation, in so doing assess its usefulness in planning the tailored resection of epileptic focus. Methods: Four external anatomic fiducial makers were used for co-registration of volumetric pre-implant brain MRI and post-implant head CT using Curry 5.0 software in 19 epilepsy presurgical candidates. The location of subdural electrodes derived from the co-registration was compared to that obtained by intra-operative digital photographs by using gyral/sulcal patterns and cortical vasculature as anatomic markers. Results: The mean localization error was 4.3 +/- 2.5 mm in all 19 patients. However, the mean localization error was 3.1 +/- 1.3 mm in 13 patients with all four reliable fiducial markers; whereas the mean localization error was 6.8 +/- 2.4 mm in 6 patients with two or three reliable fiducial markers. Conclusion: Visualization of subdural electrode positions on a patient's cortex can be accurately performed in the routine clinical setting by 3D CT/MRI co-registration. However, the accuracy of co-registration is dependent upon having reliable surface fiducial markers. In practice, confirmation of location accuracy, such as with intra-operative digital photographs, is necessary for planning of tailored resective surgery. Significance: The combination of 3D CT/MRI co-registration and intra-operative digital photography techniques provides a practical and effective algorithm for the localization and validation of implanted subdural electrodes. (C) 2009 International Federation of Clinical Neurophysiology. Published by Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:748 / 753
页数:6
相关论文
共 50 条
  • [1] Proposal for a new MEG-MRI co-registration: A 3D laser scanner system
    Hironaga, Naruhito
    Hagiwara, Koichi
    Ogata, Katsuya
    Hayamizu, Mariko
    Urakawa, Tomokazu
    Tobimatsu, Shozo
    CLINICAL NEUROPHYSIOLOGY, 2014, 125 (12) : 2404 - 2412
  • [2] Automatic co-registration of OPM-MEG and MRI using a 3D laser scanner
    Gao, Zhenfeng
    Cao, Fuzhi
    An, Nan
    Ning, Xiaolin
    MEASUREMENT, 2023, 223
  • [3] Combined multimodal co-registration of PET/CT and MRI images increases diagnostic accuracy in squamous cell carcinoma staging
    Stecco, Alessandro
    Ciolfi, Silvio
    Buemi, Francesco
    Cassara, Alessia
    Sacchetti, Gian Mauro
    Brambilla, Marco
    Carriero, Alessandro
    RADIOLOGIA MEDICA, 2016, 121 (06): : 502 - 509
  • [4] MEG-MRI CO-REGISTRATION USING 3D GENERALIZED HOUGH TRANSFORM
    Lin, Sheng-Kai
    Lo, Rong-Chin
    Lee, Ren-Guey
    BIOMEDICAL ENGINEERING-APPLICATIONS BASIS COMMUNICATIONS, 2020, 32 (04):
  • [5] Combined multimodal co-registration of PET/CT and MRI images increases diagnostic accuracy in squamous cell carcinoma staging
    Alessandro Stecco
    Silvio Ciolfi
    Francesco Buemi
    Alessia Cassarà
    Gian Mauro Sacchetti
    Marco Brambilla
    Alessandro Carriero
    La radiologia medica, 2016, 121 : 502 - 509
  • [6] MEG and 3D Digitization Co-registration of Head Using 3D Hough Transform
    Lin, Sheng Kai
    Lo, Rong Chin
    INTERNATIONAL CONFERENCE ON BIOLOGICAL, MEDICAL AND CHEMICAL ENGINEERING (BMCE 2013), 2013, : 653 - 656
  • [7] AUTOMATIC CO-REGISTRATION OF MEG AND 3D DIGITIZATION USING 3D GENERALIZED HOUGH TRANSFORM
    Lin, Sheng-Kai
    Lo, Rong-Chin
    Lee, Ren-Guey
    BIOMEDICAL ENGINEERING-APPLICATIONS BASIS COMMUNICATIONS, 2020, 32 (03):
  • [8] Imaging of prostate cancer: a platform for 3D co-registration of in-vivo MRI ex-vivo MRI and pathology
    Orczyk, Clement
    Mikheev, Artem
    Rosenkrantz, Andrew
    Melamed, Jonathan
    Taneja, Samir S.
    Rusinek, Henry
    MEDICAL IMAGING 2012: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2012, 8316
  • [9] Automatic co-registration of 3D multi-sensor point clouds
    Persad, Ravi Ancil
    Armenakis, Costas
    ISPRS JOURNAL OF PHOTOGRAMMETRY AND REMOTE SENSING, 2017, 130 : 162 - 186
  • [10] Co-registration of FLAIR- and T1-weighted MRI for image guided extended lesionectomy in epilepsy surgery
    Mahvash, M
    König, R
    Urbach, H
    von Oertzen, J
    Meyer, B
    Schaller, C
    CARS 2004: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2004, 1268 : 1283 - 1283