摘要
Purpose The goal of this study was to find the optimal combination of collimator, photopeak and scatter correction for(177)Lutetium (Lu-177) SPECT/CT imaging. Methods Three experiments [sphere-to-background ratios (SBR) 50:1, 10:1, and 2:1] were performed with the NEMA Image Quality phantom filled with(177)Lu-trichloride. SPECT/CT acquisitions were performed with the medium-energy low-penetration (MELP) collimator and(99m)Tc/Krypton collimator. For each acquisition six reconstructions, all with attenuation correction (AC), were made: the 113-keV photopeak only, the 208-keV photopeak only and both photopeaks combined, each with or without scatter correction (SC). Image quality was assessed using contrast-to-noise ratios (CNR), quantification accuracy by means of recovery coefficients (RCs) and the spatial resolution using line profiles. Results With SBR 50:1 and 10:1, both collimators met theRose criterion(CNR > 5), whereas the MELP collimator showed a higher CNR for the 2:1 ratio. The RC(mean)was higher with the MELP collimator, most explicit after the 208-keV AC/SC reconstruction for all acquisitions. The line profiles showed a better spatial resolution for the MELP collimator and the 208-keV AC/SC reconstructions. Conclusion Lu-177 SPECT/CT image quality and quantification was most optimal when acquired with the MELP collimator and reconstructed using the 208-keV photopeak, with AC and SC.
引用
收藏
页码:272 / 277
页数:6
相关论文
共 17 条
[1]   A phantom study: Should 124I-mIBG PET/CT replace 123I-mIBG SPECT/CT? [J].
Beijst, Casper ;
de Keizer, Bart ;
Lam, Marnix G. E. H. ;
Janssens, Geert O. ;
Tytgat, Godelieve A. M. ;
de Jong, Hugo W. A. M. .
MEDICAL PHYSICS, 2017, 44 (05) :1624-1631
[2]  
Cherry S R., 2012, Physics in Nuclear Medicine, V4th, P43
[3]   The conflict between treatment optimization and registration of radiopharmaceuticals with fixed activity posology in oncological nuclear medicine therapy [J].
Chiesa, C. ;
Gleisner, K. Sjogreen ;
Flux, G. ;
Gear, J. ;
Walrand, S. ;
Bacher, K. ;
Eberlein, U. ;
Visser, E. P. ;
Chouin, N. ;
Ljungberg, M. ;
Bardies, M. ;
Lassmann, M. ;
Strigari, L. ;
Konijnenberg, M. W. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 (11) :1783-1786
[4]   Improving quantitative dosimetry in 177 Lu-DOTATATE SPECT by energy window-based scatter corrections [J].
de Nijs, Robin ;
Lagerburg, Vera ;
Klausen, Thomas L. ;
Holm, Soren .
NUCLEAR MEDICINE COMMUNICATIONS, 2014, 35 (05) :522-533
[5]  
Dewaraja YK, 2012, J NUCL MED, V53, P1310, DOI 10.2967/jnumed.111.100123
[6]   177Lu-PSMA Radioligand Therapy for Prostate Cancer [J].
Fendler, Wolfgang P. ;
Rahbar, Kambiz ;
Herrmann, Ken ;
Kratochwil, Clemens ;
Eiber, Matthias .
JOURNAL OF NUCLEAR MEDICINE, 2017, 58 (08) :1196-1200
[7]   Partial volume effect of SPECT images in PRRT with 177Lu labelled somatostatin analogues: A practical solution [J].
Finocchiaro, Domenico ;
Berenato, Salvatore ;
Grassi, Elisa ;
Bertolini, Valentina ;
Castellani, Gastone ;
Lanconelli, Nico ;
Versari, Annibale ;
Spezi, Emiliano ;
Iori, Mauro ;
Fioroni, Federica .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 57 :153-159
[8]   177Lu-[DOTA0,Tyr3] Octreotate Therapy in Patients With Disseminated Neuroendocrine Tumors: Analysis of Dosimetry With Impact on Future Therapeutic Strategy [J].
Garkavij, Michael ;
Nickel, Mattias ;
Sjogreen-Gleisner, Katarina ;
Ljungberg, Michael ;
Ohlsson, Tomas ;
Wingardh, Karin ;
Strand, Sven-Erik ;
Tennvall, Jan .
CANCER, 2010, 116 (04) :1084-1092
[9]   The Relevance of Dosimetry in Precision Medicine [J].
Lassmann, Michael ;
Eberlein, Uta .
JOURNAL OF NUCLEAR MEDICINE, 2018, 59 (10) :1494-1499
[10]   MIRD Pamphlet No. 26: Joint EANM/MIRD Guidelines for Quantitative 177Lu SPECT Applied for Dosimetry of Radiopharmaceutical Therapy [J].
Ljungberg, Michael ;
Celler, Anna ;
Konijnenberg, Mark W. ;
Eckerman, Keith F. ;
Dewaraja, Yuni K. ;
Sjogreen-Gleisner, Katarina ;
Bolch, Wesley E. ;
Brill, A. Bertrand ;
Fahey, Frederic ;
Fisher, Darrell R. ;
Hobbs, Robert ;
Howell, Roger W. ;
Meredith, Ruby F. ;
Sgouros, George ;
Zanzonico, Pat ;
Bacher, Klaus ;
Chiesa, Carlo ;
Flux, Glenn ;
Lassmann, Michael ;
Strigari, Lidia ;
Walrand, Stephan .
JOURNAL OF NUCLEAR MEDICINE, 2016, 57 (01) :151-162