An intercomparison of 11 amorphous silicon EPIDs of the same type: implications for portal dosimetry

被引:27
作者
Winkler, Peter [1 ]
Georg, Dietmar [1 ]
机构
[1] Med Univ Vienna, Dept Radiotherapy & Radiobiol, Div Med Radiat Phys, A-1090 Vienna, Austria
关键词
D O I
10.1088/0031-9155/51/17/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of electronic portal imaging devices (EPIDs) for portal dosimetry requires knowledge of their dosimetric properties. The pixel value response of amorphous silicon EPIDs of type Elekta iViewGT (TM) is known to be nonlinear with dose. However, it is not clear whether these nonlinearities vary with time and from one detector to another, respectively. In the present study, the dose response characteristics of 11 iViewGT EPIDs were investigated with respect to dose rate, total dose and field size. It was found that each detector needs to be individually calibrated, not only in terms of absolute sensitivity but also with respect to its relative response variations with exposure parameters. Doubling the dose rate typically increased the EPID signal between 1.4% and 2.8%. Changing the number of monitor units from 30 to 500 was accompanied by an increase in detector sensitivity between 1.7% and 2.8%. The EPID scatter factors were always within +/- 1%. It was observed that the dose-response behaviour was not stable with respect to time. Particularly within the first weeks of operation, detector ageing caused variations in both absolute sensitivity and relative response curves. It is recommended to establish a quality assurance programme if the amorphous silicon EPIDs are intended to be used for clinical portal dosimetry.
引用
收藏
页码:4189 / 4200
页数:12
相关论文
共 18 条
[11]   Three-dimensional IMRT verification with a flat-panel EPID [J].
Steciw, S ;
Warkentin, B ;
Rathee, S ;
Fallone, BG .
MEDICAL PHYSICS, 2005, 32 (02) :600-612
[12]   Beam-characteristics during beam start-up: Minimum number of MU for IMRT [J].
Steenhuijsen, J .
RADIOTHERAPY AND ONCOLOGY, 2005, 76 :S180-S180
[13]   THE DETERMINATION OF PHANTOM AND COLLIMATOR SCATTER COMPONENTS OF THE OUTPUT OF MEGAVOLTAGE PHOTON BEAMS - MEASUREMENT OF THE COLLIMATOR SCATTER PART WITH A BEAM-COAXIAL NARROW CYLINDRICAL PHANTOM [J].
VANGASTEREN, JJM ;
HEUKELOM, S ;
VANKLEFFENS, HJ ;
VANDERLAARSE, R ;
VENSELAAR, JLM ;
WESTERMANN, CF .
RADIOTHERAPY AND ONCOLOGY, 1991, 20 (04) :250-257
[14]   Dosimetric IMRT verification with a flat-panel EPID [J].
Warkentin, B ;
Steciw, S ;
Rathee, S ;
Fallone, BG .
MEDICAL PHYSICS, 2003, 30 (12) :3143-3155
[15]   Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method [J].
Wendling, M ;
Louwe, RJW ;
McDermott, LN ;
Sonke, JJ ;
van Herk, M ;
Mijnheer, BJ .
MEDICAL PHYSICS, 2006, 33 (02) :259-273
[16]   Dose-response characteristics of an amorphous silicon EPID [J].
Winkler, P ;
Hefner, A ;
Georg, D .
MEDICAL PHYSICS, 2005, 32 (10) :3095-3105
[17]  
WINKLER P, 2006, P 9 INT WORKSH EL PO
[18]   Quantitative measurement of MLC leaf displacements using an electronic portal image device [J].
Yang, Y ;
Xing, L .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (08) :1521-1533