Utility of porous graphitic carbon stationary phase in quantitative liquid chromatography/tandem mass spectrometry bioanalysis: quantitation of diastereomers in plasma

被引:21
作者
Xia, Yuan-Qing
Jemal, Mohammed
Zheng, Naiyu
Shen, Xiaohang
机构
[1] Bristol Myers Squibb Co, Pharmaceut Res Inst, Bioanalyt & Discovery Analyt Sci, Princeton, NJ 08543 USA
[2] Bristol Myers Squibb Co, Pharmaceut Res Inst, Bioanalyt & Discovery Analyt Sci, New Brunswick, NJ 08903 USA
关键词
D O I
10.1002/rcm.2517
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A major challenge in selecting an appropriate stationary phase for diastereomeric separation is that it is difficult to predict which of the commercially available stationary phases could achieve the required liquid chromatographic (LC) separation. This work describes the selection and evaluation of a porous graphitic carbon (PGC) column coupled with tandem mass spectrometry (MS/MS) for the simultaneous quantitation of an experimental drug candidate (I), its two diastereomeric metabolites (II and III), and its demethylated metabolite (IV) in rat plasma. In addition, we investigated the PGC column for the separation of another drug candidate (VD, its two diastereomeric metabolites (VII and VIII) and its ketone metabolite (IX). The PGC column showed excellent chromatographic resolution for the two diastereomers II and 111, as well as for VII and VIII. In contrast, the required resolution for the diastereomers 11 and III could not be achieved using silica-bonded C-18, C-30, phenyl, perfluorinated, polar embedded and polar end-capped phases. The PGC column showed ruggedness with excellent reproducibility of retention times, peak symmetry and response over a period of more than 400 injections of a plasma acetonitrile-precipitation extract. Excellent accuracy and precision were achieved, with accuracy of 94-108% and intra- and inter-run precision within 9%. This work indicates that PGC is a valuable addition to the repertoire of LC columns used for quantitative LC/MS/MS bioanalysis, especially where the separation and quantitation of diastereomeric analytes is involved. Copyright (c) 2006 John Wiley & Sons, Ltd.
引用
收藏
页码:1831 / 1837
页数:7
相关论文
共 23 条
  • [1] POROUS GRAPHITIC CARBON SHOWS PROMISE FOR THE RAPID CHROMATOGRAPHIC ANALYSIS OF POLAR DRUG METABOLITES
    AYRTON, J
    EVANS, MB
    HARRIS, AJ
    PLUMB, RS
    [J]. JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1995, 667 (01): : 173 - 178
  • [2] High-throughput mass spectrometric analysis of xenobiotics in biological fluids
    Bakhtiar, R
    Ramos, L
    Tse, FLS
    [J]. JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2002, 25 (04) : 507 - 540
  • [3] Porous graphitic carbon chromatography-tandem mass spectrometry for the study of isoprostanes in human cerebrospinal fluid
    Bohnstedt, KC
    Karlberg, B
    Basun, H
    Schmidt, S
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 827 (01): : 39 - 43
  • [4] Determination of isoprostanes in urine samples from Alzheimer patients using porous graphitic carbon liquid chromatography-tandem mass spectrometry
    Bohnstedt, KC
    Karlberg, B
    Wahlund, LO
    Jönhagen, ME
    Basun, H
    Schmidt, S
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 796 (01): : 11 - 19
  • [5] Systematic LC/MS metabolite identification in drug discovery
    Clarke, N
    Rindgen, D
    Korfmacher, W
    Cox, K
    [J]. ANALYTICAL CHEMISTRY, 2001, 73 (15) : 430A - 439A
  • [6] Inamoto Y, 1998, BIOMED CHROMATOGR, V12, P239, DOI 10.1002/(SICI)1099-0801(199809/10)12:5<239::AID-BMC806>3.0.CO
  • [7] 2-X
  • [8] Liquid chromatography of guanidino compounds using a porous graphite carbon column and application to their analysis in serum
    Inamoto, Y
    Inamoto, S
    Hanai, T
    Tokuda, M
    Hatase, O
    Yoshii, K
    Sugiyama, N
    Kinoshita, T
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1998, 707 (1-2): : 111 - 120
  • [9] JEMAL M, 2006, IN PRESS CURR DRUG M, V7
  • [10] Utilisation of reversed-phase high-performance liquid chromatography as an alternative to silver-ion chromatography for the separation of cis- and trans-C18:1 fatty acid isomers
    Juanéda, P