Developments in and experience of kilovoltage X-ray cone beam image-guided radiotherapy

被引:39
作者
Moore, C. J. [1 ]
Amer, A.
Marchant, T.
Sykes, J. R.
Davies, J.
Stratford, J.
McCarthy, C.
MacBain, C.
Henry, A.
Price, P.
Williams, P. C.
机构
[1] Christie Hosp, N Western Med Phys Dept, Manchester M20 4BX, Lancs, England
[2] Christie Hosp, Radiotherapy Res Facil, Manchester M20 4BX, Lancs, England
[3] Christie Hosp, Acad Dept Radiat Oncol, Manchester M20 4BX, Lancs, England
关键词
D O I
10.1259/bjr/68255935
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper offers a realistic review of kilovoltage X-ray cone beam tomography integrated with the treatment machine for image-guided radiotherapy in the light of experience taking a commercial system from prototype development into clinical use. It shows that key practicalities cannot be ignored, in particular the regular characterization of mechanical flex during gantry rotation, the mapping of defects in flat panel image transducers and their response to X-ray exposure. The number of X-ray projections and the doses required for clinically useful cone beam reconstruction at different therapy sites are considered in the context of imaging that is fit for purpose. Three roles for cone beam tomography in radiotherapy are identified: patient setup in three dimensions (3D), where even low dose cone beam tissue detail is superior to megavoltage imaging; disease targeting where, despite wide field scatter and slow scanning, it is possible to generate images that are suitable for tumour delineation even at challenging sites; adaptive treatment planning, where calibrated cone beam images have been shown to provide sufficient target detail to support "plan of the day" selection and have the potential for planning with bulk corrections. With frequent use in mind, the need to limit patient dose during setup, yet maximize much needed image quality in the target zone, is considered. Finally, it is noted that the development of cone beam tomography for radiotherapy is far from complete, with X-ray source, image transducer, reconstruction algorithms and techniques for image profile collection still being researched. (C) 2006 The British Institute of Radiology.
引用
收藏
页码:S66 / S78
页数:13
相关论文
共 28 条
[1]   Second cancer risk, concomitant exposures and IRMER(2000) [J].
Aird, EGA .
BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (924) :983-985
[2]  
Amer A, 2004, RADIOTHER ONCOL, V73, pS230
[3]  
Amer A, 2004, IEEE INFOR VIS, P92
[4]  
AMER A, 2004, RADIOTHER ONCOL S1, V73
[5]  
AMER A, UK RAD ONC C 2005 11
[6]   Initial performance evaluation of an indirect-detection, active matrix flat-panel imager (AMFPI) prototype for megavoltage imaging [J].
Antonuk, LE ;
El-Mohri, Y ;
Huang, WD ;
Jee, KW ;
Siewerdsen, JH ;
Maolinbay, M ;
Scarpine, VE ;
Sandler, H ;
Yorkston, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (02) :437-454
[7]   Electronic portal imaging devices: a review and historical perspective of contemporary technologies and research [J].
Antonuk, LE .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (06) :R31-R65
[8]   Accurate technique for complete geometric calibration of cone-beam computed tomography systems [J].
Cho, YB ;
Moseley, DJ ;
Siewerdsen, JH ;
Jaffray, DA .
MEDICAL PHYSICS, 2005, 32 (04) :968-983
[9]  
*EC, 1998, 1626 EC EUR
[10]  
*EN, 1998, 6060121 EN IEC