Highly Undersampled Contrast-Enhanced MRA with Iterative Reconstruction: Integration in a Clinical Setting

被引:43
作者
Stalder, Aurelien F. [1 ]
Schmidt, Michaela [1 ]
Quick, Harald H. [2 ,3 ]
Schlamann, Marc [4 ]
Maderwald, Stefan [2 ]
Schmitt, Peter [1 ]
Wang, Qiu [5 ]
Nadar, Mariappan S. [5 ]
Zenge, Michael O. [1 ]
机构
[1] Siemens Healthcare, D-91052 Erlangen, Germany
[2] Univ Duisburg Essen, Erwin L Hahn Inst MR Imaging, Duisburg, Germany
[3] Univ Hosp Essen, High Field & Hybrid MR Imaging, Essen, Germany
[4] Univ Hosp Essen, Dept Diagnost & Intervent Radiol & Neuroradiol, Essen, Germany
[5] Siemens Corp Technol, Princeton, NJ USA
关键词
sparse data sampling; iterative reconstruction; sparse MRI; contrast-enhanced MRA; intracranial MRA; MAGNETIC-RESONANCE ANGIOGRAPHY; CAROTID-ARTERY STENOSIS; DIGITAL-SUBTRACTION-ANGIOGRAPHY; ROTATIONAL ANGIOGRAPHY; INTRACRANIAL ANEURYSMS; DISEASE; STROKE; ABNORMALITIES; METAANALYSIS; EXPERIENCE;
D O I
10.1002/mrm.25565
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To integrate, optimize, and evaluate a three-dimensional (3D) contrast-enhanced sparse MRA technique with iterative reconstruction on a standard clinical MR system. Methods: Data were acquired using a highly undersampled Cartesian spiral phyllotaxis sampling pattern and reconstructed directly on the MR system with an iterative SENSE technique. Undersampling, regularization, and number of iterations of the reconstruction were optimized and validated based on phantom experiments and patient data. Sparse MRA of the whole head (field of view: 265 x 232 x 179 mm 3) was investigated in 10 patient examinations. Results: High-quality images with 30-fold undersampling, resulting in 0.7 mm isotropic resolution within 10 s acquisition, were obtained. After optimization of the regularization factor and of the number of iterations of the reconstruction, it was possible to reconstruct images with excellent quality within six minutes per 3D volume. Initial results of sparse contrast-enhanced MRA (CEMRA) in 10 patients demonstrated high-quality whole-head first-pass MRA for both the arterial and venous contrast phases. Conclusion: While sparse MRI techniques have not yet reached clinical routine, this study demonstrates the technical feasibility of high-quality sparse CEMRA of the whole head in a clinical setting. Sparse CEMRA has the potential to become a viable alternative where conventional CEMRA is too slow or does not provide sufficient spatial resolution. (C) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:1652 / 1660
页数:9
相关论文
共 48 条
[1]  
Adcock B, ARXIV201313020561
[2]   Carotid artery stenosis: Intraindividual correlations of 3D time-of-flight MR angiography, contrast-enhanced MR angiography, conventional DSA, and rotational angiography for detection and grading [J].
Anzalone, N ;
Scomazzoni, F ;
Castellano, R ;
Strada, L ;
Righi, C ;
Politi, LS ;
Kirchin, MA ;
Chiesa, R ;
Scotti, G .
RADIOLOGY, 2005, 236 (01) :204-213
[3]   3D Reconstruction of Histological Sections: Application to Mammary Gland Tissue [J].
Arganda-Carreras, Ignacio ;
Fernandez-Gonzalez, Rodrigo ;
Munoz-Barrutia, Arrate ;
Ortiz-De-Solorzano, Carlos .
MICROSCOPY RESEARCH AND TECHNIQUE, 2010, 73 (11) :1019-1029
[4]   A Fast Iterative Shrinkage-Thresholding Algorithm for Linear Inverse Problems [J].
Beck, Amir ;
Teboulle, Marc .
SIAM JOURNAL ON IMAGING SCIENCES, 2009, 2 (01) :183-202
[5]  
Bernstein MA, 2001, MAGNET RESON MED, V46, P955, DOI 10.1002/mrm.1282
[6]  
Borisch I, AJNR AM J NEURORADIO, V24, P1117
[7]   Whole-Heart Imaging Using Undersampled Radial Phase Encoding (RPE) and Iterative Sensitivity Encoding (SENSE) Reconstruction [J].
Boubertakh, R. ;
Prieto, C. ;
Batchelor, P. G. ;
Uribe, S. ;
Atkinson, D. ;
Eggers, H. ;
Sorensen, T. S. ;
Hansen, M. S. ;
Razavi, R. S. ;
Schaeffter, T. .
MAGNETIC RESONANCE IN MEDICINE, 2009, 62 (05) :1331-1337
[8]   Practical noninvasive neurovascular imaging of the neck arteries in patients with stroke, transient ischemic attack, and suspected arterial disease that may lead to ischemia, infarction, or flow abnormalities [J].
Brobeck, BR ;
Forero, NP ;
Romero, JM .
SEMINARS IN ULTRASOUND CT AND MRI, 2006, 27 (03) :177-193
[9]  
Bui Trung D, 2005, Vasc Endovascular Surg, V39, P481, DOI 10.1177/153857440503900604
[10]   Quantification of paper mass distributions within local picking areas [J].
Chinga, Gary ;
Syverud, Kristin .
NORDIC PULP & PAPER RESEARCH JOURNAL, 2007, 22 (04) :441-446