Dual-feedback mid-infrared cavity-enhanced absorption spectroscopy for H2CO detection using a radio-frequency electrically-modulated interband cascade laser

被引:32
作者
He, Qixin [1 ,2 ]
Zheng, Chuantao [1 ,2 ]
Lou, Minhan [2 ]
Ye, Weilin [2 ,3 ]
Wang, Yiding [1 ]
Tittel, Frank K. [2 ]
机构
[1] Jilin Univ, Coll Elect Sci & Engn, State Key Lab Integrated Optoelect, 2699 Qianjin St, Changchun 130012, Jilin, Peoples R China
[2] Rice Univ, Dept Elect & Comp Engn, 6100 Main St, Houston, TX 77005 USA
[3] Shantou Univ, Coll Engn, 243 Daxue Rd, Shantou 515063, Peoples R China
基金
美国国家科学基金会; 中国国家自然科学基金;
关键词
FREQUENCY STABILIZATION; DIODE-LASER; FORMALDEHYDE DETECTION; AMBIENT-AIR; SPECTROMETER; TEMPERATURE; LOCKING; SNO2;
D O I
10.1364/OE.26.015436
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A mid-infrared cavity-enhanced sensor system was demonstrated for the detection of formaldehyde (H2CO) using a continuous-wave (cw) interband cascade laser (ICL) centered at 3599 nm. A compact Fabry-Perot (F-P) cavity with a physical size of 38 x 52 x 76 mm(3) was developed consisting of two concave mirrors with a radius of curvature of 80 mm and a reflectivity of 99.8% at 3.6 mu m. Different from the widely reported electro-optical (EO) external modulation based Pound-Drever-Hall (PDH) locking technique, a radio-frequency electrical internal modulation based PDH technique was used for locking the laser mode to the cavity mode. A dual-feedback control on the laser current and on the piezo transducer (PZT) displacement was utilized for further stabilizing mode locking. A 20 m effective optical path length was achieved with a cavity length of 2 cm and a finesse of 1572. The effectiveness and sensitivity of the sensor system were demonstrated by targeting an absorption line at 2778.5 cm(-1) for H2CO measurements. A linear relation between the cavity transmitted signal amplitude and the H2CO concentration was obtained within the range of 0-5 ppm. A 1 sigma detection limit of 25 parts-per-billion (ppb) was achieved with an averaging time of 1 s based on Allan-Werle variance analysis. The reported dual-feedback RF modulation based PDH technique led to a method for gas detection using a similar experimental setup and measurement scheme. (C) 2018 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
引用
收藏
页码:15436 / 15444
页数:9
相关论文
共 27 条
[1]   Optical feedback cavity enhanced absorption spectroscopy with diode lasers [J].
Baran, Stuart G. ;
Hancock, Gus ;
Peverall, Robert ;
Ritchie, Grant A. D. ;
van Leeuwen, Nicola J. .
ANALYST, 2009, 134 (02) :243-249
[2]   An introduction to Pound-Drever-Hall laser frequency stabilization [J].
Black, ED .
AMERICAN JOURNAL OF PHYSICS, 2001, 69 (01) :79-87
[3]   Real-time multiplexed digital cavity-enhanced spectroscopy [J].
Boyson, Toby K. ;
Dagdigian, Paul J. ;
Pavey, Karl D. ;
Fitzgerald, Nicholas J. ;
Spence, Thomas G. ;
Moore, David S. ;
Harb, Charles C. .
OPTICS LETTERS, 2015, 40 (19) :4560-4562
[4]   Simple piezoelectric-actuated mirror with 180 kHz servo bandwidth [J].
Briles, Travis C. ;
Yost, Dylan C. ;
Cingoz, Arman ;
Ye, Jun ;
Schibli, Thomas R. .
OPTICS EXPRESS, 2010, 18 (10) :9739-9746
[5]   PRECISE RF TUNING FOR CW DYE-LASERS [J].
BURGHARDT, B ;
JITSCHIN, W ;
MEISEL, G .
APPLIED PHYSICS, 1979, 20 (02) :141-146
[6]   High-performance midinfrared quantum cascade lasers [J].
Capasso, Federico .
OPTICAL ENGINEERING, 2010, 49 (11)
[7]   Rapid formaldehyde monitoring in ambient air by means of mid-infrared cavity leak-out spectroscopy [J].
Dahnke, H ;
Von Basum, G ;
Kleinermanns, K ;
Hering, P ;
Mürtz, M .
APPLIED PHYSICS B-LASERS AND OPTICS, 2002, 75 (2-3) :311-316
[8]   Ppb-level formaldehyde detection using a CW room-temperature interband cascade laser and a miniature dense pattern multipass gas cell [J].
Dong, Lei ;
Yu, Yajun ;
Li, Chunguang ;
So, Stephen ;
Tittel, Frank K. .
OPTICS EXPRESS, 2015, 23 (15) :19821-19830
[9]   LASER PHASE AND FREQUENCY STABILIZATION USING AN OPTICAL-RESONATOR [J].
DREVER, RWP ;
HALL, JL ;
KOWALSKI, FV ;
HOUGH, J ;
FORD, GM ;
MUNLEY, AJ ;
WARD, H .
APPLIED PHYSICS B-PHOTOPHYSICS AND LASER CHEMISTRY, 1983, 31 (02) :97-105
[10]   Cavity enhanced absorption spectroscopy using a broadband prism cavity and a supercontinuum source [J].
Johnston, Paul S. ;
Lehmann, Kevin K. .