Dosimetric properties of radiophotoluminescent glass rod detector in high-energy photon beams from a linear accelerator and Cyber-Knife

被引:80
|
作者
Araki, F [1 ]
Moribe, N
Shimonobou, T
Yamashita, Y
机构
[1] Kumamoto Univ, Dept Radiol Technol, Sch Hlth Sci, Kumamoto 8620976, Japan
[2] Kumamoto Univ Hosp, Dept Radiol, Kumamoto 8620976, Japan
关键词
radiophotoluminescent glass rod dosimeter; dosimetry; output factors; small field measurements;
D O I
10.1118/1.1758351
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A fully automatic radiophotoluminescent glass rod dosimeter (GRD) system has recently become commercially available. This article discusses the dosimetric properties of the GRD including uniformity and reproducibility of signal, dose linearity, and energy and directional dependence in high-energy photon beams. In addition, energy response is measured in electron beams. The uniformity and reproducibility of the signal from 50 GRDs using a Co-60 beam are both +/-1.1% (one standard deviation). Good dose linearity of the GRD is maintained for doses ranging from 0.5 to 30 Gy, the lower and upper limits of this study, respectively. The GRD response is found to show little Co-60 beam, 4 MV (TPR1020 = 0.617) and 10 MV energy dependence in photon energies of a 10 (TPR1020 = 0.744) x-ray beams. However, the GRD responses for 9 MeV (mean energy, (E) over bar (z) = 3.6 MeV) and 16 MeV ((E) over bar (z) = 10.4 MeV) electron beams are 4%-5% lower than that for a Co-60 beam in the beam quality dependence. The measured angular dependence of GRD, ranging from 0degrees (along the long axis of GRD) to 120degrees is within 1.5% for a 4 MV x-ray beam. As applications, a linear accelerator-based radiosurgery system and Cyber-Knife output factors are measured by a GRD and compared with those from various detectors including a p-type silicon diode detector, a diamond detector, and an ion chamber. It is found that the GRD is a very useful detector for small field dosimetry, in particular, below 10 mm circular fields. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:1980 / 1986
页数:7
相关论文
共 50 条
  • [1] Dosimetric properties of radiophotoluminescent glass detector in low-energy photon beams
    Kadoya, Noriyuki
    Shimomura, Kouhei
    Kitou, Satoshi
    Shiota, Yasuo
    Fujita, Yukio
    Dobashi, Suguru
    Takeda, Ken
    Jingu, Keiichi
    Matsushita, Haruo
    Namito, Yoshihito
    Ban, Syuichi
    Koyama, Syuji
    Tabushi, Katsuyoshi
    MEDICAL PHYSICS, 2012, 39 (10) : 5910 - 5916
  • [2] Investigation of the dosimetric characteristics of radiophotoluminescent glass dosimeter for high-energy photon beams
    Oonsiri, Puntiwa
    Kingkaew, Sakda
    Vannavijit, Chulee
    Suriyapee, Sivalee
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2019, 12 (01) : 65 - 71
  • [3] DOSIMETRIC PROPERTIES OF RADIOPHOTOLUMINESCENT GLASS DOSIMETER IN HIGH-ENERGY ELECTRON BEAMS
    Son, K.
    Jung, H.
    Shin, S. H.
    Lee, H. H.
    Kim, M. S.
    Ji, Y. H.
    Kim, B.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S407 - S407
  • [4] Evaluation of the dosimetric characteristics of a radiophotoluminescent glass dosimeter for high-energy photon and electron beams in the field of radiotherapy
    Son, Kihong
    Jung, Haijo
    Shin, Sang Hun
    Lee, Hyun-Ho
    Kim, Mi-Sook
    Ji, Young Hoon
    Kim, Kum Bae
    RADIATION MEASUREMENTS, 2011, 46 (10) : 1117 - 1122
  • [5] Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams
    Ravichandran, Ramamoorthy
    Binukumar, John Pichy
    Al Amri, Iqbal
    Davis, Cheriyathmanjiyil Antony
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 291 - 303
  • [6] Beam quality of high-energy photon beams at the Ghent University linear accelerator
    Palmans, H
    Mondelaers, W
    Thierens, H
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (17): : L15 - L18
  • [7] DOSIMETRIC ASPECTS OF THE THERAPEUTIC PHOTON BEAMS FROM A DUAL-ENERGY LINEAR-ACCELERATOR
    ALGHAZI, MSAL
    ARJUNE, B
    FIEDLER, JA
    SHARMA, PD
    MEDICAL PHYSICS, 1988, 15 (02) : 250 - 257
  • [8] Dosimetric Properties of a New Solid Water High Equivalency Phantom for High-Energy Photon Beams
    Araki, F.
    Ohno, T.
    Onitsuka, R.
    Shimohigashi, Y.
    MEDICAL PHYSICS, 2015, 42 (06) : 3410 - 3410
  • [9] A diamond detector in the dosimetry of high-energy electron and photon beams
    Laub, WU
    Kaulich, TW
    Nüsslin, F
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (09): : 2183 - 2192
  • [10] DOSIMETRY OF HIGH-ENERGY ELECTRON-BEAMS FROM A LINEAR-ACCELERATOR
    BRIOT, E
    DUTREIX, A
    JOURNAL DE RADIOLOGIE D ELECTROLOGIE ET DE MEDECINE NUCLEAIRE, 1976, 57 (05): : 447 - 454