Dosimetric properties of radiophotoluminescent glass rod detector in high-energy photon beams from a linear accelerator and Cyber-Knife

被引:81
作者
Araki, F [1 ]
Moribe, N
Shimonobou, T
Yamashita, Y
机构
[1] Kumamoto Univ, Dept Radiol Technol, Sch Hlth Sci, Kumamoto 8620976, Japan
[2] Kumamoto Univ Hosp, Dept Radiol, Kumamoto 8620976, Japan
关键词
radiophotoluminescent glass rod dosimeter; dosimetry; output factors; small field measurements;
D O I
10.1118/1.1758351
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A fully automatic radiophotoluminescent glass rod dosimeter (GRD) system has recently become commercially available. This article discusses the dosimetric properties of the GRD including uniformity and reproducibility of signal, dose linearity, and energy and directional dependence in high-energy photon beams. In addition, energy response is measured in electron beams. The uniformity and reproducibility of the signal from 50 GRDs using a Co-60 beam are both +/-1.1% (one standard deviation). Good dose linearity of the GRD is maintained for doses ranging from 0.5 to 30 Gy, the lower and upper limits of this study, respectively. The GRD response is found to show little Co-60 beam, 4 MV (TPR1020 = 0.617) and 10 MV energy dependence in photon energies of a 10 (TPR1020 = 0.744) x-ray beams. However, the GRD responses for 9 MeV (mean energy, (E) over bar (z) = 3.6 MeV) and 16 MeV ((E) over bar (z) = 10.4 MeV) electron beams are 4%-5% lower than that for a Co-60 beam in the beam quality dependence. The measured angular dependence of GRD, ranging from 0degrees (along the long axis of GRD) to 120degrees is within 1.5% for a 4 MV x-ray beam. As applications, a linear accelerator-based radiosurgery system and Cyber-Knife output factors are measured by a GRD and compared with those from various detectors including a p-type silicon diode detector, a diamond detector, and an ion chamber. It is found that the GRD is a very useful detector for small field dosimetry, in particular, below 10 mm circular fields. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:1980 / 1986
页数:7
相关论文
共 19 条
  • [1] AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
    Almond, PR
    Biggs, PJ
    Coursey, BM
    Hanson, WF
    Huq, MS
    Nath, R
    Rogers, DWO
    [J]. MEDICAL PHYSICS, 1999, 26 (09) : 1847 - 1870
  • [2] [Anonymous], 1983, MED PHYS, V10, P741
  • [3] Measurements of Gamma-Knife helmet output factors using a radiophotoluminescent glass rod dosimeter and a diode detector
    Araki, F
    Ikegami, T
    Ishidoya, T
    Kubo, HD
    [J]. MEDICAL PHYSICS, 2003, 30 (08) : 1976 - 1981
  • [4] *AS TECHN GLASS CO, 2000, EXPL MAT RPL GLASS D
  • [5] Shuttle dose at the Vienna Leksell Gamma Knife
    Ertl, A
    Zehetmayer, M
    Schoggl, A
    Kitz, K
    Koschuch, R
    Stadtmann, H
    Kindl, P
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06) : 1567 - 1578
  • [6] A comparison of dosimetry techniques in stereotactic radiosurgery
    Heydarian, M
    Hoban, PW
    Beddoe, AH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) : 93 - 110
  • [7] DOSE-RATE DEPENDENCE OF A PTW DIAMOND DETECTOR IN THE DOSIMETRY OF A 6 MV PHOTON-BEAM
    HOBAN, PW
    HEYDARIAN, M
    BECKHAM, WA
    BEDDOE, AH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (08) : 1219 - 1229
  • [8] Application of a newly developed photoluminescence glass dosimeter for measuring the absorbed dose in individual mice exposed to low-dose rate 137Cs γ-rays
    Hoshi, Y
    Nomura, T
    Oda, T
    Iwasaki, T
    Fujita, K
    Ishikawa, T
    Kato, A
    Ikegami, T
    Sakai, K
    Tanooka, H
    Yamada, T
    [J]. JOURNAL OF RADIATION RESEARCH, 2000, 41 (02) : 129 - 137
  • [9] Dosimetry and mechanical accuracy of the first rotating gamma system installed in North America
    Kubo, HD
    Araki, F
    [J]. MEDICAL PHYSICS, 2002, 29 (11) : 2497 - 2505
  • [10] Evaluation and characterization of parallel plate microchamber's functionalities in small beam dosimetry
    Lee, HR
    Pankuch, M
    Chu, JC
    Spokas, JJ
    [J]. MEDICAL PHYSICS, 2002, 29 (11) : 2489 - 2496