A pitfall of false lumen embolization in chronic aortic dissection: intimal injury caused by the embolization device edge

被引:18
作者
Furukawa, Tomokuni [1 ]
Uchida, Naomichi [1 ]
Yamane, Yoshitaka [1 ]
Yamada, Kazunori [1 ]
机构
[1] Akane Fdn Tsuchiya Gen Hosp, Dept Cardiovasc Surg, Hiroshima, Japan
关键词
COVERED STENT DEVICES;
D O I
10.1093/icvts/ivw279
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report a case of intimal injury caused by the occluder device in the false lumen (FL) after treatment of refractory chronic aortic dissection with FL embolization. We speculate that the intimal injury was due to the disproportionate stress from the FL. We covered the new entry by an additional stent graft in the true lumen. The deployment of a stent device in both lumens at the level of embolization might be indispensable for FL embolization. © The Author 2016. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.
引用
收藏
页码:153 / 155
页数:3
相关论文
共 5 条
  • [1] Combined proximal stent grafting plus distal bare metal stenting for management of aortic dissection: Superior to standard endovascular repair?
    Hofferberth, Sophie C.
    Newcomb, Andrew E.
    Yii, Michael Y.
    Yap, Kelvin K.
    Boston, Raymond C.
    Nixon, Ian K.
    Mossop, Peter J.
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2012, 144 (04) : 956 - 962
  • [2] Outcomes after false lumen embolization with covered stent devices in chronic dissection
    Idrees, Jahanzaib
    Roselli, Eric E.
    Shafii, Susan
    Reside, Joshua
    Lytle, Bruce W.
    [J]. JOURNAL OF VASCULAR SURGERY, 2014, 60 (06) : 1507 - 1513
  • [3] Kölbel T, 2013, J ENDOVASC THER, V20, P484, DOI 10.1583/13-4318.1
  • [4] Periscope graft to extend distal landing zone in ruptured thoracoabdominal aneurysms with short distal necks
    Rancic, Zoran
    Pfammatter, Thomas
    Lachat, Mario
    Hechelhammer, Lukas
    Frauenfelder, Thomas
    Veith, Frank J.
    Criado, Frank J.
    Mayer, Dieter
    [J]. JOURNAL OF VASCULAR SURGERY, 2010, 51 (05) : 1293 - 1296
  • [5] Use of Covered Stent Devices for False Lumen Embolization in Chronic Dissection: A Novel Approach
    Roselli, Eric E.
    Idrees, Jahanzaib
    Reside, Joshua
    Shafii, Susan
    [J]. ANNALS OF THORACIC SURGERY, 2014, 98 (02) : 737 - 739