Performance of the Hp(10) and Hp(0.07) Measurable Electronic Pocket Dosimeter for Gamma- and Beta-Rays

被引:7
作者
Takahashi, Masa [1 ]
Sekiguchi, Masato [1 ]
Miyauchi, Hideaki [1 ]
Tachibana, Haruo [1 ]
Yoshizawa, Michio [1 ]
Kato, Toru [2 ]
Yamaguchi, Akihito [2 ]
机构
[1] Japan Atom Energy Agcy, Tokai, Ibaraki 3191195, Japan
[2] ALOKA CO LTD, Mitaka, Tokyo 1818622, Japan
关键词
electronic dosimeter; Hp(10); Hp(0.07); response; Monte Carlo simulation;
D O I
10.1080/00223131.2008.10875828
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
For individual monitoring for external exposure of workers, electronic dosimeters have been widely used as supplementary dosimeters in nuclear facilities. In this study, an electronic pocket dosimeter for measuring the H-p(10) and H-p(0.07), which was recently developed by Aloka Co., Ltd., has been examined in terms of (1) energy dependences for gamma- and beta-rays and (2) applicability in actual radiation workplaces. For the latter purpose, performance tests were carried out in hot cells of the Reactor Fuel Examination Facility in the Japan Atomic Energy Agency. From the examinations, the followings were found; (1) the energy responses vary within +/- 20 % in the intended energy range for both radiations. (2) At the actual radiation workplaces, the readings of the dosimeter tend to show slightly higher H-p(10) and lower H-p(0.07) than those of other types of dosimeters, i.e. glass dosimeters and thermoluminescence dosimeters.
引用
收藏
页码:225 / 228
页数:4
相关论文
共 5 条
[1]  
*ICRU MEAS, 56 ICRU MEAS
[2]  
ISO, 1999, 403731999 ISO
[3]  
ITO M, 2001, 2001048 JAERITECH
[4]  
*NAGASE CO LTD, 1985, TL BUDG TECHN NOT
[5]  
Nelson W., 1985, SLAC265