Is zirconia a viable alternative to titanium for oral implant? A critical review

被引:180
作者
Sivaraman, Karthik [1 ]
Chopra, Aditi [2 ]
Narayan, Aparna I. [1 ]
Balakrishnan, Dhanasekar [1 ]
机构
[1] Manipal Univ, Manipal Coll Dent Sci, Dept Prosthodont, Manipal 576104, Karnataka, India
[2] Manipal Univ, Manipal Coll Dent Sci, Dept Periodontol, Manipal, Karnataka, India
关键词
Zirconia; Titanium; Oral implants; Implant materials; Osseointegration; LOW-TEMPERATURE DEGRADATION; OSTEOBLAST-LIKE CELLS; HARD-TISSUE RESPONSE; ALL-CERAMIC SYSTEMS; IN-VIVO; DENTAL IMPLANTS; FLEXURAL STRENGTH; FRACTURE STRENGTH; 2-PIECE ZIRCONIA; SURFACE-ROUGHNESS;
D O I
10.1016/j.jpor.2017.07.003
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: Titanium based implant systems, though considered as the gold standard for rehabilitation of edentulous spaces, have been criticized for many inherent flaws. The onset of hypersensitivity reactions, biocompatibility issues, and an unaesthetic gray hue have raised demands for more aesthetic and tissue compatible material for implant fabrication. Zirconia is emerging as a promising alternative to conventional Titanium based implant systems for oral rehabilitation with superior biological, aesthetics, mechanical and optical properties. This review aims to critically analyze and review the credibility of Zirconia implants as an alternative to Titanium for prosthetic rehabilitation. Study selection: The literature search for articles written in the English language in PubMed and Cochrane Library database from 1990 till December 2016. The following search terms were utilized for data search: "zirconia implants" NOT "abutment", "zirconia implants" AND "titanium implants" AND "osseointegration", "zirconia implants" AND compatibility. Results: The number of potential relevant articles selected were 47. All the human in vivo clinical, in vitro, animals' studies were included and discussed under the following subheadings: Chemical composition, structure and phases; Physical and mechanical properties; Aesthetic and optical properties; Osseointegration and biocompatibility; Surface modifications; Peri-implant tissue compatibility, inflammation and soft tissue healing, and long-term prognosis. Conclusions: Zirconia implants are a promising alternative to titanium with a superior soft-tissue response, biocompatibility, and aesthetics with comparable osseointegration. However, further long-term longitudinal and comparative clinical trials are required to validate zirconia as a viable alternative to the titanium implant. (c) 2017 Japan Prosthodontic Society. Published by Elsevier Ltd.
引用
收藏
页码:121 / 133
页数:13
相关论文
共 129 条
[41]  
Glauser R, 2004, INT J PROSTHODONT, V17, P285
[42]   Influence of grinding, sandblasting, polishing and heat treatment on the flexural strength of a glass-infiltrated alumina-reinforced dental ceramic [J].
Guazzato, M ;
Albakry, M ;
Quach, L ;
Swain, MV .
BIOMATERIALS, 2004, 25 (11) :2153-2160
[43]   Osteoblast cell adhesion on a laser modified zirconia based bioceramic [J].
Hao, L ;
Lawrence, J ;
Chian, KS .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2005, 16 (08) :719-726
[44]   Relative translucency of six all-ceramic systems. Part I: Core materials [J].
Heffernan, MJ ;
Aquilino, SA ;
Diaz-Arnold, AM ;
Haselton, DR ;
Stanford, CM ;
Vargas, MA .
JOURNAL OF PROSTHETIC DENTISTRY, 2002, 88 (01) :4-9
[45]   Relative translucency of six all-ceramic systems. Part II: Core and veneer materials [J].
Heffernan, MJ ;
Aquilino, SA ;
Diaz-Arnold, AM ;
Haselton, DR ;
Stanford, CM ;
Vargas, MA .
JOURNAL OF PROSTHETIC DENTISTRY, 2002, 88 (01) :10-15
[46]   Response of osteoblast-like SAOS-2 cells to zirconia ceramics with different surface topographies [J].
Hempel, Ute ;
Hefti, Thomas ;
Kalbacova, Marie ;
Wolf-Brandstetter, Cornelia ;
Dieter, Peter ;
Schlottig, Falko .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (02) :174-181
[47]   STABILITY OF TETRAGONAL ZRO2 PARTICLES IN CERAMIC MATRICES [J].
HEUER, AH ;
CLAUSSEN, N ;
KRIVEN, WM ;
RUHLE, M .
JOURNAL OF THE AMERICAN CERAMIC SOCIETY, 1982, 65 (12) :642-650
[48]   Fracture strength after dynamic loading of endodontically treated teeth restored with different post-and-core systems [J].
Heydecke, G ;
Butz, F ;
Hussein, A ;
Strub, JR .
JOURNAL OF PROSTHETIC DENTISTRY, 2002, 87 (04) :438-445
[49]  
Hoffmann O, 2008, INT J ORAL MAX IMPL, V23, P691
[50]   Influence of implant diameters on the integration of screw implants - An experimental study in rabbits [J].
Ivanoff, CJ ;
Sennerby, L ;
Johansson, C ;
Rangert, B ;
Lekholm, U .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1997, 26 (02) :141-148