Agreement and repeatability of four different devices to measure noninvasive tear breakup time (NIBUT)

被引:31
作者
Bandlitz, Stefan [1 ,2 ]
Peter, Barbara [3 ]
Pflugi, Tanja [3 ]
Jaeger, Kai [1 ]
Anwar, Aaisha [2 ]
Bikhu, Paramdeep [2 ]
Nosch, Daniela S. [3 ]
Wolffsohn, James S. [2 ]
机构
[1] Hohere Fachschule Augenopt Koln, Cologne Sch Optometry, Bayenthalgurtel 6-8, D-50968 Cologne, Germany
[2] Aston Univ, Optometry & Vis Sci, Birmingham, W Midlands, England
[3] Univ Appl Sci & Arts Northwestern Switzerland, Inst Optometry, FHNW, Olten, Switzerland
关键词
Non-invasive breakup time; Tear film stability; Dry eye; NIBUT; Repeatability; Tearscope; Polaris; Easy-tear; Keratograph; FILM STABILITY; DRY EYE; UP TIME; FLUORESCEIN; VALUES;
D O I
10.1016/j.clae.2020.02.018
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: Since tear film stability can be affected by fluorescein, the Dry Eye Workshop (DEWSII) recommended non-invasive measurement of tear breakup time (NIBUT). The aim of this study was to investigate the agreement and repeatability of four different instruments in the measurement of NIBUT. Methods: 72 participants (mean 24.2 +/- 3.6 years) were recruited for this multi-centre, cross-sectional study. NIBUT was measured three times from one eye using each of the instruments in randomized order on two separate sessions during a day, separated by at least 2 h. NIBUT was performed at three sites (Switzerland, Germany and UK) using three subjective instruments, Tearscope Plus (Keeler, Windsor, UK) (TS), Polaris (bon Optic, Lubeck, Germany) (POL), EasyTear Viewplus (Easytear, Rovereto, Italy) (ET) and the objective Keratograph 5 M (Oculus Optikgerate GmbH, Wetzlar, Germany) (KER). As the latter instrument only analyses for 24 s, all data was capped at this value. Results: NIBUT measurements (average of both sessions) between the four instruments were not statistically significantly different: TS (median 10.4, range 2.0-24.0 s), POL (10.1, 1.0-24.0 s), ET (10.6, 1.0-24.0 s) and KER (11.1, 2.6-24.0 s) (p = 0.949). The objective KER measures were on average (1.2 s +/- 9.6 s, 95 % confidence interval) greater than the subjective evaluations of NIBUT with the other instruments (mean difference 0.4 s +/- 7.7 s, 95 % confidence interval), resulting in a higher limits of agreement. The slope was -0.08 to 0.11 indicating no bias in the difference between instruments with the magnitude of the NIBUT. Repeated measurements from the two sessions were not significantly different for TS (p = 0.584), POL (p = 0.549), ET (p = 0.701) or KER (p = 0.261). Conclusions: The four instruments evaluated for their measurement of tear stability were reasonably repeatable and give similar average results.
引用
收藏
页码:507 / 511
页数:5
相关论文
共 27 条
[1]  
Abelson MB, 2002, ADV EXP MED BIOL, V506, P1121
[2]  
Bandlitz S., 2016, Optom Pract, V17, P81
[3]   Clinical evaluation of the Oculus Keratograph [J].
Best, N. ;
Drury, L. ;
Wolffsohn, J. S. .
CONTACT LENS & ANTERIOR EYE, 2012, 35 (04) :171-174
[4]   RELIABILITY OF THE TEAR BREAK-UP TIME TECHNIQUE OF ASSESSING TEAR STABILITY AND THE LOCATIONS OF THE TEAR BREAK-UP IN HONG-KONG CHINESE [J].
CHO, P ;
BROWN, B ;
CHAN, I ;
CONWAY, R ;
YAP, M .
OPTOMETRY AND VISION SCIENCE, 1992, 69 (11) :879-885
[5]   Agreement between Automated and Traditional Measures of Tear Film Breakup [J].
Cox, Stephanie M. ;
Nichols, Kelly K. ;
Nichols, Jason J. .
OPTOMETRY AND VISION SCIENCE, 2015, 92 (09) :E257-E263
[6]   Agreement and repeatability of objective systems for assessment of the tear film [J].
Fernandez, Joaquin ;
Rodriguez-Vallejo, Manuel ;
Martinez, Javier ;
Tauste, Ana ;
Garcia-Montesinos, Javier ;
Pinero, David P. .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2018, 256 (08) :1535-1541
[7]   Repeatability of Noninvasive Keratograph 5M Measurements Associated With Contact Lens Wear [J].
Garcia-Montero, Maria ;
Rico-del-Viejo, Laura ;
Lorente-Velazquez, Amalia ;
Martinez-Alberquilla, Irene ;
Luis Hernandez-Verdejo, Jose ;
Madrid-Costa, David .
EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2019, 45 (06) :377-381
[8]   Preocular tear film characteristics of nonwearers and soft contact lens wearers [J].
Guillon, M ;
Styles, E ;
Guillon, JP ;
Maissa, C .
OPTOMETRY AND VISION SCIENCE, 1997, 74 (05) :273-279
[9]   Noninvasive Assessment of Tear Stability with the Tear Stability Analysis System in Tear Dysfunction Patients [J].
Gumus, Koray ;
Crockett, Charlene Hong ;
Rao, Kavita ;
Yeu, Elizabeth ;
Weikert, Mitchell P. ;
Shirayama, Mariko ;
Hada, Shigeki ;
Pflugfelder, Stephen C. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (01) :456-461
[10]   Assessment of Tear Film Stability in Dry Eye With a Newly Developed Keratograph [J].
Hong, Jiaxu ;
Sun, Xinghuai ;
Wei, Anji ;
Cui, Xinhan ;
Li, Yimin ;
Qian, Tingting ;
Wang, Wentao ;
Xu, Jianjiang .
CORNEA, 2013, 32 (05) :716-721