Impact of Target Lesion Coronary Calcification on Stent Expansion - An Optical Coherence Tomography Study

被引:204
作者
Kobayashi, Yuhei [1 ]
Okura, Hiroyuki [1 ]
Kume, Teruyoshi [1 ]
Yamada, Ryotaro [1 ]
Kobayashi, Yukari [1 ]
Fukuhara, Kenzo [1 ]
Koyama, Terumasa [1 ]
Nezuo, Shintaro [2 ]
Neishi, Yoji [1 ]
Hayashida, Akihiro [1 ]
Kawamoto, Takahiro [1 ]
Yoshida, Kiyoshi [1 ]
机构
[1] Kawasaki Med Sch, Kurashiki, Okayama, Japan
[2] Kawasaki Hosp, Okayama, Japan
关键词
Coronary artery disease; Drug-eluting stents; Optical coherence tomography; DRUG-ELUTING STENT; INTRAVASCULAR ULTRASOUND ASSESSMENT; IMPLANTATION; THROMBOSIS; UNDEREXPANSION; OUTCOMES;
D O I
10.1253/circj.CJ-14-0108
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Stent underexpansion remains a concern as a cause of drug-eluting stent (DES) failure. Although coronary calcification is considered to be a contributing factor in stent underexpansion, previous intravascular ultrasound studies have failed to demonstrate this relationship. We investigated whether stent expansion could be predicted by coronary calcification as assessed by optical coherence tomography (OCT). Methods and Results: We enrolled 51 de novo native coronary artery lesions treated by a single 2nd-generation DES (3 types). Prior to stent deployment, the arc and area of calcium at the target lesion were measured using OCT. After successful stent implantation, OCT imaging was repeated to assess minimal stent diameter and area (MSD and MSA). Stent expansion was defined as MSD (or MSA) divided by the values predicted by the manufacturers' compliance charts. Patients were divided into 4 groups according to the median values of the arc and area of calcium. Mean stent expansion was 73.3 +/- 8.7% for MSD and 65.2 +/- 12.0% for MSA. Stent expansion defined by MSD was significantly different among the 4 groups (P=0.02). A similar trend was observed for stent expansion defined by MSA (P=0.16). Conclusions: The extent of target lesion calcification as assessed by OCT may be an important determinant of the expansion of 2nd-generation DES.
引用
收藏
页码:2209 / 2214
页数:6
相关论文
共 24 条
[11]   Randomized comparison of the nobori biolimus A9-eluting stent with the sirolimus-eluting stent in patients with stenosis in native coronary arteries [J].
Kadota, Kazushige ;
Muramatsu, Toshiya ;
Iwabuchi, Masashi ;
Saito, Shigeru ;
Hayashi, Yasuhiko ;
Ikari, Yuji ;
Nanto, Shinsuke ;
Fujii, Kenshi ;
Inoue, Naoto ;
Namiki, Atsuo ;
Kimura, Takeshi ;
Mitsudo, Kazuaki .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2012, 80 (05) :789-796
[12]   Impact of a Prolonged Delivery Inflation Time for Optimal Drug-Eluting Stent Expansion [J].
Kawasaki, Tomohiro ;
Koga, Hisashi ;
Serikawa, Takeshi ;
Orita, Yoshiya ;
Ikeda, Shinsuke ;
Mito, Takahiro ;
Gotou, Yoshitaka ;
Shintani, Yoshiaki ;
Tanaka, Atsushi ;
Tanaka, Hidenori ;
Fukuyama, Takaya ;
Koga, Nobuhiko .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2009, 73 (02) :205-211
[13]   Assessment of coronary arterial plaque by optical coherence tomography [J].
Kume, T ;
Akasaka, T ;
Kawamoto, T ;
Watanabe, N ;
Toyota, E ;
Neishi, Y ;
Sukmawan, R ;
Sadahira, Y ;
Yoshida, K .
AMERICAN JOURNAL OF CARDIOLOGY, 2006, 97 (08) :1172-1175
[14]   Natural History of Stent Edge Dissection, Tissue Protrusion and Incomplete Stent Apposition Detectable Only on Optical Coherence Tomography After Stent Implantation - Preliminary Observation [J].
Kume, Teruyoshi ;
Okura, Hiroyuki ;
Miyamoto, Yoshinori ;
Yamada, Ryotaro ;
Saito, Ken ;
Tamada, Tomoko ;
Koyama, Terumasa ;
Neishi, Yoji ;
Hayashida, Akihiro ;
Kawamoto, Takahiro ;
Yoshida, Kiyoshi .
CIRCULATION JOURNAL, 2012, 76 (03) :698-703
[15]   Assessment of the coronary calcification by optical coherence tomography [J].
Kume, Teruyoshi ;
Okura, Hiroyuki ;
Kawamoto, Takahiro ;
Yamada, Ryotaro ;
Miyamoto, Yoshinori ;
Hayashida, Akihiro ;
Watanabe, Nozomi ;
Neishi, Yoji ;
Sadahira, Yoshito ;
Akasaka, Takashi ;
Yoshida, Kiyoshi .
EUROINTERVENTION, 2011, 6 (06) :768-772
[16]   Volumetric Characterization of Human Coronary Calcification by Frequency-Domain Optical Coherence Tomography [J].
Mehanna, Emile ;
Bezerra, Hiram G. ;
Prabhu, David ;
Brandt, Eric ;
Chamie, Daniel ;
Yamamoto, Hirosada ;
Attizzani, Guilherme F. ;
Tahara, Satoko ;
Van Ditzhuijzen, Nienke ;
Fujino, Yusuke ;
Kanaya, Tomoaki ;
Stefano, Gregory ;
Wang, Wei ;
Gargesha, Madhusudhana ;
Wilson, David ;
Costa, Marco A. .
CIRCULATION JOURNAL, 2013, 77 (09) :2334-2340
[17]   An optimal diagnostic threshold for minimal stent area to predict target lesion revascularization following stent implantation in native coronary lesions [J].
Morino, Y ;
Honda, Y ;
Okura, H ;
Oshima, A ;
Hayase, M ;
Bonneau, HN ;
Kuntz, RE ;
Yock, PG ;
Fitzgerald, PJ .
AMERICAN JOURNAL OF CARDIOLOGY, 2001, 88 (03) :301-+
[18]   Impact of final Stent dimensions on long-term results following sirolimus-eluting stent implantation - Serial intravascular ultrasound analysis from the SIRIUS trial [J].
Sonoda, S ;
Morino, Y ;
Ako, J ;
Terashima, M ;
Hassan, AHM ;
Bonneau, HN ;
Leon, MB ;
Moses, JW ;
Yock, PG ;
Honda, Y ;
Kuntz, RE ;
Fitzgerald, PJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (11) :1959-1963
[19]   A Prospective, Randomized Evaluation of a Novel Everolimus-Eluting Coronary Stent [J].
Stone, Gregg W. ;
Teirstein, Paul S. ;
Meredith, Ian T. ;
Farah, Bruno ;
Dubois, Christophe L. ;
Feldman, Robert L. ;
Dens, Joseph ;
Hagiwara, Nobuhisa ;
Allocco, Dominic J. ;
Dawkins, Keith D. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (16) :1700-1708
[20]   Optimizing stent expansion with new stent delivery systems [J].
Takano, Y ;
Yeatman, LA ;
Higgins, JR ;
Currier, JW ;
Ascencio, E ;
Kopelson, KA ;
Tobis, JM .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 38 (06) :1622-1627