Interventional atrial septal defect closure using a totally bioresorbable occluder matrix - Development and preclinical evaluation of the BioSTAR device

被引:63
作者
Jux, Christian
Bertram, Harald
Wohlsein, Peter
Bruegmann, Michael
Paul, Thomas
机构
[1] Univ Gottingen, Dept Pediat Cardiol & Pediat Intens Care Med, D-37099 Gottingen, Germany
[2] Hannover Med Sch, D-3000 Hannover, Germany
[3] Sch Vet Med, Inst Pathol, Hannover, Germany
关键词
D O I
10.1016/j.jacc.2006.02.057
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES We sought to test the hypothesis that interventional atrial septal defect (ASD) closure can be performed safely and effectively using a bioresorbable occluder matrix. BACKGROUND The ideal septal occluder scaffold should promote the healthiest and most complete heating response while eventually facilitating the full resorption of the material and leaving "native" tissue behind, thus minimizing the potential for future complications from chronic foreign body and maintaining the possibility for later unobstructed transseptal access to the left atrium. METHODS The STARFlex occluders (NMT Medical Inc., Boston, Massachusetts) were modified by substituting the conventional polyester fabric for a bioengineered, acellular type-1 collagen matrix derived from porcine submucosa with a heparin-coated surface (BioSTAR occluder, NMT Medical Inc.). Comparative transcatheter closure of ASDs was performed in young sheep (n = 36). Gross pathology and histopathology were obtained after follow-up periods ranging from 7 days to 2 years. RESULTS The STARFlex (control) devices were encapsulated time-dependently by ingrown fibrous tissue. Histology showed a mild but chronically persisting foreign body reaction. By contrast, BioSTAR devices exhibited a mild-to-moderate transient cellular immune response. Heparin coating of the BioSTAR surface improved the biocompatibility of the device by reducing surface thrombogencity. A remodeling process of the collagen scaffold, starting after 30 days in vivo, resulted in the full replacement of the matrix by host tissue after 2 years of follow-up. CONCLUSIONS The BioSTAR device is the first septal occluder with a totally bioresorbable matrix that is fully replaced by host tissue during the healing process. The promising results of this study support testing of the BioSTAR device in clinical trials.
引用
收藏
页码:161 / 169
页数:9
相关论文
共 37 条
[11]  
Edelman David S, 2002, JSLS, V6, P203
[12]  
Franklin M E Jr, 2004, Hernia, V8, P186
[13]   Adsorption of inflammatory cytokines using a heparin-coated extracorporeal circuit [J].
Fujita, M ;
Ishihara, M ;
Ono, K ;
Hattori, H ;
Kurita, A ;
Shimizu, M ;
Mitsumaru, A ;
Segawa, D ;
Hinokiyama, K ;
Kusama, Y ;
Kikuchi, M ;
Maehara, T .
ARTIFICIAL ORGANS, 2002, 26 (12) :1020-1025
[14]   HEPARIN-COATED CIRCUITS REDUCE THE INFLAMMATORY RESPONSE TO CARDIOPULMONARY BYPASS [J].
GU, YJ ;
VANOEVEREN, W ;
AKKERMAN, C ;
BOONSTRA, PW ;
HUYZEN, RJ ;
WILDEVUUR, CRH .
ANNALS OF THORACIC SURGERY, 1993, 55 (04) :917-922
[15]  
Hausdorf G, 2001, J Interv Cardiol, V14, P69, DOI 10.1111/j.1540-8183.2001.tb00715.x
[16]   Remodeling of an acellular collagen graft into a physiologically responsive neovessel [J].
Huynh, T ;
Abraham, G ;
Murray, J ;
Brockbank, K ;
Hagen, PO ;
Sullivan, S .
NATURE BIOTECHNOLOGY, 1999, 17 (11) :1083-1086
[17]   Experimental ASD closure using autologous cell-seeded interventional closure devices [J].
Jux, C ;
Bertram, H ;
Wohlsein, P ;
Brügmann, M ;
Wüboldt, P ;
Fink, C ;
Paul, T ;
Hausdorf, G .
CARDIOVASCULAR RESEARCH, 2002, 53 (01) :181-191
[18]  
KADLER K, 1994, PROTEIN PROFILE, V1, P519
[19]   Effect of biologically active coating on biocompatibility of Nitinol devices designed for the closure of intra-atrial communications [J].
Kong, XQ ;
Grabitz, RG ;
van Oeveren, W ;
Klee, D ;
van Kooten, TG ;
Freudenthal, F ;
Qing, M ;
von Bernuth, G ;
Seghaye, MC .
BIOMATERIALS, 2002, 23 (08) :1775-1783
[20]   Healing response to the Clamshell device for closure of intracardiac defects in humans [J].
Kreutzer, J ;
Ryan, CA ;
Gauvreau, K ;
Van Praagh, R ;
Anderson, JM ;
Jenkins, KJ .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2001, 54 (01) :101-111