Label-free detection of C-reactive protein using a carbon nanofiber based biosensor

被引:108
作者
Gupta, Rakesh K. [1 ,2 ]
Periyakaruppan, Adaikkappan [1 ,3 ]
Meyyappan, M. [1 ]
Koehne, Jessica E. [1 ]
机构
[1] NASA Ames Res Ctr, Moffett Field, CA 94035 USA
[2] GGM Sci Coll, Dept Elect, Jammu 18004, Jammu & Kashmir, India
[3] NASA Johnson Space Ctr, Houston, TX 77058 USA
关键词
Biosensors; Carbon nanofibers; C-reactive protein; Nanoelectrode array; Electrical impedance spectroscopy etc; NANOELECTRODE ARRAYS; ELECTRODE ARRAYS; CARDIAC MARKERS; BIOMARKERS;
D O I
10.1016/j.bios.2014.03.027
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
We report the sensitive detection of C-reactive protein (CRP), a biomarker for cardiac disease, using a carbon nanofiber based biosensor platform. Vertically aligned carbon nanofibers were grown using plasma enhanced chemical vapor deposition to fabricate nanoelectrode arrays in a 3 x 3 configuration. Cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) were used for the CRP detection. The CV responses show a 25% reduction in redox current upon the immobilization of anti-CRP on the electrode where as a 30% increase in charge transfer resistance is seen from EIS. Further reduction in redox current and increase in charge transfer resistance result from binding of CRP on anti-CRP immobilized surface, proportional to the concentration of the CRP target. The detection limit of the sensor is found to be similar to 90 pM or similar to 11 ng/ml, which is in the clinically relevant range. Control tests using non-specific myoglobin antigen confirmed the specificity of the present approach. Published by Elsevier B.V.
引用
收藏
页码:112 / 119
页数:8
相关论文
共 34 条
[1]   Future biomarkers for detection of ischemia and risk stratification in acute coronary syndrome [J].
Apple, FS ;
Wu, AHB ;
Mair, J ;
Ravkilde, J ;
Panteghini, M ;
Tate, J ;
Pagani, F ;
Christenson, RH ;
Mockel, M ;
Danne, O ;
Jaffe, AS .
CLINICAL CHEMISTRY, 2005, 51 (05) :810-824
[2]   Vertically aligned carbon nanofiber electrode arrays for nucleic acid detection [J].
Arumugam, Prabhu U. ;
Yu, Edmond ;
Riviere, Roger ;
Meyyappan, M. .
CHEMICAL PHYSICS LETTERS, 2010, 499 (4-6) :241-246
[3]   Wafer-scale fabrication of patterned carbon nanofiber nanoelectrode arrays: A route for development of multiplexed, ultrasensitive disposable biosensors [J].
Arumugam, Prabhu U. ;
Chen, Hua ;
Siddiqui, Shabnam ;
Weinrich, Jarret A. P. ;
Jejelowo, Ayodeji ;
Li, Jun ;
Meyyappan, M. .
BIOSENSORS & BIOELECTRONICS, 2009, 24 (09) :2818-2824
[4]   CYCLIC AND DIFFERENTIAL PULSE VOLTAMMETRIC BEHAVIOR OF REACTANTS CONFINED TO ELECTRODE SURFACE [J].
BROWN, AP ;
ANSON, FC .
ANALYTICAL CHEMISTRY, 1977, 49 (11) :1589-1595
[5]   An optimised electrochemical biosensor for the label-free detection of C-reactive protein in blood [J].
Bryan, Thomas ;
Luo, Xiliang ;
Bueno, Paulo R. ;
Davis, Jason J. .
BIOSENSORS & BIOELECTRONICS, 2013, 39 (01) :94-98
[6]   Vertically Aligned Carbon Nanofibers: Interconnecting Solid State Electronics with Biosystems [J].
Cassell, Alan M. ;
Li, Jun ;
Nguyen-Vu, Thuy-Duong Barbara ;
Koehne, Jessica E. ;
Chen, Hua ;
Andrews, Russell ;
Meyyappan, M. .
JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2009, 9 (08) :5038-5046
[7]   Protein Detection Using Arrayed Microsensor Chips: Tuning Sensor Footprint to Achieve Ultrasensitive Readout of CA-125 in Serum and Whole Blood [J].
Das, Jagotamoy ;
Kelley, Shana O. .
ANALYTICAL CHEMISTRY, 2011, 83 (04) :1167-1172
[8]   Function of C-reactive protein [J].
Du Clos, TW .
ANNALS OF MEDICINE, 2000, 32 (04) :274-278
[9]   Electrochemical investigations of the interaction of C-reactive protein (CRP) with a CRP antibody chemically immobilized on a gold surface [J].
Hennessey, Hooman ;
Afara, Nadia ;
Omanovic, Sasha ;
Padjen, Ante L. .
ANALYTICA CHIMICA ACTA, 2009, 643 (1-2) :45-53
[10]   Immunodetection of pentamer and modified C-reactive protein using surface plasmon resonance biosensing [J].
Hu, WP ;
Hsu, HY ;
Chiou, A ;
Tseng, KY ;
Lin, HY ;
Chang, GL ;
Chen, SJ .
BIOSENSORS & BIOELECTRONICS, 2006, 21 (08) :1631-1637